Bel-Cypher  N 

11  Mounting  Rotation  Unit  ASSY  on  Sliding  Unit  ASSY 

11.  1.  By  holding  carriage  holder,  with  two  men,  hook  the  Rotation  Unit  ASSY  onto  Sliding  Unit. 

Align  rotation  unit  by  using  taper  pins.   

 

 

    11.  2.  Insert  taper  pins  with  a  hammer. 

 

   

11.  3.  Tighten  and  Fix  hexagon  bolts. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Hook 

Rotation Unit ASSY

 

Pin 

(1): Taper Pin             

φ

5X25                2sets. 

(2): hexagon bolts     

φ

8X25                4sets. 

(3): Spring                   

φ

8                        4sets. 

(4):Flat             

   

   

φ

8                        4sets. 

Содержание Bel-Cypher N

Страница 1: ...tice Read this installation Manual thoroughly before Installation The classification is shown as follows According to the type of protection against electric shock Class According to the degree of pro...

Страница 2: ...and certification procedures for the Bel Cypher N X Ray The instructions contained in this book should be thoroughly read and understood before attempting to install the Bel Cypher N unit After the i...

Страница 3: ...surement of equipment 05 1 3 6 Checking of each components and accessories 06 1 3 7 Installation Procedure 07 1 14 8 Methods to install LAN Card for Panoramic Radiography 08 1 3 9 Methods to install a...

Страница 4: ...Installation Manual 4 After installation read Operation Manual to understand operation procedures 5 Discharge Be sure to observe Installation Manual If accidents or troubles of the equipment happen d...

Страница 5: ...ly conducted by an au thorized service personnel only Meanings DANGER Explains danger that may cause serious adverse effect to a human body WARNING Explains an instruction where personal injury or phy...

Страница 6: ...ore than 20kg Heavy units should be brought up by more than 2 people Do not put things in area where equipment functions In order to avoid influence by other electric equipment use separated power sou...

Страница 7: ...ill in wall and a floor 2 3 Others 1 Ethanol for disinfections 2 Waste Clothe 3 Cleanser 3 User supplied Network Card NIC and PC NOTE IN GENERAL MAJOR STRUCTURAL MODIFICATIONS ARE NOT REQUIRED HOWEVER...

Страница 8: ...ic Bitewing Adult Child TMJ Exposure Time Panorama 10 sec Panoramic Bitewing 4 6 sec TMJ 2 5 sec x 4 Magnification Ratio Panorama 1 2 to 1 3 Panoramic Bitewing 1 2 to 1 3 TMJ 1 2 Image receiver Digita...

Страница 9: ...ach parts and Dimension of equipment Names of each parts of equipment Arm Section X Ray Generator Pillar Chinrest Unit Exposure Switch Rotation Unit Control Unit Digital Sensor Grip Base Sliding Unit...

Страница 10: ...Bel Cypher N 2 Dimensions of equipment with Free Standing Base...

Страница 11: ...Bel Cypher N Dimensions of equipment without Free Standing Base...

Страница 12: ...sher 6 4 pcs Flat Washer 6 4 pcs Hexagon socket flat screw M6 x 16 4 pcs Washer 4 pcs 13 Fixing bolts for Sliding Unit ASSY and Rotation Unit ASSY Hexagon Bolt M8 x 30 4 pcs Flat Washer 8 4 pcs Spring...

Страница 13: ...Bel Cypher N Accessories 2 1 Accessories 1 Head Holding Rods Panorama 2 Chinrest Panorama 3 Bite Block Panorama 4 Ear Rods TMJ 4 divisions...

Страница 14: ...Bel Cypher N 2 2 Expendable Supplies Bite Block Cover 3 X ray Exposure Switch with Holder 4 CD containing TWAIN Driver...

Страница 15: ...r waste clothe on the pillow for protecting scratch of Pillar Then put the Pillar on the pillows Sliding Unit ASSY is down side 2 Remove pillar cover from Pillar Assy Take out pillows and put them on...

Страница 16: ...Free Standing Base Type Floor Type 4 Attach the Wall Bracket to upper part of the Pillar by 2 flat head cap screws 1 Cap Screw M10 X 30 6 sets 2 Spring 10 6 sets 3 Washer 10 6 sets 1 Flat head screw...

Страница 17: ...liding Unit up then insert the Steel Rod to lower hole 6 Lift Sliding Unit set 2 counter weight and adjusting weight s in upper counter weight frame then attach the cover Opening Shaft Nut Nylon washe...

