PN 4277248DD
3-21
OPERATION PRINCIPLES
XB ANALYSIS
3
See the Glossary for terms used with the
XB Analysis.
XM Analysis
XM Analysis is a quality-control method that uses an Exponentially Weighted Moving
Average (EWMA) of CBC, Diff, NRBC and Reticulocyte Parameters and compares them with
with known target values, to monitor instrument performance. The first form of moving
average statistical analysis in hematology was XB Analysis.
Extended QC
Extended QC Rules are derived from the German Quality Control Guidelines for the Medical
laboratory, known in Germany as Rili-BÄK. Rili-BÄK (Guidelines of the Federal Chamber of
Physicians), was first published in 1987 and amended in 1990 and 1993 covering clinical
chemistry, immunochemistry and other tests, but not hematology. In 2003, the guidelines
were extended to include hematology.
Users can enable/disable Extended QC Rules for 5C Cell control.
Table 3.1 Effect of Directly-Measured Parameters on the Red Cell Indices
Index
Directly-Measured Parameter
MCV
RBC
HGB
Increased
Decreased
Increased
Decreased
Increased
Decreased
MCV
HIGH
LOW
NORMAL
NORMAL
NORMAL
NORMAL
MCH
NORMAL
NORMAL
LOW
HIGH
HIGH
LOW
MCHC
LOW
HIGH
LOW
HIGH
HIGH
LOW
Содержание COULTER LH 750 System
Страница 26: ...PN 4277248DD 1 12 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Страница 54: ...PN 4277248DD 3 22 OPERATION PRINCIPLES XB ANALYSIS ...
Страница 84: ...PN 4277248DD 4 30 SPECIFICATIONS CHARACTERISTICS BAR CODE LABEL SPECIFICATIONS ...
Страница 90: ...PN 4277248DD 5 6 HAZARDS DISPOSAL OF ELECTRICAL INSTRUMENTATION ...
Страница 116: ...PN 4277248DD 6 INDEX INDEX X XB Analysis definition GLOSSARY 16 XM Analysis definition GLOSSARY 16 ...
Страница 118: ......