PN 4277248DD
4-21
SPECIFICATIONS/CHARACTERISTICS
BAR-CODE LABEL SPECIFICATIONS
4
Table 4.25 Code-Related Specifications
Checksum Algorithm
Beckman Coulter strongly recommends the use of bar-code checksums to provide automatic
checks for read accuracy.
The algorithm for determining the checksum for each code is as follows:
Interleaved 2-of-5
This code requires 3 to 15 data digits plus a checksum or 16 digits without checksum. The
number of printed characters must always be even in length. Therefore, printers will
automatically add a leading zero as a filler if an odd number of characters is chosen.
To determine the value of the checksum character:
1.
Identify even- and odd-positioned characters in the message with the right-hand message
character
always
defined as an even-positioned character.
2.
Sum the numeric values of the odd-positioned characters.
3.
Sum the numeric values of the even-positioned characters and multiply the total by 3.
4.
Sum the odd and even totals from steps 2 and 3.
5.
Determine the smallest number which, when added to the sum in step 4, results in a
multiple of 10.
Code Type
Analyzer
Readable
Digit Length
*
Narrow Element Width
Wide
Element/Narrow
Element Ratio
SlideMaker Label
Scannable Bar Code
Characters
*
Inches (Minimum)
µm (Minimum)
Interleaved 2-of-5
with Checksum
3 to 15
0.010
254.0
(2.21 to 3):1
11
Interleaved 2-of-5
without Checksum
4 to 16
0.010
254.0
(2.21 to 3):1
10
NW7 with
Checksum
3 to 11
0.010
254.0
3:1
8
12 to 16
0.007
177.8
3:1
Codabar with
Checksum
3 to 11
0.010
254.0
3:1
8
12 to 16
0.007
177.8
3:1
Codabar/NW7
without Checksum
4 to 11
0.010
254.0
3:1
9
12 to 16
0.007
177.8
3:1
Code 39 with
Checksum
3 to 9
0.010
254.0
(2.21 to 3):1
8
10 to 16
0.007
177.8
3:1
Code 39 without
Checksum
4 to 9
0.010
254.0
(2.21 to 3):1
9
10 to 16
0.007
177.8
3:1
Code 128: Character
Sets A & B
4 to 14
0.010
254.0
(2 to 4):1
9 or 11 with hardware
modification
15 to 16
0.007
177.8
(2 to 4):1
Code 128: Character
Set C
4 to 16
0.010
254.0
(2 to 4):1
10
*
Although the LH 700 Series analyzer with SlideMaker accepts up to 16-character bar codes, the SlideMaker only prints the human-readable
information on the slide for bar-code labels with 12 or more digits.
IMPORTANT
Use of bar codes is an extremely accurate and effective method of positive patient
identification. Certain features, such as checksum digits, maximize accuracy in reading Codabar, Code 39
and Interleaved 2-of-5 labels. Use checksums to provide protection against occasional misread errors
caused by problems such as damaged or misapplied labels. If you must use bar-codes without
checksums, Beckman Coulter recommends that you verify each bar-code reading to assure correct patient
identification.
Содержание COULTER LH 750 System
Страница 26: ...PN 4277248DD 1 12 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Страница 54: ...PN 4277248DD 3 22 OPERATION PRINCIPLES XB ANALYSIS ...
Страница 84: ...PN 4277248DD 4 30 SPECIFICATIONS CHARACTERISTICS BAR CODE LABEL SPECIFICATIONS ...
Страница 90: ...PN 4277248DD 5 6 HAZARDS DISPOSAL OF ELECTRICAL INSTRUMENTATION ...
Страница 116: ...PN 4277248DD 6 INDEX INDEX X XB Analysis definition GLOSSARY 16 XM Analysis definition GLOSSARY 16 ...
Страница 118: ......