Страница 18: ...t 2 counter weight in lower counter weight frame then attach the cover 8 Attach pillar cover 1 Bind Screw M3 6 2 sets 2 Nylon washer 3 2 sets Pillar cover 1 Screw M4 10 2 sets Opening Counter Weights...

Страница 19: ...strike anchors Tighten M8 nuts Confirm that the apparatus is securely fixed When coach bolts diameter 8mm length 40mm are used Move the apparatus to the place where the Bel Cypher N is installed Tight...

Страница 20: ...e from the hole located at the bottom of the column to the grounding terminal in the X Ray room Connect Earth Wire to facility Earth Ground in accordance with National Electric Code and local codes Se...

Страница 21: ...holder with two men hook the Rotation Unit ASSY onto Sliding Unit Align rotation unit by using taper pins 11 2 Insert taper pins with a hammer 11 3 Tighten and Fix hexagon bolts Hook Rotation Unit AS...

Страница 22: ...Bel Cypher N 12 Remove Carriage Holder from Rotation unit Assy Take the carriage holder off the rotation unit assembly 1 Nut 8 4sets 2 Spring Washer 8 4sets 3 Cap screw M6 4sets...

Страница 23: ...Bel Cypher N 13 Connect the wire harness from Sliding Unit to the PC board of Rotation Unit...

Страница 24: ...Bel Cypher N 14 Connect the wire harness from Rotation Unit to wire harness from Sliding Unit...

Страница 25: ...ll Chinrest unit to sliding unit 15 1 Align positioning pins of Chinrest Unit ASSY to positioning holes of Sliding Unit ASSY 15 2 Attach Chinrest Unit by cap screw M5 20 4 sets Spring washer 5 4 sets...

Страница 26: ...unit cover Left side Put the sliding unit cover from front Slowly rotate arm to 90 degrees position Turn the ball screw in order to bring rota tion unit to forward end 1 Bind Screw M3 16 2 sets 2 Bin...

Страница 27: ...handle assy insert handle to detent pin fix handle by cap screw Connect CNSM Fix by a cap screw M6 40 18 Put the rotating unit cover on rotating Unit Put Frankfurt plane knob Attaches cover Fix with...

Страница 28: ...Bel Cypher N 19 Attach the Upper Cover on the Pillar 20 Insert Chinrest and Head Holding Rods or Ear Rods to Chinrest ASSY Pillar cover Hole Spiral code Chinrest Unit Head Holding Rods...

Страница 29: ...tall a LAN Card for Panoramic radiograph 1 Change User Accounts and Safety Rem If Operating System is Windows XP skip this section 1 Start Control Panel click User Accounts and Family Safety 2 Choose...

Страница 30: ...the bar to Never notify then click OK 5 After user account control setting is changed O S requests to reboot PC If you need to add an expanded LAN card go to next section install LAN card without rebo...

Страница 31: ...PC power plug into the outlet and turn the power of PC ON 5 After starting up of the Windows the driver is installed automatically If the driver is not installed automatically follow instructions desc...

Страница 32: ...er N 8 Install wizard comes up Click Next 9 License agreement screen comes up choose I accept the terms in the license agreement then click Next 10 Setup Options screen comes up do not make change cli...

Страница 33: ...en comes up click Install 12 When below screen is shown click Continue Anyway 13 After installation is completed click Finish 14 Connect LAN cable of Bel Cypher N to the PC Make sure to connect LAN ca...

Страница 34: ...Cypher N 9 Methods to install a TWAIN driver in imaging software 1 Set IP address 1 Open Control Panel 2 Choose Network and Internet 3 Choose Network and Sharing Center 4 Choose Change adapter settin...

Страница 35: ...owing IP address 7 Set IP address and subnet mask as follows You need to set following values first You need to set following values first You need to set following values first You need to set follow...

Страница 36: ...o Bel Cypher N Then choose property If you use the NIC supplied with Bel Cypher right click Intel R Gigabit CT Desktop Adapter and choose property 3 Property screen of NIC comes up 4 Choose Advanced t...

Страница 37: ...Bel Cypher N 3 DCAM API Install 1 Run setup exe in Common DCAM API of Library CD 2 click Install 3 Click Next 4 Check I accept the terms of the license agreement then click Next...

Страница 38: ...elow screen is shown click Finish to finish installation 4 Install DirectX 1 Run DXSETUP exe in Common DXInst of Library CD 2 When below screen is shown check I accept the agreement then click Next 3...

