B30162AA
3-5
Troubleshooting
Other Possible Problems
3
Other Possible Problems
Possible malfunctions that may not be indicated by diagnostic messages are described in
along with probable causes and corrective actions required. Possible causes for each problem are
listed in the probable order of occurrence. Perform the recommended corrective action in
sequence, as listed. If you are unable to correct the problem, call Beckman Coulter Field Service.
Table 3.2
Troubleshooting Chart
Problem
Possible Reason
Recommended Action
Display is blank
1.
No power to the mains supply.
2.
Power cord is not plugged in.
3.
Mains power is switched off.
4.
Door is not closed correctly.
1.
Check the fuse in the mains supply.
2.
Plug in the power cord correctly.
3.
Switch the mains power on.
4.
Close the door.
Centrifuge cannot be
started; The
displayed set speed
value cannot be
changed.
1.
Several .
2.
The door lock is not closed
correctly.
1.
Cycle power. If the error persists, contact Beckman
Coulter Field Service.
2.
Open and close the door. If the error persists,
contact Beckman Coulter Field Service.
Centrifuge
decelerates during
operation and
displays an error from
1 to 18 after powering
on.
Several
Cycle power. If the error occurs again, contact
Beckman Coulter Field Service.
Centrifuge
decelerates during
operation and
displays error 19 after
powering on.
Several .
Quit by pressing the door key.
Temperature value
cannot be reached
(applies to 20R
Refrigerated
centrifuges only).
Dirty condenser.
Contact Beckman Coulter Field Service.
Содержание Microfuge 20 Series
Страница 4: ...B30162AA iv Revision Status ...
Страница 8: ...B30162AA viii Safety Notice Chemical and Biological Safety ...
Страница 14: ...xiv Tables ...
Страница 48: ...B30162AA 2 18 Operation Saving and Using Programs ...
Страница 56: ...B30162AA 3 8 Troubleshooting Accessing the Centrifuge in Case of Power Failure ...
Страница 62: ...B30162AA 4 6 Care and Maintenance Supply List ...
Страница 68: ...B30162AA A 6 Installation Test Run ...
Страница 72: ...B30162AA Warranty 2 Beckman Coulter Inc Microfuge 20 Series Centrifuge Warranty ...