Test procedure of the test sample
Please observe the information on the display.
o
Depending on the test procedure used,
individual measurements must be enabled
by tapping the "
Start
" button.
o
In case of negative individual results, the
measurement will be stopped
automatically.
11.3.3.1 Stopping an automatic test
Tap the "
Stop
" button to stop the test procedure.
You will get back to the “
Automatic test
” screen.
11.3.3.2 Interrupting an automatic test
Touch the "
Pause
" button to interrupt the
measurement
(e. g. in case of a changed position of the alligator clip or
test probe or to prevent an incorrect measurement).
Touch the “
OK
" button to continue the
measurement.
Attention! The testing voltage remains during an
interrupted or paused test and is still applied to the test
sample!
11.3.3.3 Resetting measured values
Touch the "
Reset
" button to delete incorrect
measured values and to reset the test duration.
(Resetting the measured values only works when the
test is running, not when the test is interrupted.)
When the test procedure has been completed, the "
Test result
"
screen opens.
Analysis of test results
The message "
Test successful
" or
“
Test failed
” is displayed.
Scroll to the end of the test results and
choose the test type
(
periodic testing, test after modification,
after
repair
or
receiving inspection
).
You can enter an additional comment in
the “
Note
" line.
You can choose between the following buttons:
“
”
(if option available)
“
Discard
”
“
Save
”
11.3.4.1 Print
Touch the “
” button to print the test and
follow the instructions on the touchscreen.
(The appliance tester must be connected to the optional
log printer via Bluetooth
®
.)
11.3.4.2 Discard
Touch the “
Discard
” button to discard the test
and confirm the confirmation prompt on the
display with "
Yes
".
You will get back to the “
Automatic test
” screen.
11.3.4.3 Save
Touch the "
Save
" button to save the test.
The test data will be saved automatically in the database and can
be read out via a PC.
You will get back to the “
Automatic test
” screen.