Attached to the ceiling framework to
provide 200lb (90kg) dead weight capacity
(C) In suspended ceilings
(2) Electrical preparation
Refer to below figure for the location of electrical feed
opening in pallet, provide 3 wire, circuit (fuse or breaker)
through flexible conduit with enough slack to protrude at
least 2” (50mm) below the pallet when installed.
A readily accessible shut-off switch for this circuit is
recommended. Use wiring suitable for 194˚F (90˚C) service.
6-5/16
(160)
4-1/4 (108)
24 (610)
8 (203)
16 (406)
65-1/4
(
1657)
11/16 (16.5)
3-5/8
(92)
4-15/16
(125)
φ
1-9/16 (40) DIA
(Electrical Feed Opening)
φ
3/8 (9.5) DIA HOLE
φ
1 (25.4) DIA C'Bore (8Holes)
Head of
Chair
3 Installation
Coach bolt
(M8 x 75, plated)
Unit : inch (mm)
26