External Loopback

The Communications menu option that performs an
external communications loopback test. This test
verifies the internal communications through the serial
ports and external cables.

File 13

A storage file containing QC data that
Bayer Diagnostics field service engineers use during
troubleshooting. You can delete the contents of the file,
but you cannot access the file data.

File 14

A temporary storage file for discarded QC results. The
system does not place data into an active QC file or
update the statistics. You can recall the data in this file
using the Data Recall function if it was accidentally
discarded.

fixed time
calibrations

A method of determining intervals between calibrations
in which you determine the maximum time between
one-point and two-point calibrations. You determine
fixed time calibration through the Calibration Setup
menu.

F-keys

The five function keys (F1 through F5) located below
the screen that are associated with the labels displayed
on the bottom line of the screen. The key functions can
change as you move from screen to screen.

flag

A symbol that appears on some screens and printed
reports that indicates that the system detects an
unexpected result or measurement. For example, the
single up arrow (

) on a patient report indicates that the

sample result is above the reference range or that the
calibration result is above the high drift limit.

flexible time
calibrations

A method of determining calibration intervals. The
system uses an algorithm to determine the time between
calibrations based on sensor status and previous
calibration results (calibration drift).

Fluid Detector

The Fluidics System menu option that lets you test fluid
detectors 1, 1A, and 2 with air, clear liquid, or opaque
liquid (blood) and test fluid detectors 3 and 4 with air or
clear liquid.

fluid detector

A sensor that detects fluid in the sample path. There are
five fluid detectors: 1, 1A, 2, 3, and 4.

Fluidics System

The Troubleshooting menu option that lets you test the
fluidic system.

Содержание Rapidlab 800

Страница 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...

Страница 2: ...re trademarks of Bayer Corporation APV is a servicemark of Bayer Corporation Cidex is a trademark of Johnson Johnson Corporation IBM and AT are trademarks of International Business Machines Corporatio...

Страница 3: ...3 20 20 0 20 3 2 2 5 5 3 5 3 5 5 5 5 5 3 5 5 0 0 1 3 5 34 3 5 34 3 5 34 5 34 5 2 3 5 34 3 5 34 5 34 2 5 34 1 5 2 5 34 50 5...

Страница 4: ...5 5 0 0 5 5 34 5 34 5 0 5 5 5 0 5 5 5 0 5 5 3 5 2 3 5 3 5 3 5 3 5 0 3 5 0 5 3 5 5 0 0 5 3 5 0 1 5 3 5 0 5 5...

Страница 5: ...0 1 02 1 2 8 3 01 2 1 0 1 8 5 1 8 0 1 1 8 5 1 8 601 2 0 1 8 11 1 601 8 1 0 8 1 1 0 8 2 51 3 0 8 3 12 1 8 1 0 8 1 6 21 4 8 5 1 1 2 0 7 0 8 5 1 3 2 0 8 5 0 8 5 1 1 51 3 0 8 5 1 01 0 8 5 1 10 8 5 1 201 1...

Страница 6: ...0 0 4 4 3 0 0 4 4 3 0 4 4 3 0 0 4 1 4 4 3 2 4 4 3 4 4 3 4 4 3 4 4 3 2 4 4 3 4 4 3 2 4 4 3 0 4 4 3 2 4 4 3 2 4 4 3 4 4 3 4 4 3 2 4 4 3 4 4 3 4 4 3 2 4 4 3 2 4 4 3 4 4 3 0 4 4 3 1 4 4 3 2 4 4 3 4 4 3 2...

Страница 7: ...0 13 2 0 30 9 02 30 9 2 2 12 3 07 02 30 9 4 7 1 02 30 9 0 2 7 3 0 30 9 0 2 3 0 30 9 2 0 2 0 30 9 0 2 2 0 0 0 30 9 0 2 2 0 13 2 0 30 9 5 2 0 2 02 30 9 2 3 0 30 9 2 0 30 9 21 4 1 0 30 9 0 2 1 3 2 0 30...

Страница 8: ...8 1 9 5 2 2 9 2 9 8 1 1 2 2 9 2 9 8 1 1 1 1 9 2 9 8 3 1 01 11 9 2 9 8 3 1 01 21 1 3 9 2 9 8 3 1 0 1 02 1 2 9 2 9 8 1 0 2 0 9 2 9 8 3 1 2 12 0 9 2 9 8 1 2 12 0 9 2 9 8 3 9 2 9 8 3 1 11 9 2 9 8 3 1 1 9...

Страница 9: ...6 0 6 5 1 6 0 6 5 6 0 6 5 1 1 2 3 6 0 6 5 1 6 0 6 5 1 0 6 0 6 5 1 6 0 6 5 0 6 0 6 5 1 0 6 0 6 5 1 0 6 0 6 5 0 6 0 6 5 1 4 0 6 0 6 5 6 0 6 5 1 0 6 0 6 5 1 0 6 0 6 5 6 0 6 5 0 6 0 6 5 6 0 6 5 4 6 0 6 5...

Страница 10: ...13 63 17 0 4536 5 104 31 2 3 0 13 2 3160 2 3 563 0413 10 0 63 17 0 4536 5 104 31 5 3 5 3 10 0 63 17 0 4536 5 104 31 5 23 0 1 5 10 0 63 19 2 0 45 6 0 63 4 100 5 0 5 19 2 5 63 45 0 13 5 10 3 0 13 0 63 5...

Страница 11: ...7 1 7 5 4 0 0 7 1 7 5 4 0 0 7 1 7 5 4 0 0 7 1 7 5 4 0 0 7 1 7 5 0 0 7 1 7 5 01 7 1 7 5 01 7 1 7 5 01 7 1 7 5 01 7 1 7 5 00 7 1 7 5 00 7 1 7 5 4 0 0 7 1 7 5 01 0 7 1 7 5 0 0 0 7 1 7 5 0 1 7 1 7 5 0 2 7...

Страница 12: ...45 6 0 6 45 6 0 6 45 6 0 6 45 6 0 6 45 6 0 6 45 2 6 0 6 45 2 6 0 6 45 6 0 6 45 3 0 6 0 6 45 0 6 0 6 45 6 0 6 45 1 6 0 6 45 6 0 6 45 6 1 6 0 6 45 6 0 6 45 6 0 6 45 6 0 6 45 6 0 6 45 6 0 6 45 3 6 0 6 4...

Страница 13: ......

Страница 14: ......

Страница 15: ...perform routine maintenance and troubleshooting respiratory therapists who use an 800 system to routinely analyze blood gas samples supervisors who customize their 800 system to meet their laboratory...

Страница 16: ...About the System analyze samples QC and calibrators Section 2 Operating the System perform system maintenance Section 3 Maintaining the System identify the appropriate corrective action to resolve ope...

Страница 17: ...install the system Appendix H Installation review the analytic principles of the sensors and the calculations performed by the system Appendix I Operating Principles review the technical bulletin abo...

Страница 18: ...damage or loss of data Biohazard statements alert you to potentially biohazardous conditions The note symbol is used with important information that requires your attention This symbol indicates that...

Страница 19: ...risk of exposure to potential electrical hazards This symbol indicates that the input electricity is alternating current This symbol alerts you to important information about the fuses This symbol id...

Страница 20: ......

Страница 21: ......

Страница 22: ......

Страница 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...

Страница 24: ...4 0 0 4 1 0 4 4 4 1 4 4 4 4 4 0 4 0 4 0 4 4 4 0 4 4 2 4 0 2 4 4 0 3 4 0 3 4 4 3 0 3 4 0 0 3 4 0 0 3 4 4 1 34 4...

Страница 25: ......

Страница 26: ......

Страница 27: ...actate pH Na K Ca Cl tHb FO2Hb 845 855 865 pO2 pCO2 pH tHb FO2Hb FHHb FMetHb FCOHb pO2 pCO2 pH Na K Ca Cl tHb FO2Hb FHHb FMetHb FCOHb pO2 pCO2 glucose lactate pH Na K Ca Cl tHb FO2Hb FHHb FMetHb FCOHb...

Страница 28: ...bin ctO2 Hb oxygen capacity of hemoglobin O2CAP partial pressure of oxygen at 50 saturation p50 sulfhemoglobin concentrations greater than 1 5 arterial oxygen content ctO2 a venous oxygen content ctO2...

Страница 29: ...line assistance using an integrated Help program optional bar coding capabilities for patient ID and accession numbers managing patient results and QC data flags for results that fall outside expected...

Страница 30: ...fety automatic reagent path cleaning cycle expanding system capabilities advanced on board data management bidirectional communication with your information systems and Chiron Diagnostics data managem...

Страница 31: ...onents of the 800 system The 860 864 and 865 systems have five reagent bottles and the 840 844 845 850 854 and 855 systems have four reagent bottles Screen Keypad Measurement Module Window Sample Entr...

