PN B32164AC
8
JA-14.50 Fixed-Angle Rotor
Preparation and Use
Figure 6
Installing the Rotor
Rotor Lid Configurations
The rotor has two available lid configurations.
•
The dual-locking lid (B19947) allows the lid to be attached to and removed from the rotor
outside the centrifuge, so that tubes can be loaded and unloaded under a safety hood. The dual-
locking lid consists of a daisy knob that attaches the lid to the rotor body, and a tie-down knob
that attaches the rotor to the centrifuge drive hub.
1.
Top View
2.
Side View
1.
Tie-down Knob
2.
Daisy Knob
3.
Lubricate Threads
4.
Dual-Locking Lid
1
4
3
2