Operation
Manifold Specifications
All Lifeline Manifold systems are designed in
accordance with current revision of NFPA 99.
There are three categories of Lifeline manifolds for
liquid containers depending upon delivery
pressure.
Following gas types are available for each
delivery pressure:
55 PSI Delivery Pressure
• Oxygen
• Nitrous Oxide
• Carbon Dioxide
• Argon
100 PSI Delivery Pressure
• Oxygen
180 PSI Delivery Pressure
• Nitrogen
Refer to Table 1, page 27, for manifold
specifications.
Precautions
WARNING:
• Tampering with gas-specific connections shall
be prohibited. Do not alter, remove or modify
gas-specific connections.
• Before connecting liquid container to left bank
of manifold or high-pressure cylinder to right
bank of manifold and reserve header,
momentarily open and close
container/cylinder valve to blow out dirt and
debris.
• After connecting container/cylinder to
manifold, open container/cylinder valve
s-l-o-w-l-y to allow heat of compression to
dissipate.
• Always secure high-pressure cylinders with
racks, straps, or chains. Unrestrained
cylinders may fall over and damage or break
off cylinder valve.
• Do not repeatedly bend, sharply bend, or
twist copper pigtails as damage to tubing may
occur.
• Do not bend flexible pigtails into a radius
smaller than 3”.
• Service to be performed by qualified medical
equipment technician.