(6) Setting the lower volume limit of the dental light
intensity
1. Turn on the power of the unit.
2. Turn the light intensity control volume of the dental
light counterclockwise to its lowest setting.
3. Push the switch (SW1, CHECK) on the light control
PCB in the power supply box.
LED2 (Yellow) on the light control PCB brinks.
4. Push the switch (SW2, ON_OFF) on the light control
PCB.
LED2 (Yellow) on the light control PCB turns off.
3 Installation
Light intensity
control volume
Light control PCB
SW1
SW2
LED2
3-2-6 AL-D107G
Installation instruction, see in cabinet installation instractions.