Страница 39: ...ly then run Setup bat stored in a CD 2 Set Bel Cypher N TWAIN Driver of Install Disc 3 NET Framework 3 5 SP1 setup display such as Fig 2 1 starts automatically Click Accept Fig2 1 NET Framework 3 5 SP...

Страница 40: ...e Setup screen 5 The dialogue such like Fig 2 3 is displayed NET Framework 3 5 SP1 and Visual C Run time is installed Fig 2 3 Visual C Run time Install screen 6 After the installation was completed th...

Страница 41: ...splayed Click Next Fig 2 5 Select Installation Folder screen 8 The dialogue such like Fig 2 6 is displayed Click Next Fig 2 6 Confirm Installation screen 9 The dialogue such like Fig 2 7 is displayed...

Страница 42: ...Bel Cypher N 10 The dialogue such like Fig 2 8 is displayed Click Close Fig 2 8 Installation Complete screen 11 Installation is completed...

Страница 43: ...0 101 Subnet mask 255 255 0 0 Default IP address Subnet mask for PC NIC IP address 169 254 87 101 Subnet mask 255 255 0 0 6 1 Change IP address of sensor Set up for PC and set up for Bel Cypher should...

Страница 44: ...Bel Cypher N 5 Click DeviceDiscovery 6 When IP address for sensor is displayed click Open 7 When Open Success is displayed then click IP Change 8 IP address change screen comes up...

Страница 45: ...close window by clicking on upper right corner to close IP_Address_Change screen 13 Click Close and Memory save 14 Click on upper right corner to close GigE_IP_Tool_ _ exe if OS is 64bit GigE_IP_Tool_...

Страница 46: ...program of Bel Cypher N 2 When above TWAIN screen is shown click L M or S then hit Ctrl Alt i simultaneously 3 IP address set up screen comes up 4 Enter new IP Adress and SubnetMask Set IP address an...

Страница 47: ...Bel Cypher N 5 Click Settings 6 When Turn on the the power of the X ray equipment again is displayed then turn off and on the power of Bel Cypher N...

Страница 48: ...ble click Internet Protocol Version 4 TCP IPv4 in The connection uses the following items 4 Enter IP address and Subnet mask Set IP address and Subnet Mask as NIC is included within sensor and Bel Cyp...

Страница 49: ...The lowest specifications of P C CPU Intel Core2 Duo E6850 3 00GHz 4MB 1333MHz FSB Chipset Intel X38 Express Video Controller NVIDIA Quadro FX380 256MB or equivalent Memory 4GB OS Windows XP Professio...

Страница 50: ...ktop Adapter We validated below three patterns Direct connection not through HUB Connect through 100M HUB Connect through Gigabit HUB NIC setting is done by auto negotiation Below HUBs are used for va...

Страница 51: ...Figure 1 Cover radiation aperture with lead Turn on the power of PC that is connected to Bel Cypher N Hit Ctrl Alt T simultaneously to get into Test mode Set exposure condition by referring the Opera...

Страница 52: ...ration Manual 2 2 Confirm the operation with X ray 2 2 1 Cover radiation aperture with lead 2 2 2 Set exposure condition by referring the Operation Manual 1 Exposure Orbit Panorama 5 Manual Exposure 6...

Страница 53: ...age Vac 2 Does 1 meet the rating description on the Controller plate OK NG 3 Rating Values of the Circuit Protector on the Rotation Unit Voltage V Current A 4 Are there any problems when the power plu...

Страница 54: ...with up down switch OK NG 4 Does up down operation of main body have a prob lem Allophone OK NG 5 Does Sliding Unit stop at the highest and the lowest position OK NG 6 Positioning Beams in Panorama 6...

Страница 55: ...A at Panoramic mode Allophone OK NG Vibration OK NG 10 Check in T M J 10 1 Does equipment work normally with exposure op eration at 0kV 0mA at TMJ mode Allophone OK NG Vibration OK NG 10 2 After the f...

Страница 56: ...KARA BELMONT USA Inc 101 Belmont Drive Somerset NJ 08873 Toll Free 800 223 1192 Toll Free Fax 800 280 7504 www belmontequip com TAKARA COMPANY CANADA LTD 2706 South Sheridan Way Mississauga Ontario Ca...

Страница 57: ...Bel Cypher N 12 Revision data of this manual This manual was created on Oct 2008 Revised July 2014 Document number B02 T170...