Страница 32: ...nt Pump Sample Entry Reagent Bottles Keypad Screen Measurement Module Door Latch Sample Pump Waste Pump Measurement Module Gas Tubing Waste Bottle Measurement Module Door Latch Analyze Key Reagent Man...

Страница 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...

Страница 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...

Страница 35: ...Contrast Knob Lamp Cover Keyboard Port Parallel Port Paper Spool Bar Code Scanner Port Serial Port 3 Serial Port 1 Serial Port 2 Network Port Air Filter Cover Base Model Interface...

Страница 36: ...F2 F1 F3 F4 F5 1 2 3 4 5 6 7 8 9 0 Enter Home Help CIBA CORNING Contrast Knob Arrow Keys F keys Help Key Home Key Enter Key Numeric Keys Screen Paper Spool Key Paper Advance Key Printer...

Страница 37: ...sis Refer to Appendix I Operating Principles for more information about each sensor Sample Ground Temperature Sensor The sample ground temperature sensor provides an electrical grounding mechanism for...

Страница 38: ...Measurement Module Door Measurement Module Door Latch Spring Loaded Latch Measurement Module Door Latch Release Pin Contact Assembly pH Sensor Reference Sensor Spring loaded Latch Measurement Module T...

Страница 39: ...omponents spectrophotometrically measure tHb and its derivatives Most of the measurement components are enclosed in the CO ox module The sample chamber is visible when the CO ox cover is raised The la...

Страница 40: ...nents direct the fluidic components in response to the user interface and automatic system activities The CO ox module uses the fluidic components of the base model and additional components The sampl...

Страница 41: ...865 systems Gas Tubing Reagent Manifold Gas Fittings Reagents The system pumps are the reagent pump the sample pump the waste pump and the CO ox pump if the CO ox module is attached Each pump consist...

Страница 42: ...O ox sample chamber sense the presence of a sample in the CO ox module The waste components include the waste bottle and latch the waste outlets the waste outlet cover and the waste detector The waste...

Страница 43: ...nd then provides power to the rest of the system Real Time Processor RTP Sends commands to other PC boards to perform system activities the RTP controls the fluidic system acquires data from the measu...

Страница 44: ...nt of energy generated by the optics and provides a coarse alignment of the diode array with the polychromator CO ox Measurement Board CMB Provides control signals to and collects and limits the signa...

Страница 45: ...hat you use to direct system activities and obtain operating status information F2 F1 F3 F4 F5 1 2 3 4 5 6 7 8 9 0 Enter Home Help CIBA CORNING Contrast Knob Arrow Keys F keys Help Key Home Key Enter...

Страница 46: ...the direction of the selected arrow Enter accepts selected options and moves from field to field on the Ready screen it displays the Reagent Levels screen Help Help accesses the Help screen Home Home...

Страница 47: ...the labels can vary with each screen You can use an IBM AT compatible 101 key alpha numeric keyboard to operate an 800 system Use the following keys on the keyboard to simulate the system keys on the...

Страница 48: ...results to a connected LIS HIS or data management system This key appears on the Results screen only if Auto Send is off Do Not Send prevent the transmission of sample results to a connected LIS HIS o...

Страница 49: ...Screen elements are the components on the screen that enable you to interact with the system software...

Страница 50: ...a green box with black text turned on and in calibration a yellow box with black text and a slash across the label turned on but out of calibration not visible turned off or not installed The Analyze...

Страница 51: ...arrow keys to select the menu option you want If dots appear after the option another menu exists for that option If no dots appear after the option the screen associated with the option appears when...

Страница 52: ...0 2 0 9 0 0 9 2 9 0 2 23 9 23 9 2 2 2 9 02 0 21 9 21 3 1 9 0 2 0 1 9 0 2 23 9 0 2 2 1 9 00 2 9 0 2 23 02 9 1 9 22 1 9 2 1 9 2 207 0 2 9 0 2 21 9 2 04 1 9 1 3 1 9 712 23 9 712 2 2 1 9 2 7 9 1 2 2 1 9 3...

Страница 53: ...on the numeric keypad you press to use menu codes NOTE During some procedures you may need to move to the Main Menu before you can press the menu code The menu code for these procedures contains the p...

Страница 54: ...re you press Help In the Menu mode press Help to display a scroll list of Menu mode topics Menu mode Help consists of an index of topics Each topic describes a menu or menu selection The topics do not...

Страница 55: ...the appropriate sample handling documents are in place Use sterile technique at all times to avoid infecting the sample site The following indicate some of the procedural limitations encountered with...

Страница 56: ...n analyze samples obtained from the following sources arterial blood Arterial blood is commonly recommended for use in blood gas studies because it accurately reflects acid base physiology and oxygena...

Страница 57: ...ed correctly arterial venous mixed venous and capillary blood samples are also suitable for analyte determinations For further information refer to the Bayer Diagnostics brochure Specimen Collection f...

Страница 58: ...pping Fill the capillary tubes completely and mix the samples thoroughly Collect venous blood using vacuum tubes containing lithium heparin Fill the tubes completely and mix the sample by gently inver...

Страница 59: ...g the sample in contact with the ice water slurry You can store samples in an ice water slurry for up to 2 hours without significant change in values for pH and pCO2 however this will affect the K val...

Страница 60: ...le You can autoclave the waste bottle before discarding it but you cannot use it after autoclaving After autoclaving seal the waste bottle with the disposable cap and then discard the waste bottle Dis...

Страница 61: ...ing occur Reagents other than those recommended are used Reagents are used after the expiration date on the label Reagents are not used or stored according to Bayer Diagnostics recommendations Standar...

Страница 62: ...eaning compound Cleaning Solution 2 cleaning compound Conditioner ammonium bifluoride NH4HF2 CO oximeter Slope dyes and viscosity adjuster Deproteinizer 0 1N hydrochloric acid proteolytic enzyme Mater...

Страница 63: ...e slope point two point metabolite calibration Cleaning Solution 1 Cleaning Solution 1 when alternated with Cleaning Solution 2 provides thorough cleaning of the reagent path Cleaning Solution 2 Clean...

Страница 64: ...L Zero Store Wash G L Zero upright at 4 to 25 C Discard Wash G L Zero 60 days after opening Cal G L Store Cal G L upright at 4 to 25 C Discard Cal G L 30 days after opening Cleaning Solution 1 Store...

Страница 65: ...1 Cleaning Solution 1 is ready to use Invert the bottle several times to mix thoroughly before use Cleaning Solution 2 Cleaning Solution 2 is ready to use Invert the bottle several times to mix thorou...

Страница 66: ...observe the following precautions when handling the tanks Secure tanks to a wall a bench or the floor or place them in a tank base support stand Use calibration gases only for the calibration of clini...

Страница 67: ...ing measurement to the established calibration curve The 800 system performs calibrations and sample measurements of pH blood gases electrolytes and metabolites at 37 0 0 15 C and hemoglobin and its d...

Страница 68: ...the required calibration automatically The system does not accept samples until the required calibration finishes In addition to the regularly scheduled one and two point calibrations the 860 system h...

Страница 69: ...s manually from the Analyze mode when the system is not analyzing samples and from the Menu mode You perform calibrations from the Analyze mode by pressing Calibrate You perform calibrations in the Me...

Страница 70: ...not in calibration Until a successful calibration completes all subsequent measurements indicate the not in calibration condition on the screen and on the printed reports The system automatically rep...

Страница 71: ...You can view print and transmit stored calibration data using the Recall option You cannot edit calibration data You can search for the data by calibration type or calibration date and time...

Страница 72: ...s U S federal regulations state that each laboratory must establish QC procedures to document and to evaluate system performance that ensure the accuracy and the reliability of patient results and rep...

Страница 73: ...lets you compare your system performance with systems in other laboratories Participation in interlaboratory QC survey and proficiency testing programs can identify systematic errors not detected by...

Страница 74: ...ials are not expired or have not deteriorated Visible signs of deterioration include color changes or cloudiness of the reagents or QC materials Ensure that you followed the operating procedures recom...

Страница 75: ...ation Sequence numbers run from 100 through 99999 and then repeat If the auto ID option is on the 800 system automatically identifies the QC sample results and selects the appropriate QC file to store...

Страница 76: ...er to Archiving QC Data and Viewing Archived QC Data in Section 5 The 800 system provides Levey Jennings charts which are visual representations of QC performance Levey Jennings charts allow rapid det...

Страница 77: ...hived QC Data in Section 5 Archiving QC data saves the previous month s QC file results and statistics on a diskette You can use this diskette to retrieve historical QC data which you can view or use...

Страница 78: ...l values are not adversely affected when you perform major preventive maintenance or replace critical components on your system when controls reflect unusual trends or are consistently outside your la...

Страница 79: ...method Typically linear regression and least squares evaluation are used to compare results The slope and intercept of the regression line can be used to match the results of the new system to an exis...

Страница 80: ...gure 1 12 describes the conventions used in the diagrams 7 3 6 8 Wash Clean Cal G L Cal Slope Ambient Air Measurement Module Sample Pump S Sample Port Reagent Pump R CO ox Pump CO Waste Pump W Sample...

