DD, PD
Filter with liquid trap, automatically
drained
12/ 15
Pressure regulator with gauge
PDp
Dust filter
13
Safety valve
1, 8
Flow restriction
18
Non-return valve
2, 9, 20
Pressure transducer
21
Pressure switch
3
Reservoir open to atmosphere
with inlet pipe above fluid level
QDT1, QDT2
QDT filter (basic version, only
activated carbon) / QDT+ filter
(activated CO catalyst
(add-on option))
4, 16, 19, 22, 23
Valve
CD1, CD 2
CD dryer
5
Silencer
IN
Inlet
6
Shuttle valve
OUT
Outlet
7
Desiccant
Description
Compressed air enters water trap (WSD), eliminating liquid water from the air. Next, the air flows through
DD and PD filters, removing contaminating particles and oil mist. The air then enters the CD dryer, eliminating
water damp and producing dry air with a pressure dew point of minimum -31 ˚C (-23.8 ˚F). Next is the QDT
filter where oil vapour and smells of the dried air are reduced (active carbon) and where the conversion of
CO into CO
2
takes place (catalyst in QDT+). Finally the air flows through the PDp filter, hereby eliminating
remaining dust particles for absolute purity.
Pressure switch (21) operates as an hardwired alarm in case a failure with the ES-MED controller should
occur. Should the ES-MED fail to operate the compressors, resulting in a decrease of the pressure, each
compressor will start to operate independently once their outlet pressure reaches a certain critical pressure
level (emergency forced local). The pressure switch, installed on the inlet of the dMED, thus linked to the
outlet pressure of each compressor will also be triggered at this pressure level, resulting in a Plant fault being
transmitted to the building management system.
Specifications of the outlet air can be found in section
Reference conditions and limitations
.
Instruction book
12
2920 7099 70
Содержание dMED 100
Страница 2: ......
Страница 15: ...2 3 WSD and filters WSD and filters in the air flow WSD and filter overview Instruction book 2920 7099 70 13...
Страница 40: ...Inputs Instruction book 38 2920 7099 70...
Страница 45: ...Instruction book 2920 7099 70 43...
Страница 54: ...4 Installation 4 1 Dimension drawings Instruction book 52 2920 7099 70...
Страница 55: ...Instruction book 2920 7099 70 53...
Страница 56: ...Instruction book 54 2920 7099 70...
Страница 57: ...Instruction book 2920 7099 70 55...
Страница 58: ...Instruction book 56 2920 7099 70...
Страница 59: ...Instruction book 2920 7099 70 57...
Страница 60: ...Instruction book 58 2920 7099 70...
Страница 61: ...Instruction book 2920 7099 70 59...
Страница 62: ...References on drawings Instruction book 60 2920 7099 70...
Страница 93: ......
Страница 94: ...No 2920 7099 70 2012 01 Printed in Belgium www beaconmedaes com...