45
Class II, RInsu-2
Sec. <-> test body
Connection diagram
VDE 0701-0702, VDE 0544-4, VDE 0751-1
Circuit diagram
VDE 0701-0702, VDE 0544-4, VDE 0751-1
Connect the test probes according to the connection diagram and
scan the portable test sample.
The testing voltage is applied to the PE socket. The insulating
resistance is measured between the secondary side and the body
of the test sample.
Class II, RInsu-3
LN <-> sec.
Connection diagram
VDE 0701-0702, VDE 0544-4, VDE 0751-1
Circuit diagram
VDE 0701-0702, VDE 0544-4, VDE 0751-1
Insert the shock-proof plug of the portable test sample into the
test socket of the appliance tester.
The testing voltage is applied to the bridged LN conductor.
The insulating resistance is measured between LN and the
secondary side.
Class III, RInsu-1
Input <-> test body
Connection diagram
VDE 0701-0702
Circuit diagram
VDE 0701-0702
Connect the test probes according to the connection diagram and
scan the portable test sample.
The testing voltage is applied to the PE socket.
The insulating resistance is measured between the input (
IN
) and
the body of the test sample.
Class III, RInsu-2
Output <-> test body
Connection diagram
VDE 0701-0702
Circuit diagram
VDE 0701-0702
Connect the test probes according to the connection diagram and
scan the portable test sample.
The testing voltage is applied to the PE socket.
The insulating resistance is measured between the output (
OUT
)
and the body of the test sample.