PN B32164AC
16
JA-14.50 Fixed-Angle Rotor
Preparation and Use
Tubes for JA-14.50 Rotors Manufactured after October 2016
NOTE
JA-14.50 rotors and lids manufactured after October 2016 include the letter “T” after the serial number
(see
and
).
CAUTION
The tubes listed in Table 3 should be used one time only. Tube damage and sample
leakage will result from repeated use.
Do not use glass tubes in this rotor.
Table 3
Adapters to support tubes from other manufacturers for the JA-14.50 Rotor
a
manufactured after October 2016
Type and
Volume
Tube Description
Required
Adapter
Effective
r
max
(mm)
Effective
r
min
(mm)
Max
Speed/RCF/
k
Factor
conical
50 mL
Polypropylene conical tube with cap
bcdef
N/A
159.7
98.8
g
conical
15 mL
Polypropylene conical tube with cap
bce
B89429
(Qty 2 ea
B19962)
156.2
92.2
conical
5 mL
Polypropylene conical tube with tethered
cap
h
B89431
(Qty 2 ea
B19964)
125.1
95.7
conical
50 mL
Spin filtration device
ij
N/A
159.7
98.8
conical
15 mL
Spin filtration device
B89429
(Qty 2 ea.
B19962)
156.2
92.2
a. Use only the items listed here and observe fill volumes and maximum run speeds.
b. Manufactured by Greiner.
c. Manufactured by BD Falcon.
d. Manufactured by VWR.
e. Manufactured by Corning.
f. Manufactured by NUNC.
g. Use manufactures' recommendations for tube and bottle RCF limits. The corresponding RPM and k factor can be calculated using the equations
on pages 1 and 15, respectively.
h. Manufactured by Eppendorf.
i. Manufactured by Pall.
j. Manufactured by Millipore.