Bel-Cypher N
2. Confirmation of movement with X-ray irradiation.
2.1.
Check listed items by referring to the Operation Manual.
2.2. Confirm the operation with X-ray
2.2. 1. Cover radiation aperture with lead.
2.2. 2. Set exposure condition by referring the Operation Manual.
1) Exposure Orbit
→
Panorama
5) Manual Exposure
6) Tube Voltage: 60kV
7) Tube Current: 2mA
2.2.3. Irradiate X-Ray and confirm the operation.
Figure 2
CP11
CP12
CPCG