MD3150/3060
Copyright BEGLE GmbH, All rights reserved.
Page 10
Internet: www.bue.de
Mode E:
Calibration mode: Only for service
Mode display
Small display
Large display
E
10,000 digit of A/D value
1,000 digit of A/D value
The letter 'E' is shown in the mode display.
The
small
+
large displays
show the count value of the internal A/D converter. This
mode is only used for service work.
IMPORTANT:
If you enter calibration mode accidentally, hold the minus key pressed until the
software returns to the normal operating mode (Mode 0).
Maintenance, disposal
Please observe the following instructions :
•
Spent batteries are hazardous waste and must be disposed as prescribed by law.
•
Do not use carbon-containing chemicals or fuel, alcohol or similar fluids for cleaning.
••••
Damage due to non-compliance with the above instructions is not covered by
guarantee.
Troubleshooting and remedy
Problem
Possible remedy
The device does not display
temperature but only meaningless
characters
You are probably in calibration mode. Press the
minus key to return to normal operating mode.
Not all modes can be entered
An interval measurement has not been completed
and is still active; see description of mode 3
When the device is switched on it does
not go to temperature display mode
(MD3060)
No sensor connected or sensor defective