Testing the alarms
Before starting the plant, the alarms can be tested. The 4 alarms will automatically cycle for 3 seconds without
actual faults. This can be used to test the transmission to the control room alarm display. Navigate to the
following submenu and press the Enter button:
When the test icon is greyed out, the plant needs to be stopped first. Refer to the paragraph Stopping to stop
the ES-MED.
Starting the ES-MED
To start the system, after having completed the CAN and ES setup described above, carry out the following
steps:
1. Put the compressor's main switch to ON and the local/LAN switch to LAN. The emergency forced local
warning (see section
Interface icons and menu structure
) is present on the display and the red alarm LED
is on, since the pressure is lower than the preset emergency pressure level.
Once the pressure is above this level (on the display of the compressor controller), press the escape button
on the compressor controller. The warning disappears and the red LED extinguishes.
2. Then carry on to start the ES system (full access mode, see above).
Instruction book
2920 7099 70
29
Содержание dMED 100
Страница 2: ......
Страница 15: ...2 3 WSD and filters WSD and filters in the air flow WSD and filter overview Instruction book 2920 7099 70 13...
Страница 40: ...Inputs Instruction book 38 2920 7099 70...
Страница 45: ...Instruction book 2920 7099 70 43...
Страница 54: ...4 Installation 4 1 Dimension drawings Instruction book 52 2920 7099 70...
Страница 55: ...Instruction book 2920 7099 70 53...
Страница 56: ...Instruction book 54 2920 7099 70...
Страница 57: ...Instruction book 2920 7099 70 55...
Страница 58: ...Instruction book 56 2920 7099 70...
Страница 59: ...Instruction book 2920 7099 70 57...
Страница 60: ...Instruction book 58 2920 7099 70...
Страница 61: ...Instruction book 2920 7099 70 59...
Страница 62: ...References on drawings Instruction book 60 2920 7099 70...
Страница 93: ......
Страница 94: ...No 2920 7099 70 2012 01 Printed in Belgium www beaconmedaes com...