Bel-Cypher N
10. Post-Installation Instructions
1. Check listed items by referring to the Operation Manual.
1) Measurement of Feedback voltage and tube current
①
Make sure the main power is off. Remove covers from Rotation Unit. (Figure 1)
②
Cover radiation aperture with lead
③
Turn on the power of PC that is connected to Bel-Cypher N. Hit Ctrl + Alt + T simultaneously to get
into Test mode.
④
Set exposure condition by referring the Operation Manual.
1) Exposure Orbit
→
Panorama
2) Manual Exposure
3) Tube Voltage: 70kV
4) Tube Current: 8mA
⑤
Irradiate X-ray, then measure the feedback voltage between CP11 & CPCG (Tube voltage),and
between CP12 & CPCG (Tube current). (Figure 2)