Страница 81: ...e and sample pumps start and the sample moves to fluid detector FD 1 and 1A The CO ox module pump moves part of the sample to FD5 Cal Slope Ambient Air 1 1a 2 4 3 R S W 7 3 6 8 Wash Clean Cal G L 5 CO...

Страница 82: ...easurement is being made The foam wash and vent valves open the reagent pump starts and foam wash moves to the outer sample port The reagent pump stops the waste pump starts and the foam wash is remov...

Страница 83: ...the CO ox module is attached wash segments are split at the sample connector so the CO ox sample path is thoroughly cleaned This sequence repeats 11 times to create wash segments that thoroughly clea...

Страница 84: ...ple port The reagent pump stops the waste pump starts and the foam wash is emptied from the sample port This sequence repeats four times and is followed by a solid wash segment 7 3 6 8 Wash Clean Cal...

Страница 85: ...into the measurement module The reagent pump introduces another solid wash segment into the sample port This sequence repeats 11 times to create wash segments that thoroughly clean the measurement mod...

Страница 86: ...he wash valve opens and a wash segment is introduced into the reagent manifold The sample and waste pumps start The wash valve closes The vent valve opens Cal Slope Ambient Air 1 1a 2 4 3 R S W 7 3 6...

Страница 87: ...contamination of the sample connector by ambient air Cal Slope Ambient Air 1 1a 2 4 3 R S W 7 3 6 8 Wash Clean 5 CO H SC 4 After the gases are measured the 7 3 valve wash bypass valve and diverter val...

Страница 88: ...moves until the trailing edge is at fluid detector 1 The pumps stop and the analytes are measured The CO ox zero calibration is performed Cal Slope Ambient Air 1 1a 2 4 3 R S W 7 3 6 8 Wash Clean 5 CO...

Страница 89: ...reagent pump stops the waste pump remains on and the sample pump moves the wash segment from the sample port into the measurement module An air segment follows The wash valve opens and the reagent pum...

Страница 90: ...he wash valve opens and a wash segment is introduced into the reagent manifold The sample and waste pumps start The wash valve closes The vent valve opens Cal Slope Ambient Air 1 1a 2 4 3 R S W 7 3 6...

Страница 91: ...5 CO H SC 4 The Slope gas valve opens and Slope gas flows into the inner sample port The sample pump aspirates the Slope gas at a controlled rate through the measurement module where pO2 and pCO2 are...

Страница 92: ...es toward the sample port When the main segment reaches the sample port the sample pump starts and moves the segment to the measurement module until it is detected by fluid detector 1 and 1A The CO ox...

Страница 93: ...to remove any residual 7 3 reagent from the pathways Cal Slope Ambient Air 1 1a 2 4 3 R S W 7 3 6 8 Wash Clean 5 CO H SC 10 The pre segments move through the sample port into the waste Cal Slope Ambie...

Страница 94: ...wash The reagent pump starts and moves foam wash to the inner sample port The reagent pump stops the waste pump starts and the foam wash is emptied from the sample port This sequence repeats four tim...

Страница 95: ...e reagent pump introduces another solid wash segment into the sample port This sequence repeats 11 times to create wash segments that thoroughly clean the measurement module Cal Slope Ambient Air 1 1a...

Страница 96: ......

Страница 97: ......

Страница 98: ......

Страница 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...

Страница 100: ......

Страница 101: ...to instrument and operator to operator variability The system automatically identifies the sampling device If the system cannot identify the device a message box appears prompting you to select one of...

Страница 102: ...ry or hospital information system or a data management system you may see a message box prompting you to begin a patient sample assay The information system can also send a disable sampling order to t...

Страница 103: ...our system requires a password to access selected menu options the system prompts you to enter the Menu Options password Arterial samples analyzed for blood gas and CO oximeter values can determine th...

Страница 104: ...rs for the sample press Change Parameters to select another panel Refer to Selecting Parameter Panels page 2 29 for more information BIOHAZARD Refer to Appendix A Protecting Yourself from Biohazards f...

Страница 105: ...s and then returns to the Ready screen If the system continues to display the Results screen after the wash finishes press to return to the Ready screen If the system determines that there is insuffic...

Страница 106: ...erent set of parameters for the sample press Change Parameters to select another panel Refer to Selecting Parameter Panels page 2 29 for more information BIOHAZARD Refer to Appendix A Protecting Yours...

Страница 107: ...e has reached point A If not turn the sample pump clockwise to move the leading edge of the sample to point A as shown in Figure 2 4 b Press Point A Sample Pump 3 When prompted remove the sample devic...

Страница 108: ...and are continuously updated until analysis is complete The system then displays the final results and prints the report 8 The system performs a wash at the end of analysis and then returns to the Rea...

Страница 109: ...he remaining parameters after you move the sample into position Refer to Selecting Parameter Panels page 2 29 for more information NOTE The tHb parameters are not measured when you use the Micro Sampl...

Страница 110: ...is If the syringe contains sufficient volume the system analyzes the sample automatically 4 Press The Position Sample Manually message box appears 5 Turn the sample pump clockwise to move the leading...

Страница 111: ...sor Sample Pump 10 Press Results appear on the screen and are continuously updated until analysis is complete The system then displays the final results and prints the report 11 The system performs a...

Страница 112: ...ompted remove the sample device 6 Type the required information in the Patient Information screen and then press Refer to Entering Patient Sample Data page 2 27 if you need more information to complet...

Страница 113: ...hen you select pH only the system measures everything except pCO2 and pO2 On systems with the CO ox module the system measures everything except pCO2 pO2 and tHb parameters BIOHAZARD Refer to Appendix...

Страница 114: ...he sample remains in contact with the sample ground temperature sensor 5 Turn the sample pump clockwise until the leading edge of the sample fills the reference sensor as shown in Figure 2 11 Referenc...

Страница 115: ...s a wash at the end of analysis and then returns to the Ready screen If the system continues to display the Results screen after the wash finishes press to return to the Ready screen If the system ana...

Страница 116: ...he sample press Change Parameters to select another panel Refer to Selecting Parameter Panels page 2 29 for more information BIOHAZARD Refer to Appendix A Protecting Yourself from Biohazards for recom...

Страница 117: ...ys the final results and prints the report 6 The system performs a wash at the end of analysis and then returns to the Ready screen If the system continues to display the Results screen after the wash...

Страница 118: ...azards for recommended precautions when working with biohazardous materials 1 Prepare the sample and insert the syringe in the sample port as shown in Figure 2 13 Sample Port Expired Gas Sample 2 Pres...

Страница 119: ...hen displays the final results and prints the report 8 The system performs a wash at the end of analysis and then returns to the Ready screen If the system continues to display the Results screen afte...

Страница 120: ...els page 2 29 for more information BIOHAZARD Refer to Appendix A Protecting Yourself from Biohazards for recommended precautions when working with biohazardous materials 1 Prepare the sample and intro...

Страница 121: ...t ID and press you do not want to combine the 270 results with the 800 system results press If the 800 system completes analysis before the 270 CO oximeter completes analysis the Waiting for CO oximet...

Страница 122: ...with two different venous samples Samples analyzed for a v studies can produce the following results ctO2 a v ctO2 a v a VO2 DO2 Qsp Qt T The system determines a v study results using the following r...

Страница 123: ...ge box appears 5 Select and press The First a v Sample Selected message box appears 6 Press only one sample is found the Patient Data Search Result screen appears Press The Reporting Options message b...

Страница 124: ...ars 11 Press to view the a v study results The a v results are not automatically sent to the LIS Select Reporting Options on the patient results screen and then select Send Results to transmit the res...

Страница 125: ...t value of arterial NOTE Any field that has a symbol is a required field and must be completed 1 Type the patient ID and press You can enter patient sample data using the keypad an optional keyboard o...

Страница 126: ...a sequence number the date and the time of analysis in the status log The system returns to the Ready screen You can search for the sample results by sequence number and analysis time and then enter...

Страница 127: ...ed select the parameter panel and press Custom panels are listed as Panels 1 5 and are defined during system setup by the system administrator or authorized personnel specific parameters to analyze a...

Страница 128: ...reen the result dashes and arrow appear in red The result is below the measurement range On the screen the result dashes and arrow appear in red The patient sample did not reach endpoint The patient s...

Страница 129: ...ial Operator ID 12345678901 Patient ID 12345678901 Sex M Pt Name John Jones Physician ID 12345 Birthdate 05 30 22 Physician Dr Smith Age 70 Location ICU SYRINGE SAMPLE ACID BASE 37 C Units Reference R...

Страница 130: ...iewed See Troubleshooting Patient Results in Section 4 SulfHb 1 5 The system detects a sulfhemoglobin concentration greater than 1 5 in a CO ox sample Interfering Substance Glu The system detects the...

