![Beckman Coulter Microfuge 20 Series Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/beckman-coulter/microfuge-20-series/microfuge-20-series_instructions-for-use-manual_2719742051.webp)
B30162AA
3-3
Troubleshooting
User Messages
3
Table 3.1
Error Message Chart
Error Number
Problem
Recommended Action
Explanation
1 through 9,
System error
After rotor comes to a complete
stop, turn the power off (
O
),
then back on (I) to reset.
All of these errors stop
the centrifuge or
cause it to decelerate
brakeless.
10 through 19,
Speedometer error
After rotor comes to a complete
stop, turn the power off (
O
),
then back on (I) to reset.
20 through 29
Motor error
Turn the power off (
O
). Ensure
ventilation.
30 through 39
EEPROM error
After rotor comes to a complete
stop, turn the power off (
O
),
then back on (I) to reset.
Errors 34, 35 and 36
will cause the
centrifuge to stop;
Errors 37 and 38 will
not cause the
centrifuge to stop,
these errors will only
return error messages.
40 through 45
Temperature error
(applies only to
Microfuge 20R
refrigerated
centrifuges)
•
Allow to slow down
•
Power off
•
Allow to cool down
•
Provide better ventilation
46 through 49
Imbalance error
•
Allow to slow down
•
Power off
•
Allow to cool down
50 through 59
Door error
•
Press door key
•
Close door
•
Remove foreign matter
from the opening of the
door lock device
Errors 50 and 51 will
cause the centrifuge
to stop.
60 through 69
Process error
•
Allow to slow down
•
Cycle power
“With error 60
message “power
failure during run”,
with error 61, the
message “stop after
power on”
70 through 79
Communication error
•
Allow to slow down
•
Cycle power
Содержание Microfuge 20 Series
Страница 4: ...B30162AA iv Revision Status ...
Страница 8: ...B30162AA viii Safety Notice Chemical and Biological Safety ...
Страница 14: ...xiv Tables ...
Страница 48: ...B30162AA 2 18 Operation Saving and Using Programs ...
Страница 56: ...B30162AA 3 8 Troubleshooting Accessing the Centrifuge in Case of Power Failure ...
Страница 62: ...B30162AA 4 6 Care and Maintenance Supply List ...
Страница 68: ...B30162AA A 6 Installation Test Run ...
Страница 72: ...B30162AA Warranty 2 Beckman Coulter Inc Microfuge 20 Series Centrifuge Warranty ...