8
ENGLISH
Long period switching off
In case of absence for long periods of time, set the mode selector
(1 -
fi
g. 1a) to (OFF).
Then, close the gas tap present on the system. In this case, anti-frost
device is deactivated: empty the systems, in case of risk of frost.
4.4 Light signals and faults
The operating status of the boiler is shown on the digital display,
below is a list of the types of displays.
4.5 Adjustments
The boiler has already been adjusted by the manufacturer.
If it is necessary to adjust it again, for example after extraordinary
maintenance, after replacement of gas valve or after gas conversion,
carry out the following procedure.
The maximum output adjustment must be carried out in
the sequence indicated exclusively by quali
fi
ed personnel.
- remove the shell unscrewing the
fi
xing screws A (
fi
g. 6)
- unscrew by two turns the screw of the pressure test point down-
stream the gas valve and connecting the pressure gauge
- disconnect the compensation inlet of the air distribution box
4.5.1 Ciao S C.S.I.: Maximum power and minimum domestic
hot water adjustment
- Fully open the hot water tap
- on the control panel:
- set the mode selector to
(summer) (
fi
g. 2a)
- turn the domestic hot water temperature selector to its maximum
(
fi
g. 8a)
- power the boiler setting the system main switch to “on”
- check that the pressure on the pressure gauge is stable; or with
a milliammeter in series to the modulator, make sure that the
modulator supplies the maximum available current (120 mA for
G20 and 165 mA for LPG).
- carefully prise out the protection cap of the adjustment screws,
using a screwdriver (
fi
g. 15)
- with a fork spanner CH10 use the adjustment nut of the maximum
output in order to obtain the value indicated in table "Technical data"
- disconnect the modulator faston
- wait until the pressure on the pressure gauge is stable at minimum value
- paying attention not to press the internal shaft, use an Allen
spanner to turn the red adjustment screw for domestic hot water
minimum temperature regulation, calibrating it until the pressure
gauge reads the value indicated in the table "Technical data"
- reconnect the modulator faston
- close the domestic hot water tap
- carefully re
fi
t the protection cap of the adjustment screws.
Ciao S R.S.I.: Maximum and minimum power adjustment
- Set the function selector to
(winter) (
fi
g. 2b)
- Remove the housing and access the card
- Insert the jumpers JP1 and JP2
- Bring the trimmer P2 to the maximum value, using a screwdriver
(turn clockwise)
- Power the boiler setting the system main switch to “on”
- Check the pressure on the pressure gauge is stable; or, with a
milliammeter in series to the modulator, make sure the modulator
supplies the maximum available current (120 mA for G20 and 165
mA for LPG)
- Use a screwdriver to carefully prise out the protection cap of the
adjustment screws
- With a fork spanner CH10, use the adjustment nut of the maximum
output in order to obtain the value indicated in the table "Technical
data"
- Disconnect the modulator faston
- Wait until the pressure on the pressure gauge is stable at minimum
value
- Use an Allen spanner to set the red adjustment screw at the
minimum output, calibrating it until the pressure gauge shows the
value indicated in the table "Technical data"
- Reconnect the modulator faston
- Disconnect the boiler from the power supply
- Remove the jumpers JP1 and JP2
- Carefully re
fi
t the protection cap of the adjustment screws.
4.5.2 Minimum and maximum heating electric adjustment
The “electric adjustment” function is activated and deactivated
exclusively by the jumper (JP1) (
fi
g. 16).
ADJ appears on the display to indicate that the calibration pro-
cedure is underway.
The function can be enabled in the following way:
- by powering the card with the jumper JP1 inserted and the mode
selector in winter position, independently from the possible pres-
ence of other operation request.
- by inserting the jumper JP1, with the mode selector in winter posi-
tion, without heat request in progress.
BOILER STATUS
DISPLAY
Stand-by
-
OFF status
OFF
ACF module lockout alarm
A01
ACF electrical fault alarm
A01
Limit thermostat alarm
A02
Air pressure switch alarm
A03
H2O pressure switch alarm
A04
NTC domestic water fault (C.S.I. and R.S.I only
with external storage heater with probe)
A06
NTC heating fault
A07
Parasite
fl
ame
A11
Electric calibration min and max heating
ADJ
Transient awaiting ignition
88
°C
fl
ashing
Air pressure switch intervention
fl
ashing
H2O pressure switch intervention
fl
ashing
Preheating Function active (only C.S.I.)
Preheating heat request (only C.S.I.)
fl
ashing
External probe present
Domestic water heat request
60
°C
Heating heat request
80
°C
Anti-freeze heat request
Flame present
To restore operation (deactivate alarms):
Faults A 01-02-03
Position the function selector to
(OFF), wait 5-6 seconds then
set it to the required position
(summer mode) or
(winter
mode).
If the reset attempts do not reactivate the boiler, contact the Techni-
cal Assistance Centre.
Fault A 04
In addition to the fault code, the digital display shows the symbol .
Check the pressure value indicated by the water gauge:
if it is less than 0.3 bar, position the function selector to (OFF)
and adjust the
fi
lling tap (L
fi
g. 13 for C.S.I. - external for R.S.I.) until
the pressure reaches a value between 1 and 1.5 bar.
Then position the mode selector to the desired position
(sum-
mer) or
(winter).
The boiler will perform one purge cycle lasting approximately 2
minutes.
If pressure drops are frequent, request the intervention of the Tech-
nical Assistance Service.
Fault A 06 (only C.S.I.)
The boiler operates normally but cannot reliably maintain a con-
stant domestic hot water temperature, which remains set at around
50°C. Contact the Technical Assistance Centre.
Fault A 07
Contact the Technical Assistance Centre.