Страница 131: ...lds were not completed and Auto Send is turned on Insufficient COox Sample Appears when there is not enough sample to fill the CO ox module sample chamber and the analysis cannot be performed Insuffic...

Страница 132: ...Sample Temperature Out of Range The measurement module temperature is not in range at the end of measurement sequence __ Sent Appears when you press Send at the end of analysis...

Страница 133: ...his day Use this screen to enter the criteria that you want the system to use to search for patient sample data The system locates only the samples that meet all the criteria you specify For example i...

Страница 134: ...alysis Time From field and type the end time in the Analysis Time To field by sequence and accession number type the sequence number in the first Sample Sequence Number field 3 Press more than one sam...

Страница 135: ...e 2 19 Refer to the following procedures to edit patient data print the report or the log or transmit the report If you press Home while in the Patient Data Search Criteria screen all data is cleared...

Страница 136: ...edit fields that are entered by the system such as Analysis Date and Time and Sequence Number 1 Use the procedure described in Recalling Patient Sample Data page 2 35 to locate the sample you want to...

Страница 137: ...he sample you want as described in Recalling Patient Sample Data page 2 35 2 Press The Reporting Options message box appears print a sample report a Select Print Report and press b Press Refer to Figu...

Страница 138: ...ing with biohazardous materials 1 Press The Sample Type menu appears as shown in Figure 2 22 2 Select and press If you are using Bayer Diagnostics controls you can scan the bar code on the ampule now...

Страница 139: ...ss Done Press to assign the results to the file The system then displays the final results and prints the report Continue with step 8 you did not scan the ampule and Auto ID is off the QC File Informa...

Страница 140: ...ata the QC Data Mismatch message box appears Press OK to enter the correct data If you scan a bar code other than the QC ampule bar code the Bar Code Not Recognized message box appears Press OK The QC...

Страница 141: ...is manual Ensure that you handled and sampled the quality control materials according to the procedures recommended by the manufacturer If required analyze the quality control materials again Use this...

Страница 142: ...the results to the correct file 2 Type the Operator ID and press If you want to change text in the QC File Information screen after you press Done press The QC File Information screen reappears Make y...

Страница 143: ...screen appears as shown in Figure 2 25 The Analysis Date fields contain the current day s date If you do not complete any other fields the system recalls all QC samples for this day Use this screen to...

Страница 144: ...ate to the present date type the start date in the Analysis Date From field and type nothing in the Analysis Date To field for a specific range of dates type the start date in the Analysis Date From f...

Страница 145: ...ecord and press to view the previous report that appears on the log Select Search Criteria Screen and press to view the QC Data Search Criteria screen Press to close this message box one sample is fou...

Страница 146: ...ress until the status you want appears 3 Press The system displays the QC Data Search Result screen which now contains the changes you made 4 Press to return to the Ready screen If you change the file...

Страница 147: ...2345 Analysis Date Apr 12 1994 Expiration Dec 31 1995 Analysis Time 11 42 Results Units Target Range pH 7 613 7 594 7 634 pCO2 21 0 mmHg 19 8 23 8 pO2 148 8 mmHg 142 9 162 9 Na 155 9 mmol L 150 6 160...

Страница 148: ...or below the target range or exceeds action range QC sample result is above or below the action range ACCEPTED INTO QC FILE __ Indicates the QC result status accepted and the file number in which the...

Страница 149: ...System 850 1001 File 1 QC ID 1234 Level 1 Lot number 93419 Expiration date Jul 12 1995 Month to Date Statistics Apr 1 1994 To Apr 12 1994 Mean SD CV n pH 7 12 0 015 0 2 45 pCO2 80 4 3 1 3 9 48 pO2 49...

Страница 150: ...The QC Data Search Criteria screen appears 2 Type only the file number and press The QC Data Search Criteria Log screen appears 3 Select NOTE You can view the previous month s Levey Jennings charts b...

Страница 151: ...t Limit 2SD Target Value mean 3SD The Levey Jennings chart for the previous month displays only the data points for QC samples remaining in the file The Levey Jennings chart for the current month disp...

Страница 152: ...se to view a Levey Jennings chart the chart that appears is for the file number of the sample that you selected on the log The following symbols appear on Levey Jennings charts The results are accepte...

Страница 153: ...s to return to the Ready screen If the Auto Repeat function is on or Flexible Time calibration intervals are selected and a parameter shows excessive drift the system prints a report and repeats the a...

Страница 154: ...o not press Home the system automatically returns to the Ready screen after 45 seconds If the Auto Repeat function is turned on or Flexible Time calibration intervals are selected and a parameter show...

Страница 155: ...nd press b Select and press The tHb Slope Calibration screen appears as shown in Figure 2 33 2 Enter the slope value for the slope reagent and press The last entered value is displayed To remove this...

Страница 156: ...he Wash screen appears The system returns to the Ready screen at the end of the wash The maximum time between tHb slope calibrations is 90 days One day before a tHb slope calibration is scheduled a st...

Страница 157: ...1 Access the Barometer Calibration screen from the Menu screen a Select and press b Select and press 2 Type the correct atmospheric pressure and press The acceptable range for atmospheric pressure is...

Страница 158: ...nalyze samples until the required calibration completes You cannot defer a one point calibration that is initiated following rapid sampling You can interrupt or defer the one point calibration that oc...

Страница 159: ...n Date fields from the earliest date to a specific date leave the Calibration Date From field blank and type the specific date in the Calibration Date To field from a specific date to the present date...

Страница 160: ...calibration select the report you want and press The Calibration Data Search Result screen appears as shown in Figure 2 35 print a calibration report a Select the report you want b Press c Press The...

Страница 161: ...libration Data Search Result screen print the calibration report a Press b Press The calibration report prints as shown in Figure 2 36 transmit the results to an LIS or a data management system a Pres...

Страница 162: ...mol L Glucose 180 180 0 mg dL Lactate 0 00 0 02 0 mmol L Atmospheric pressure appears on all systems 840 850 860 840 850 860 7 Press when you finish more than one report is found the Done Options mess...

Страница 163: ...Drift __ Sensor drift is beyond predefined limits during a one point or a two point calibration D5 No Endpoint __ Calibration did not reach endpoint D3 Slope Error __ Sensor slope is beyond predefined...

Страница 164: ......

Страница 165: ......

Страница 166: ......

Страница 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...

Страница 168: ...6 0 6 3 6 45 6 6 6 3 6 0 0 0 6 6 3 0 6 0 0 6 6 4 6 1 0 6 0 0 6 6 6 6 6 2 0 4 6 2 6 0 0 6 6 4 6 4 6 1 2 6 6...

Страница 169: ......

Страница 170: ......

Страница 171: ......

Страница 172: ......

Страница 173: ...ppear at the top of the scroll list 2 Use the arrow keys to highlight the task you want to record as completed 3 Press Task Done The Maintenance Task Done dialog box appears 4 Enter your operator ID a...

Страница 174: ...ear at the top of the report The Due field displays the date on which the task is due next The Date Time Done field displays the date and time the maintenance task was last recorded as complete If a t...

Страница 175: ...3 Maintenance Log and press Enter The Maintenance Log screen appears as shown in Figure 3 2 2 Use the arrow keys to scroll through the task list 3 Press Previous Month to view tasks completed in the...

Страница 176: ...f household bleach NOTE Dilute household bleach 5 25 sodium hypochlorite 1 10 with reagent quality water reagent quality water44 lint free tissues or swabs BIOHAZARD Refer to Appendix A Protecting You...

Страница 177: ...3 96 b Clean the roller cage and the roller cage shaft with a 10 solution of household bleach c Rinse with reagent quality water and dry thoroughly d Reinstall the roller cage Wear safety glasses glo...

Страница 178: ...ature screen from the Menu screen a Select 3 Troubleshooting and press Enter b Select 2 Temp pAtm and press Enter 2 Press Start Test 3 Press Stop Test 4 Check the screen to verify that the sample temp...

Страница 179: ...e to your laboratory s barometer reading correct press Done incorrect a Type the correct atmospheric pressure and press Enter b Press Done to save the new atmospheric pressure and return to the Ready...

Страница 180: ...ule aspiration adapter 1 Perform a successful two point calibration 2 Prepare the High G L ampule and insert an aspiration adapter into the sample port and into the ampule 3 Press Analyze 4 When promp...

Страница 181: ...t blank sensors TB4 1 Prepare the deproteinizer as directed on the package 2 Access the Deproteinize screen from the Menu screen a Select 2 Maintenance and press Enter b Select 1 Deproteinize and pres...

Страница 182: ...on the biosensors with the contacts in the measurement module You can slide the remaining sensors to the right to create more space you want to condition the sensors press Yes The Conditioning screen...

Страница 183: ...st blank sensors c Go to step 3 3 Draw the conditioner solution into a syringe 4 Insert the syringe into the sample port 5 Press Analyze 6 When prompted remove the syringe 7 Wait 10 minutes for the co...

