MD3150/3060
Copyright BEGLE GmbH, All rights reserved.
Page 9
Internet: www.bue.de
Mode 7:
Set trigger condition or automatic switch-off of the
device
Mode display
Small display
Large display
7
OFF,
---
,
---
/ On,OFF
t r i g / A o F F
Pressing the minus key alternates between entry of 'trigger conditions' and 'automatic
switch-off' of the device.
Setting trigger condition: The
large display
shows ‘t r i g’
The 3 available trigger conditions are selected with the plus key:
-
OFF
- top segment bar:
- bottom segment bar:
Interval measurement starts directly in mode 3.
Interval measurement starts if the temperature set
in mode 6 is
exceeded
.
Interval measurement starts if the temperature set
in mode 6 is
undershot
.
Setting the device for switch-off: The
large display
shows ‘A o F F’
The device switch-off function can be selected with the plus key.
- OFF
Automatic switch-off function is disabled
- On
Automatic switch-off function is enabled, i.e., if the device does not
detect activity within 4 minutes, it switches off automatically.
Mode 8:
Manual save
Mode display
Small display
Large display
8
Memory location no.
Current temperature
The current temperature and the memory location number are displayed.
To carry out a series of measurements in manual save mode, a memory reset to
´000´ must at first be made with the minus key. Each press of the plus key saves the
current temperature value at the indicated memory location (s
mall display
).
If the measured value memory is reset but no manual saving carried out, the previous
series is not deleted, it will be deleted when a manual save is carried out.
More values can be added to an available series of measured temperature values in
manual save mode. For this, display the last-saved temperature value in the series in
mode 4 and then change to mode 8.
Pressing the plus key saves the current temperature value at the next memory
location.
Only MD3150 :
The current temperature value corresponds to the measuring channel selected in
mode 0. If 2-channel measurement mode is selected in mode 0, wait at least 2
seconds between 2 measurements in manual mode.
Please note:
Values entered in manual mode are only saved permanently when mode 8 is exited
and mode 0 adopted.