19
immunity
test
iEc 60601
test level
compliance
level
Electromagnetic environment –
guidance
conducted rF
iEc 61000-4-6
3 V rms
150 kHz
to 80 MHz
Not
applicable
Portable and mobile RF communications
equipment should be used no closer to
any part of the TH30F, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
recommended separation distance
d = 1,2 √P
d = 1,2 √P 80 MHz to 800 MHz
d = 2,3 √P 800 MHz to 2,5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey, a should be less than the
compliance level in each frequency range
b.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
rF rayonnée
iEc 61000-4-3
3 V/m
80 MHz
à 2,5 GHz
3 V/m
notE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
notE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the
location in which the TH30F is used exceeds the applicable RF compliance level above,
the TH30F should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the
TH30F.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
The TH30F is intended for use in the electromagnetic environment specified below.
The customer or the user of the TH30F should assure that it is used in such an environment.
immunity test
iEc 60601 test level
compliance
level
Electromagnetic
environment guidance
Electrostatic
discharge
(EsD)
iEc 61000-4-2
6 kV contact
8 kV air
6 kV
contact
8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast
transient/burst
iEc 61000-4-4
2 kV for power supply lines
1 kV for input/output lines
Not
applicable
Mains power quality
should be that of a typical
commercial or hospital
environment.
surge
iEc 61000-4-5
1 kV line(s) to line(s)
2 kV line(s) to earth
Not
applicable
Mains power quality
should be that of a typical
commercial or hospital
environment.