Страница 184: ...The main tank pressure must be higher than 300 psi The second stage pressure must be 3 to 5 psi 1 Check the main tank and second stage pressure for each gas tank by looking at pressure gauges on the g...

Страница 185: ...on of the fill solution in the pO2 and pCO2 sensors is normal NOTE The glucose and lactate biosensors do not require fill solution BIOHAZARD Refer to Appendix A Protecting Yourself from Biohazards for...

Страница 186: ...the pH K Cl and Ca sensors are almost full with only a bubble at the top and that the Na sensor is completely full If any sensor is not properly filled replace the fill solution as described in Filli...

Страница 187: ...h C1 is on your system replace it with C2 C2 is on your system replace it with C1 Wear safety glasses gloves and a laboratory coat when handling the reagents 1 At the Ready screen press Enter The Reag...

Страница 188: ...n the reagent fitting 7 Press Reset Levels The Reset Levels screen appears 8 Select C1 C2 and press Done Use this procedure to discard and replace the 7 3 CO ox Zero and Cal G L reagents if they are o...

Страница 189: ...A prime sequence starts followed by a wash sequence When the wash sequence finishes the Ready screen appears 9 Perform a two point calibration a Press Calibrate b Select Two point and press Enter c P...

Страница 190: ...ned and installed on the system for 60 days Materials required 6 838 Buffer Wash G L Zero Wear safety glasses gloves and a laboratory coat when handling the reagents 1 At the Ready screen press Enter...

Страница 191: ...alibration after changing the reagents to ensure that the reagents are acceptable and the system is functioning properly You can analyze quality control materials with the new reagents and compare the...

Страница 192: ...tubing according to your laboratory s biohazard protocol 7 Orient the new sample tubing as shown in the diagram on the system 8 Connect the new sample tubing to the sample probe 9 Place the sample tu...

Страница 193: ...tubing connector on the measurement module 12 Press Continue A wash sequence starts When the wash sequence sequence finishes a message box appears prompting you to perform a two point calibration 13 P...

Страница 194: ...iameter clot removal line bubble trap seal if necessary BIOHAZARD Refer to Appendix A Protecting Yourself from Biohazards for recommended precautions when working with biohazardous materials 1 Stop th...

Страница 195: ...mber Gasket Support Gasket Mounting Pin Holes Inlet Outlet CAUTION Do not touch the optical zone front or back area enclosed by the gasket of the sample chamber with your fingers Optical Zone Areas Sa...

Страница 196: ...reagent quality water d Blot both sides dry with lint free tissue 10 Clean the metal plate a Wipe both sides of the plate with a lint free tissue soaked in cleaning solution b Wipe both sides with a...

Страница 197: ...is 13 Reinstall the metal plate into the sapphire window by aligning onto the mounting pins 14 Reinstall the sample chamber a Connect the bubble trap to the sample chamber inlet port b Align the mount...

Страница 198: ...precautions when working with biohazardous materials 1 Stop the system from the Menu screen a Select 2 Maintenance and press Enter b Select 7 Stop System and press Enter 2 Disassemble the hemolyzer as...

Страница 199: ...ting pin Make sure the anvil is seated on the pin and does not rotate The anvil spring does rotate c Reconnect the sample tubing to the anvil d Reinstall the anvil cap to secure the anvil Listen for a...

Страница 200: ...er day If your laboratory analyzes more than 30 samples per day perform this maintenance more frequently Materials required air filter 1 Push the air filter cover down until it snaps out and then pull...

Страница 201: ...s required CO ox air filter 1 Push the air filter cover up until it snaps out and then pull the cover away from the system as shown in Figure 3 18 2 Remove the old air filter and discard it 3 Press th...

Страница 202: ...example information identified for an 860 also applies to an 865 1 Cut a piece of measurement module tubing to the appropriate length as shown in Figure 3 19 840 15 2 cm 6 inches 850 12 cm 4 75 inches...

Страница 203: ...ng loaded latch f Connect the right end of the tubing to the connector at position 3 g Press the tubing into the tubing groove h Continue with step 7 6 Replace the measurement module tubing a Remove t...

Страница 204: ...measurement module door 8 Press Continue A wash sequence starts When the wash sequence finishes a message box appears prompting you to perform a two point calibration 9 Press Yes to perform a two poin...

Страница 205: ...nect both pieces of tubing from the anvil c Remove the tubing guide on FD5 and gently pull the tubing through FD5 toward the sample connector d Disconnect the CO ox sample tubing from the sample conne...

Страница 206: ...g through the hole in the FD5 housing d Thread the replacement tubing through the hole in the FD5 housing e Connect the replacement tubing to the inlet on the reagent manifold f Connect the other end...

Страница 207: ...ing replace the tubing for all the pumps Replacing the tubing on only one or two pumps can affect pump calibration and cause incorrect flow rates Materials required sample pump tubing assembly waste p...

Страница 208: ...s from the CO ox pump as shown in Figure 3 24 3 Replace the reagent pump tubing as shown in Figure 3 23 a Select the replacement tubing that has the tubing collar b Grasp the left tubing cuff and pull...

Страница 209: ...e clockwise to gently work the new tubing between the platen and roller cage i Place the right tubing cuff under the right side of the platen 4 Connect the pump tubing sample pump tubing connect tubin...

Страница 210: ...g Place the tubing around the outside of the rollers h Hold the right tubing cuff below the right side of the platen and turn the roller cage clockwise to gently work both pieces of the new tubing be...

Страница 211: ...t 3 Troubleshooting and press Enter b Select 1 Fluidics System and press Enter c Select 2 Pump Flow Rate and press Enter d Press Calibrate Pump e Press Exit Test when the Pump calibration complete mes...

Страница 212: ...day perform this maintenance more frequently Materials required reagent manifold vent filter 1 Remove the old reagent manifold vent filter from the reagent manifold vent inlet and discard it 2 Insert...

Страница 213: ......

Страница 214: ......

Страница 215: ...ix A Protecting Yourself from Biohazards for recommended precautions when working with biohazardous materials Wear safety glasses gloves and a laboratory coat when performing this procedure 1 Press an...

Страница 216: ...lets as shown in Figure 3 28 a Grasp the waste outlet cover and pull it off b Wipe the outlets with a lint free tissue dampened with a 10 solution of household bleach c Clean the waste outlet cover wi...

Страница 217: ...ress Enter b Select 7 Stop System and press Enter 2 Push up the latches on the measurement module door and lift the door 3 Push the spring loaded latch to the right as shown in Figure 3 29 CAUTION Do...

Страница 218: ...arge static buildup 10 Reinstall the sensor a Gently align the top of the reference sensor with the sensor contact b Snap the sensor down into place c Press the tab on the spring loaded latch to relea...

Страница 219: ...or 3 Push the spring loaded latch to the right as shown in Figure 3 31 CAUTION Do not remove or return the sensors to the measurement module without first discharging static buildup Touch the inner su...

Страница 220: ...o release the latch 8 Close the measurement module door 9 Press Continue A wash sequence starts When the wash sequence finishes a message box appears prompting you to perform a two point calibration 1...

Страница 221: ...nded precautions when working with biohazardous materials 1 Stop the system from the Menu screen a Select 2 Maintenance and press Enter b Select 7 Stop System and press Enter 2 Disconnect the pump tub...

Страница 222: ...solution c Clean the roller cage shaft with a lint free tissue moistened with the bleach solution d Rinse the rollers the cage and the shaft with reagent water e Dry the roller cage thoroughly f Ensu...

Страница 223: ...10 with reagent quality water reagent water lint free swabs 1 Stop the system from the Menu screen a Select 2 Maintenance and press Enter b Select 7 Stop System and press Enter 2 Remove the required...

Страница 224: ...remove and clean other reagent fittings as required 11 Press Continue A wash sequence starts When the wash sequence finishes a message box appears prompting you to perform a two point calibration 12...

Страница 225: ...nter b Select 7 Stop System and press Enter 2 Pull off the sample door as shown in Figure 3 35 3 Grasp the tab on the retainer ring and firmly pull the tab toward you to rotate the ring 4 Grasp the sa...

Страница 226: ...paper from the printer c Remove the paper roll and set it aside 2 Cut a piece of standard copier paper in half lengthwise to make a strip of paper 10 8 cm 4 25 inches wide 3 Moisten the center of the...

Страница 227: ...glutaraldehyde solution to discourage microbial growth in the sample path CAUTION Do not use alcohol to perform this procedure Alcohol can damage the sensors When handling glutaraldehyde follow approp...

Страница 228: ...inish Press Cancel if you want to stop cleaning The cleaning cycle finishes The system performs an extended wash sequence The Calibrate System message box appears at the end of the wash sequence 5 Pre...

Страница 229: ...levels of quality control material to verify sensor performance 10 Empty and clean the waste bottle waste outlets and waste outlet cover as described in Emptying the Waste Bottle page 3 45 Materials r...

Страница 230: ...actate sensors TB4 c Go to step 2 Do not expose the reference sensor to bleach Replace the reference sensor with the test blank reference sensor TB5 Do not substitute a new reference sensor 2 Replace...

