JA-14.50 Fixed-Angle Rotor
Description
Description
Beckman Coulter JA-14.50 rotors are manufactured in a facility that maintains certifications to both
ISO 9001:2008 and ISO 13485:2003. They are for use with the specified Beckman Coulter centrifuges.
The JA-14.50 fixed-angle rotor is designed to hold up to sixteen 50-mL tubes at a 35-degree angle
(see
). Used in Beckman Coulter Avanti J series centrifuges, the JA-14.50 rotor develops
centrifugal forces that can efficiently pellet cells, bacteria, and food products; separate proteins,
viruses, and subcellular fractions, and do phase separation and binding studies.
The JA-14.50 rotor with dual-locking lid was tested
*
to demonstrate containment of microbiological
aerosols under normal operating conditions of the associated Beckman Coulter centrifuge, when
used and maintained as instructed. The rotor is also available with a single-locking lid.
Figure 1
The JA-14.50 Fixed Angle Rotor
The JA-14.50 rotor has a patented fluid-containment annulus, located below the O-ring sealing
surface (see
). The annulus helps to retain liquid that may escape from leaking or overfilled
tubes, thereby minimizing the escape of liquid into the instrument chamber during centrifugation.
1.
Dual-Locking Rotor Lid
2.
Small O-ring
3.
Large O-ring
4.
Rotor Body
* Validation of microbiological containment was done at an independent third-party testing facility. Improper use or
maintenance may affect seal integrity and thus containment.
1.
Single-Locking Lid
2.
Rotor Body
3.
O-Ring
4.
Dual-Locking Lid