PN B32164AC
12
JA-14.50 Fixed-Angle Rotor
Preparation and Use
If the Rotor Sticks to the Drive Hub
If the rotor sticks to the centrifuge drive hub, use the removal tool (360998) to release it as follows
(see
.
Figure 7
Using the Rotor Removal Tool
NOTE
This procedure requires you to remove the rotor lid while the rotor is in the centrifuge. If you are using
the dual-locking lid, the biosafe seal provided by this mechanism will be broken when the lid is removed.
Consult your laboratory safety officer for the proper safety procedures to follow in this case.
1
Remove the lid.
2
Screw the removal tool into the threaded opening.
As the tool tightens down it will push against the drive hub, forcing the rotor up and off the hub.
3
Lift the rotor up and out of the centrifuge by the removal tool handle or by holding the rotor
with your hands.
4
Unscrew the removal tool and remove it from the rotor.
NOTE
Never run the rotor with the removal tool in place.
Lubricate the centrifuge drive hub with a light coat of Spinkote to prevent further sticking.
1.
Rotor Removal Tool (360998)