Страница 231: ...the module frame to discharge static buildup NOTE As you install the sensors ensure that the sensors are in the correct location Visually verify that you align the contacts on the sensors with the con...

Страница 232: ...creen a Select 2 Maintenance and press Enter b Select 8 Wash and press Enter 2 Press Home to return to the Ready screen 1 Access the Prime screen from the Menu screen a Select 2 Maintenance and press...

Страница 233: ...screen a Select 2 Maintenance and press Enter b Select 7 Stop System and press Enter 2 Perform the required maintenance task 3 Press Continue A wash sequence starts When the wash sequence finishes a...

Страница 234: ...nd press Enter 2 Press Print Statistics to print a workload statistics report The workload statistics report is shown in Figure 3 36 WORKLOAD STATISTICS REPORT Apr 12 1994 11 42 System 850 1001 Total...

Страница 235: ......

Страница 236: ......

Страница 237: ...the cap that contains the reagent septum Removing the cap damages the integrity of the reagent septum 4 Remove the plug from the cap of the new reagent bottle 5 Insert the new reagent bottle into pos...

Страница 238: ...aper in the printer Use the paper advance key to feed the paper through the printer 1 Stop the system from the Menu screen a Select 2 Maintenance and press Enter b Select 7 Stop System and press Enter...

Страница 239: ...r e Push down the printer lever Pull up the paper firmly as you close the cover to ensure that the paper does not jam under the cover f Close the printer cover g Press the paper advance key to ensure...

Страница 240: ...tations away from you You can fold the end of the paper 1 3 cm 0 5 inch and place it into the slot Press the paper advance key if there is not enough paper c Lower the paper spool until it snaps into...

Страница 241: ...turn the sensors to the measurement module without first discharging static buildup Touch the inner surface of the module frame to discharge static buildup 2 Remove the reference sensor as shown in Fi...

Страница 242: ...rvoir to the fill line as shown in Figure 3 42 CAUTION Do not overtighten the reservoir cap Overtightening can deform the gasket and cause leaks 5 Reinstall the reservoir cap and hand tighten 6 Contin...

Страница 243: ...the inner surface of the module frame to discharge static buildup 2 Remove the reference sensor as shown in Figure 3 43 a Push up the latches on the measurement module door and lift the door b Push th...

Страница 244: ...fragile and is easily damaged 4 Install the internal electrode in the sensor a Use the hex tool to remove the internal electrode from its container b Insert the internal electrode into the electrode...

Страница 245: ...with the hex tool and set it aside as shown in Figure 3 45 b Partially fill the KCl reservoir as shown in Figure 3 46 c Tap the front face of the sensor with your knuckle to release any bubbles d Gra...

Страница 246: ...ce and press Enter b Select 7 Stop System and press Enter CAUTION Do not remove or return the sensors to the measurement module without first discharging static buildup Touch the inner surface of the...

Страница 247: ...n the new cassette a Insert the tip of the KCl fill solution container into the electrode compartment as shown in Figure 3 48 b Slowly fill the electrode compartment by gently squeezing the container...

Страница 248: ...your knuckle to release any bubbles d Gradually fill the KCl reservoir with KCl fill solution up to the fill line Do not overtighten the reservoir cap Overtightening can deform the gasket and cause le...

Страница 249: ...remove or return the sensors to the measurement module without first discharging static buildup Touch the inner surface of the module frame to discharge static buildup 2 Remove the reference sensor as...

Страница 250: ...Fill the electrode compartment in the new cassette as shown in Figure 3 53 a Insert the tip of the KCl fill solution container into the electrode compartment in the cassette b Partially fill the elect...

Страница 251: ...the internal electrode into the compartment and screw it into place with the hex tool ensuring that you do not cross thread the electrode 10 Fill the KCl reservoir in the sensor a Remove the reservoi...

Страница 252: ...ge 3 89 Use this procedure to fill any measurement sensors that do not have sufficient fill solution Refer to Table 3 1 to determine the appropriate fill solution to use Materials required appropriate...

Страница 253: ...ce and press Enter b Select 7 Stop System and press Enter CAUTION Do not remove or return the sensors to the measurement module without first discharging static buildup Touch the inner surface of the...

Страница 254: ...l solution remaining in the sensor NOTE Refer to Table 3 1 to determine the appropriate fill solution to use 5 Rinse the electrode compartment with 3 drops of the appropriate sensor fill solution and...

Страница 255: ...refer to Replacing the Glucose and Lactate Biosensors page 3 86 Materials required appropriate sensor BIOHAZARD Refer to Appendix A Protecting Yourself from Biohazards for recommended precautions whe...

Страница 256: ...p and out of the measurement module d Discard the sensor 3 Perform the appropriate action the chloride sensor go to step 4 other measurement sensors or the sample ground temperature sensor go to step...

Страница 257: ...own into place c Press the tab on the spring loaded latch down to release the latch d Verify that the sensors are installed from left to right in the following order in the measurement module pO2 pCO2...

Страница 258: ...of the sensor rises to 37 C gas is driven from the solution causing bubbles Remove any bubbles present Materials required glucose biosensor lactate biosensor BIOHAZARD Refer to Appendix A Protecting Y...

Страница 259: ...urement module d Discard the biosensor 3 Remove the new biosensor from the foil package 4 Ensure that the O ring is in place on the biosensor CAUTION Do not remove or return the sensors to the measure...

Страница 260: ...Lac 6 Close the measurement module door 7 Press Continue A wash sequence starts When the wash sequence finishes a message box appears prompting you to perform a two point calibration 8 Press Yes to p...

Страница 261: ...that the vent hole in the reservoir cap is free of KCl crystals 3 Replace any O ring that is worn or damaged 4 Verify that the O rings are in place NOTE The reference sensor has an O ring on both sid...

Страница 262: ...Use a suitable hand truck to move tanks Never tamper with safety devices in regulators or tanks Use these gases for the calibration of clinical and research instrumentation only Do not dispense these...

Страница 263: ...me in contact with the gas stream Gas under pressure can cause bodily injury and property damage f To avoid injury or damage position the valve outlet so that it is facing away from your face body and...

Страница 264: ...dition and in place on the regulator as shown in Figure 3 62 e Attach the gas regulator to the gas tank by aligning the regulator nipple with the valve outlet and ensure that the dowel pins on the reg...

Страница 265: ...7 Press Continue A wash sequence starts When the wash sequence finishes a message box appears prompting you to perform a two point calibration 8 Check the gas flow rate from the Menu screen a Select...

Страница 266: ...QC results after the system is recalibrated If you do not install Bayer Diagnostics Cal Gas or Slope Gas tanks ensure that you define the calibration gas values for the gases used during calibration...

Страница 267: ...ress Continue A wash sequence starts When the wash sequence finishes a message box appears prompting you to perform a two point calibration 13 Press No 14 Check the gas flow rate from the Menu screen...

Страница 268: ...ls 1 Stop the system from the Menu screen a Select 2 Maintenance and press Enter b Select 7 Stop System and press Enter 2 Disconnect the pump tubing from the reagent manifold NOTE Do not stretch the t...

Страница 269: ...n the shaft until the roller cage stops 9 Press the roller cage down until the cage snaps into place 10 Reinstall the tubing as described in Replacing the Pump Tubing page 3 39 11 Press Continue A was...

Страница 270: ...irmly pull the tab toward you to rotate the ring 4 Grasp the sample port and attached drip tray and pull it off the mount as shown in Figure 3 66 NOTE Ensure that the three O rings are in place 5 Inst...

Страница 271: ...ss Enter 2 Remove the old probe a Disconnect the sample tubing from the connector on the measurement module b Disconnect the sample tubing from the bracket on the probe mount c Rotate the sample tubin...

Страница 272: ...cket on the probe mount e Connect the sample tubing to the sample tubing connector on the measurement module 4 Press Continue A wash sequence starts When the wash sequence finishes a message box appea...

Страница 273: ...scard the capillary seal according to your laboratory protocol 4 Install the new capillary seal a Hold the seal at the top angle it toward the right and push it into the opening in the sample port b I...

Страница 274: ...ou do not adhere to the time intervals you can damage the system 2 Wait until prompted and then disconnect the power cord from the power source 3 Disconnect the power cord from the side panel of the s...

Страница 275: ...form a two point calibration 15 Press No 16 Verify the system component temperatures from the Menu screen a Select 3 Troubleshooting and press Enter b Select 2 Temp pAtm and press Enter c Press Start...

Страница 276: ...ve the old lamp from the lamp housing and discard it 6 Place a new lamp in the housing aligning the lamp with the guide pin 7 Pinch the ends of the clamp together and replace it in the bracket 8 Insta...

Страница 277: ...fuses for the voltage you use 100 120V 4A Slo Blo 5 x 20 mm 220 240V 2A Slo Blo 5 x 20 mm To prevent electrical shock or damage to the system remove power from your system before performing this proce...

