INSTALLATION
SPECIAL REQUIREMENTS: HARDWARE
Figure 2.2 Space for Small Form Factor (SFF)-type computers
Electrical Input
This instrument requires:
r
An independent protected circuit
r
A three-wire outlet furnishing the applicable line voltage, single-phase input power
r
A ground path capable of carrying the full current of the circuit (confirmed thirdwire
earth ground)
r
That the 3-m (10-ft) primary power cord on the rear of the Power Supply be plugged
directly into the electrical outlet. Do not use an extension cord.
Current-carrying capacity of 20 A is recommended, although the actual power consumption
is about 2080 W as shown in the table below:
33 cm (13 in.)
25 cm (10 in.)
CLEARANCE
60.3 cm (23.75 in.)
74.9 cm (29.5 in.)
101.6 cm (40 in.)
CLEARANCE
CLEARANCE
25 cm (10 in.)
27.9 cm (11 in.)
CLEARANCE
25.4 cm (10 in.)
38 cm (15 in.)
LH 780
30 cm (13 in.)
76.2 cm
(30 in.)
50.2 cm
(19.75 in.)
25.4 cm
(10 in.)
68.6 cm
(27 in.)
44.5 cm
(17.5 in.)
50.2 cm
(19.75 in.)
45.7 cm
(18 in.)
88.9 cm
(35 in.)
30 cm (12 in.)
CLEARANCE
61 cm
(24 in.)
66 cm
(26 in.)
38.1 cm
(15 in.)
63.5 cm
(25 in.)
disc
disc
DVD
RW
NEC
MultiSync LCD 1700m+
45.2 cm (17.8 in.)
64.5 cm
(25.4 in.)
64.8 cm
(25.5 in.)
CAUTION
Introduction of electrical interference causing the instrument to lock up or reset frequently can
occur if you do not plug the primary power cables directly into an electrical outlet. Overheating, melting and
burning of the power lines can occur if you use an extension cord with the primary power cables. Plug the
primary power cables directly into an electrical outlet. Place the instrument close enough to an electrical
outlet that an extension cord is not needed.
Instrument Components
Watts
Analyzer, Diluter, Power Supply
860
SlideMaker
430
SlideStainer
430
Computer
300
Содержание COULTER LH 750 System
Страница 26: ...PN 4277248DD 1 12 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Страница 54: ...PN 4277248DD 3 22 OPERATION PRINCIPLES XB ANALYSIS ...
Страница 84: ...PN 4277248DD 4 30 SPECIFICATIONS CHARACTERISTICS BAR CODE LABEL SPECIFICATIONS ...
Страница 90: ...PN 4277248DD 5 6 HAZARDS DISPOSAL OF ELECTRICAL INSTRUMENTATION ...
Страница 116: ...PN 4277248DD 6 INDEX INDEX X XB Analysis definition GLOSSARY 16 XM Analysis definition GLOSSARY 16 ...
Страница 118: ......