20
interruptions
and voltage
variations on
power supply
input lines
iEc 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Not
applicable
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the TH30F
requires continued
operation during power
mains interruptions, it is
recommended that the
TH30F be powered from
an uninterruptible power
supply or a battery.
power
frequency
(50/60 hz)
magnetic field
iEc 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
NOTE : UT is the a.c. mains voltage prior to application of the test level.
rEcoMMEnDED sEpArAtion DistAncEs bEtWEEn portAbLE AnD MobiLE rF
coMMUnicAtions EqUipMEnt AnD thE ME EqUipMEnt or ME sYstEM
The TH30F is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the TH30F can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the TH30F as recommended
below, according to the maximum output power of the communications equipment.
Notice Mini therm 90x120 v2.indd 20
24/07/13 10:07