Страница 278: ...the fuse holders from the compartment 7 Remove the old fuses and install new fuses with the correct rating Refer to Table 3 2 to identify the correct fuse 8 Slide the fuse holders into the fuse compar...

Страница 279: ...atures from the Menu screen a Select 3 Troubleshooting and press Enter b Select 2 Temp pAtm and press Enter c Press Start Test d Verify that the system components warm up e Press Stop Test f If the te...

Страница 280: ......

Страница 281: ......

Страница 282: ......

Страница 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...

Страница 284: ...3 140 7 7 14 085 3 0 7532 6 5535 3 14 5 785 3 3 80 3 8 0 5 3 268 27 6 03 08 6 82 7 326 67 814 03 7 67 814 82 7 326 67 09 6 67 2 7 6 03 139 2 2 2 7 26356 08 7 735 67 140 275 67 14 5 785 4 71 67 685 1...

Страница 285: ......

Страница 286: ......

Страница 287: ...ontributed to an existing system problem 1 Access the Status Event Log from the Menu screen a Select 4 Data Recall and press Enter b Select 5 Status Event Log and press Enter The Message Date fields c...

Страница 288: ...o indicate that there are more messages which can be viewed in the status log Auto Clean Pending in __ Min Appears 5 minutes before an Auto Clean sequence is due and counts down until the sequence sta...

Страница 289: ...stem Messages page 4 115 Excessive Bubbles in COox Zero Appears when the system detects a non continuous fluid 7 3 buffer in the sample chamber and cannot perform the zero See Troubleshooting System M...

Страница 290: ...page 4 118 Insufficient COox Sample Appears when there is not enough sample to fill the CO ox sample chamber so the measurement cannot be completed See Troubleshooting System Messages page 4 117 Insu...

Страница 291: ...Sample Temperature Out of Range Appears when the measurement block temperature is not in range at the end of measurement sequence See Troubleshooting System Messages page 4 120 __ Sent Appears when y...

Страница 292: ......

Страница 293: ......

Страница 294: ......

Страница 295: ...es in the status area When the system displays a D code the View Status F key appears on the screen The Not Ready screen appears if the D code prevents the system from being ready Press View Status to...

Страница 296: ...mple Path in Section 3 Check for cracks or leaks in the sample tubing If necessary replace the sample tubing as described in Replacing the Sample Tubing in Section 3 Check for leaks or crimps in the g...

Страница 297: ...heck the solution level in the affected measurement sensor and look for bubbles If the solution level is low refill the sensor as described in Filling the Measurement Sensors in Section 3 Check the so...

Страница 298: ...alve to verify that the valve is working by performing the Valves test as described on page 4 62 Replace the affected sensor as described in Replacing the Measurement or Sample Ground Temperature Sens...

Страница 299: ...is closed as described in Removing and Checking the Sensors on page 4 65 Check the biosensor contacts and measurement block for salt buildup Check behind the biosensor for fluid leaks that can come fr...

Страница 300: ...libration Deproteinize the sample path as described in Deproteinizing the Sample Path in Section 3 Repeat the tHb slope calibration Check that the sample chamber is tightly in place Refer to Cleaning...

Страница 301: ...the Sample Tubing in Section 3 Check for leaks or crimps in the gas tubing If necessary replace the gas tubing as described in Replacing the Gas Tubing in Section 3 Check the gas flow rate and valve...

Страница 302: ...itioning the Sensors in Section 3 Check the solution level in the affected measurement sensor and look for bubbles The pH K Ca and Cl sensors should be nearly full The Na sensor should be full If the...

Страница 303: ...he O rings are in place and the spring loaded latch is closed as described in Removing and Checking the Sensors on page 4 65 Check the biosensor contacts and measurement block for salt buildup Check b...

Страница 304: ...the calibration Ensure the slope reagent is appropriate for the instrument Check that the sample chamber is tightly in place The tab on the cams should be vertical Repeat the tHb slope calibration Ch...

Страница 305: ...gas tanks as described in Replacing the Gas Tanks in Section 3 Check for leaks or crimps in the gas tubing If necessary replace the gas tubing as described in Replacing the Gas Tubing in Section 3 Ch...

Страница 306: ...rs in Section 3 If K or Ca have a D4 Offset Error replace the fill solution even if the fluid levels are sufficient If more than two sensors have the D4 Offset Error check the solution level in the re...

Страница 307: ...Checking the Sensors on page 4 65 Check the expiration dates and the levels of the reagents Ensure that the reagent bottles are installed properly and in the correct location If reagents have expired...

Страница 308: ...sensor Ensure measurement module door was not opened during measurement If door was opened repeat the sample analysis Perform a gas two point calibration Deproteinize the sample path as described in D...

Страница 309: ...evel in the reference sensor and look for salt in the vent hole bubbles in the sensor or leaks If the solution level is low or salt bubbles or leaks are present clean and refill the reference sensor a...

Страница 310: ...a metabolite two point calibration Perform the Measurement Test for the affected biosensor as described in Measurement Test page 4 70 If the glucose or lactate biosensors have been on the system less...

Страница 311: ...sample door is open or closed Check that your hand does not prevent the door from closing Check that the door is installed correctly Perform the Sample Entry Test as described on page 4 67 If the D c...

Страница 312: ...s at sample port Check for cracks or leaks in the sample tubing If necessary replace the sample tubing as described in Replacing the Sample Tubing in Section 3 Check the measurement module for leaks C...

Страница 313: ...rve the sample path for obstructions 2 If you see obstructions continue with the solutions to remove obstructions Check the measurement module for obstructions and remove them as described in Removing...

Страница 314: ...orrectly as described in Removing and Checking the Sensors on page 4 65 If you see obstructions remove them as described in Removing Obstructions from the Measurement Module page 4 82 Perform the Flui...

Страница 315: ...lifiers FD3 FD4 Perform a wash Observe the fluid as it moves through the system for sufficient wash flow wash segments and no leaks at the sample port Perform the Fluid Detector Test as described on p...

Страница 316: ...the sample tubing that goes through FD5 for cracks leaks and discoloration If you see cracks or leaks replace the tubing as described in Replacing the CO ox Sample Tubing in Section 3 Perform the Flui...

Страница 317: ...ct the CO ox sample path as described in Disconnecting the CO ox Sample Path page 4 92 You can continue to analyze samples on the base model Check that the communications cable connecting the CO ox mo...

Страница 318: ...10 kPa and that the reading is stable 5 Compare the atmospheric pressure displayed to the barometer in your laboratory to verify its accuracy 6 Press 7 Press Exit Test If the reading is stable but doe...

Страница 319: ...ate and level Ensure that the reagent bottle is installed properly and in the correct location If the C1 C2 has expired is low or is incorrectly installed replace it Prime the system Perform the Fluid...

Страница 320: ...d latch is closed as described as described in Removing and Checking the Sensors on page 4 65 Check the expiration dates and the levels of the reagents Ensure that the reagent bottles are installed pr...

Страница 321: ...Obstructions from the Sample Entry Components page 4 79 Check the measurement module for leaks Check that the sensors are installed in the correct order and are aligned the O rings are in place and th...

Страница 322: ...e capillary seal as described in Removing Obstructions from the Sample Entry Components page 4 79 2 Clean the sample probe with a lint free tissue moistened with a 10 solution of household bleach and...

Страница 323: ...tart it as described in Shutting Down and Restarting the System in Section 5 6 or 7 shut down and restart the system as described in Shutting Down and Restarting the System in Section 5 10 check the t...

Страница 324: ...erature sensor is installed correctly If the sensor is not correctly installed remove and install it as described in Cleaning the Sample Ground Temperature Sensor in Section 3 Check that the location...

Страница 325: ...reagents Ensure that the reagent bottles are installed properly and in the correct location If reagents are low or are incorrectly installed replace the affected reagents and perform a two point calib...

Страница 326: ...reagents Ensure that the reagent bottles are installed properly and in the correct location If reagents are low or are incorrectly installed replace the affected reagents and perform a two point calib...

Страница 327: ...t for the port where the error occurred as described in External Loopback Test page 4 75 If the External Loopback Test fails and the cable is firmly connected to the port replace the interface cable P...

Страница 328: ...s described in Cleaning the Sample Chamber It is not necessary to remove or replace the sample chamber gasket 3 Perform a pH Lytes one point calibration Perform CO ox Optics test Clean the sample cham...

Страница 329: ...r leaks Replace the tubing if necessary as described in Replacing the CO ox Sample Tubing in Section 3 If you move the tubing or unfasten the FD5 tube locking mechanism you must perform a manual wash...

Страница 330: ...e CO ox sample path for leaks or cracks If you find cracks or leaks replace the tubing as described in Replacing the CO ox Sample Tubing in Section 3 Check the CO ox sample path for obstructions If yo...

