1 Cautions regarding installation
To avoid the risk of electric shock, this equipment must only be
connected to a supply mains with protective earth.
Failure or electric leak may result in electric shock.
Do not directly expose human eyes to LED light.
Exposure may hurt human eyes.
2. Do not drop the product, let it crash into something, or apply
strong impact.
3. Do not connect an input power supply other than with the
rated power level.
4. Make sure to connect ground wires.
5. Have the clinic equipped with double-pole switch at the
power outlet or the switchboard.
6. Adjust the levels of the unit and chair so that the light stops in
the required position.
Read the installation instructions for the unit and chair
concerning the level adjustment.
(AL-D106G / AL-D109W)
7. Once all work is completed, confirm that each part operates
normally.
WARNING