PN 4277248DD
4-11
SPECIFICATIONS/CHARACTERISTICS
PERFORMANCE CHARACTERISTICS
4
Temperature
Ambient operating range: 15.5 to 32°C (60 to 90°F)
Humidity
0 to 95% without condensation
Storing Specimens
Venipuncture Specimens
For CBC/Diff:
r
Run within 24 hours after collection, if stored at 21.2 - 25.6° C (70-78° F).
r
Run within 48 hours after collection, if stored between 2 to 8°C (35.6 to 46.4°F).
For Reticulocytes:
r
Run within 24 hours after collection, if stored at 21.2 - 25.6° C (70-78° F).
r
Run within 72 hours after collection, if stored at 2 to 8°C (35.6 to 46.4°F).
NRBC Parameters:
r
Run within 24 hours after collection, if stored at 21.2 - 25.6° C (70-78° F).
r
Stored between 2 and 8°C (35.6 and 46.4°F), up to 24 hours after collection.
NOTE:
Also refer to any pertinent manufacturer’s storage recommendations for the
venipuncture and microcollection devices used in your lab.
RDW
%
13.30
12.09
15.19
Plt
x 10
3
cells/µL
248.65
169.06
397.10
MPV
fL
9.31
7.54
11.24
NE
%
59.78
41.71
73.74
LY
%
28.67
18.39
44.42
MO
%
7.89
4.69
12.66
EO
%
2.96
0.75
8.06
BA
%
0.62
0.17
1.37
NE
x 10
3
cells/µL
3.74
1.47
7.06
LY
x 10
3
cells/µL
1.84
0.65
4.25
MO
x 10
3
cells/µL
0.50
0.17
1.21
EO
x 10
3
cells/µL
0.20
0.03
0.77
BA
x 10
3
cells/µL
0.04
0.01
0.13
RET
%
1.18
0.45
2.28
RET
x 10
6
cells/µL
0.05
0.02
0.11
MRV
fl
114.15
102.73
124.89
IRF
Ratio
0.27
0.163
0.362
Table 4.23 Normal Population Study (Continued)
Содержание COULTER LH 750 System
Страница 26: ...PN 4277248DD 1 12 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Страница 54: ...PN 4277248DD 3 22 OPERATION PRINCIPLES XB ANALYSIS ...
Страница 84: ...PN 4277248DD 4 30 SPECIFICATIONS CHARACTERISTICS BAR CODE LABEL SPECIFICATIONS ...
Страница 90: ...PN 4277248DD 5 6 HAZARDS DISPOSAL OF ELECTRICAL INSTRUMENTATION ...
Страница 116: ...PN 4277248DD 6 INDEX INDEX X XB Analysis definition GLOSSARY 16 XM Analysis definition GLOSSARY 16 ...
Страница 118: ......