3-7
Troubleshooting
Accessing the Centrifuge in Case of Power Failure
3
Figure 3.2
Removing the Emergency Access Plug
4
Insert the supplied 4 mm (P/N B33985) Allen wrench horizontally into the hole and turn it
counterclockwise until it stops. You should hear the door lock release. See
NOTE
If the door is opened via the emergency door release system during a centrifuge run, the
centrifuge will be switched off immediately and decelerate with no brake.
Figure 3.3
Manually releasing the door lock with the Allen wrench
5
Reinsert the emergency access plug.
1.
Emergency access plug and position of the opening for the
emergency door release
1.
Manual release of the door lock