9
Technical data
9.1
Fuse values
Attention
The indicated fuse value is the maximum value. The cable size used may specify fuses of
a lower value.
Values
Frequency
Voltage
Fuse specification
Maximum fuse value
IEC
50 Hz
230 V
gL/gG
2 A
CSA/UL
60 Hz
115 V
CSA HRC/UL RK5
2 A
CSA/UL
60 Hz
230 V
CSA HRC/UL RK5
2 A
9.2
Electric cable size
Attention
Local regulations remain applicable if they are stricter than the values proposed below.
The voltage drop must not exceed 5% of the nominal voltage. It may be necessary to use
cables of a larger size than those stated to comply with this requirement.
Frequency
Voltage
Minimum cable size
IEC
50 Hz
230 V
3 x 1.5 mm²
CSA/UL
60 Hz
115 V
3 x AWG14
CSA/UL
60 Hz
230 V
3 x AWG14
9.3
Device settings
Regulating and safety devices
The regulating and safety devices are factory-adjusted to obtain optimum performance of the dryer. Do not
alter the setting of any of the devices.
Instruction book
80
2920 7099 70
Содержание dMED 100
Страница 2: ......
Страница 15: ...2 3 WSD and filters WSD and filters in the air flow WSD and filter overview Instruction book 2920 7099 70 13...
Страница 40: ...Inputs Instruction book 38 2920 7099 70...
Страница 45: ...Instruction book 2920 7099 70 43...
Страница 54: ...4 Installation 4 1 Dimension drawings Instruction book 52 2920 7099 70...
Страница 55: ...Instruction book 2920 7099 70 53...
Страница 56: ...Instruction book 54 2920 7099 70...
Страница 57: ...Instruction book 2920 7099 70 55...
Страница 58: ...Instruction book 56 2920 7099 70...
Страница 59: ...Instruction book 2920 7099 70 57...
Страница 60: ...Instruction book 58 2920 7099 70...
Страница 61: ...Instruction book 2920 7099 70 59...
Страница 62: ...References on drawings Instruction book 60 2920 7099 70...
Страница 93: ......
Страница 94: ...No 2920 7099 70 2012 01 Printed in Belgium www beaconmedaes com...