Страница 331: ...ox Optics Test page 4 75 Remove the lamp cover from the CO ox module and visually verify that the lamp is on If the lamp is not on replace it as described in Replacing the CO ox Lamp in Section 3 Visu...

Страница 332: ...and press Enter 3 Press Yes 4 Wait at least 1 minute and then disconnect the power cord from the power source 5 Wait at least 10 seconds and then connect the power cord to the power source 4 7 9 12 s...

Страница 333: ...nly at power up initialization The DAC value 016FH is loaded A dark measurement is initiated the DAC is loaded with 05D0H and another dark measurement is taken The difference in ADC counts between the...

Страница 334: ...ng is to be 3 0V 3 and 3 0V 3 respectively 10 testing occurs at start up and during the CO ox pump MIT Pump MIT selects CO ox This test checks for current flow through the windings A failure indicates...

Страница 335: ...ically clears the D77 condition If the problem has not been corrected the system displays the D77 condition code again Check that the ambient temperature of the laboratory is between 15 and 32 C Perfo...

Страница 336: ...uctions from the CO ox Sample Path page 4 87 Check the CO ox sample tubing for leaks or cracks If you find cracks or leaks replace the tubing as described in Replacing the CO ox Sample Tubing in Secti...

Страница 337: ...Perform the Pump Functions Test as described in Pump Functions Test page 4 58 If the D code reappears contact your Service Representative...

Страница 338: ...ample Tubing in Section 3 Check that the gaskets in the CO ox sample chamber and hemolyzer are seated correctly If necessary reinstall either gasket as described in Cleaning the CO ox Sample Chamber o...

Страница 339: ......

Страница 340: ......

Страница 341: ...s slope gas glu lac and diverter Gas Flow Rates Tests the ability of the Cal Gas and Slope Gas valves to control the flow of gases Fluid Detector Tests the ability of the fluid detectors to detect the...

Страница 342: ...rement module 1 Access the Fluidics Functions Test screen from the Menu screen a Select 3 Troubleshooting and press Enter b Select 1 Fluidics System and press Enter c Select 1 Fluidics Functions and p...

Страница 343: ...ure to test the flow rate of the reagent pump and to calibrate the reagent pump if necessary 1 Access the Pump Flow Rate Test from the Menu screen a Select 3 Troubleshooting and press Enter b Select 1...

Страница 344: ...he reagent sample waste and CO ox pumps The flow rate is the amount of fluid the pumps can pump in a specified time Materials required for the manual pump flow rate test stopwatch 10 mL graduated cyli...

Страница 345: ...action measure the flow rate of the reagent sample or waste pump continue with step 4 return to the Menu screen continue with step 11 4 Install the test tubing on the reagent sample or waste pump as s...

Страница 346: ...to left position 4 b and to the outlet tubing connected to right position 4 test the waste pump a to the pump inlet tubing connected to left position 5 b and to the outlet tubing connected to right p...

Страница 347: ...exactly 1 minute d Press Stop Test e Compare the volume of collected fluid to the expected ranges reagent 1 51 1 85 mL min 5 94 7 26 mL min sample 1 51 1 85 mL min 5 94 7 26 mL min waste 1 51 1 85 mL...

Страница 348: ...e waste tubing connector on the pump 13 Perform the appropriate action measure the flow rate of a different pump repeat steps 4 through 9 return to the Menu screen a Press Exit Test The Menu screen ap...

Страница 349: ...or Slope Gas valves refer to Checking the Gas Flow page 4 64 1 Access the Valves Test from the Menu screen a Select 3 Troubleshooting and press Enter b Select 1 Fluidics System and press Enter c Selec...

Страница 350: ...dure Sample Port Aspiration Adapter Water Steady Stream of Bubbles 1 Fill a small container with reagent water 2 Access the Valves Test from the Menu screen a Select 3 Troubleshooting and press Enter...

Страница 351: ...Troubleshooting menu and enter the Main menu the system performs a wash 1 Stop the system from the Menu screen a Select 2 Maintenance and press Enter b Select 7 Stop System and press Enter CAUTION Tou...

Страница 352: ...t 1 Fluidics System and press Enter c Select 5 Fluid Detector and press Enter The Fluid Detector Test screen appears as shown in Figure 4 8 2 Test the fluid detectors with clear fluid and air a Select...

Страница 353: ...t to test the fluid detectors with another type of fluid 6 Press Exit Test 7 Press Exit Menu Use this procedure to test the ability of the system to move the sample door and the sample probe identify...

Страница 354: ...pulled back into the sample port when prompted c Press Start Test d Remove the sample device when prompted e Check the screen to verify that the system correctly identifies the size and the type of th...

Страница 355: ...us area This test also displays the temperature readings for the measurement module the sample path the measurement module window the preheater and the barometric pressure NOTE The 37 C temperature se...

Страница 356: ...electronics You can print stored signals from the last successful two point calibration test the sensor circuit without a sensor and with a test blank sensor 1 Access the Measurement Test from the Men...

Страница 357: ...mV 25 to 90 42 5 to 68 5 mV Dec pO2 0 300 to 0 360 nA 5 to 90 0 002 to 0 020 nA mmHg Na 35 2 to 141 0 mV 15 to 90 42 5 to 68 5 mV Dec K 41 8 to 146 6 mV 15 to 90 42 5 to 68 5 mV Dec Ca 34 7 to 137 1 m...

Страница 358: ...ppropriate sensors a Open the measurement module door b If you are testing a pH or ISE sensor circuit remove the reference sensor c Remove the measurement sensor d Close the measurement module door 3...

Страница 359: ...ement module door b Perform the appropriate action pO2 or pCO2 sensor circuit TB1 test blank sensor pH or Na sensor circuit TB2 test blank sensor K Ca or Cl sensor circuit TB3 test blank sensor Glu bi...

Страница 360: ...ic buildup before removing or returning sensors 6 Reinstall the sensors a Open the measurement module door b Remove the test blank sensor c Reinstall the measurement sensor If you want to test another...

Страница 361: ...test passed The system requires a one point calibration to zero the CO ox module when the lamp test is complete The message Calibration Overdue COox Zero is displayed when the test is run The integrat...

Страница 362: ...800 system and the external device a Remove the cable from the port on the external device b Attach the loopback connector to the end of the cable 4 Press Start Test 5 Check the screen for the message...

Страница 363: ...o test the ability of the bar code scanner to read a test pattern 1 Access the Bar Code Scanner Test from the Menu screen a Select 3 Troubleshooting and press Enter b Select 6 Barcode Scanner and pres...

Страница 364: ......

Страница 365: ......

Страница 366: ......

Страница 367: ...with biohazardous materials 2 Remove obstructions from the sample entry area a Place a gauze pad in front of the sample port b Turn the sample pump counterclockwise until the obstruction is ejected o...

Страница 368: ...structions d Push a 0 558 mm 0 022 inch diameter clot removal line through the capillary seal against the direction of sample flow e Install a new capillary seal if you cannot remove the obstruction o...

Страница 369: ...nto the mount matching the tab on the sample port to the notch in the retainer ring b Grasp the tab on the retainer ring and turn the retainer ring away from you to lock it in place c Replace the samp...

Страница 370: ...s to the measurement module without first discharging static buildup Touch the inner surface of the module frame to discharge static buildup 3 Inspect the pO2 and pCO2 gas sensors for obstructions To...

Страница 371: ...structions a Remove the sensors one at a time and inspect the sample path for obstructions b Cut a piece of tubing or use a capillary adapter and place the tubing on the tip of the syringe c Push a 0...

Страница 372: ...on is not removed fill the syringe with fresh wash reagent and inject the wash reagent into the sensor sample path to remove obstructions As you install the biosensors ensure that the biosensors are i...

Страница 373: ...are in place 7 Remove obstructions from the path through the latch as shown in Figure 4 19 Do not remove or return the sensors to the measurement module without first discharging static buildup Touch...

Страница 374: ...ce 8 Press the release tab to close the spring loaded latch 9 Ensure that the sensors are aligned correctly 10 Close the measurement module door 11 Press Continue A wash sequence starts When the wash...

Страница 375: ...or air into the CO ox sample tubing Using a needle may damage the CO ox sample tubing 1 Stop the system from the Menu screen a Select 2 Maintenance and press Enter b Select 7 Stop System and press Ent...

Страница 376: ...ing from the inlet on the reagent manifold and manually turn the CO ox pump counterclockwise to remove the obstruction NOTE If air does not remove the obstruction place the waste tubing into a beaker...

Страница 377: ...om the Measurement Module for instructions d Reinstall the sample connector e Reconnect the tubing 4 Inspect the hemolyzer for obstructions a Remove the anvil cap on the hemolyzer by turning it a quar...

Страница 378: ...tubing from the inlet on the reagent manifold f Manually turn the CO ox pump counterclockwise to remove the obstruction from the sample tubing NOTE If air does not remove the obstruction from the samp...

Страница 379: ...mber c Push a 0 016 inch diameter clot removal line into the metal tubing of the outlet port of the bubble trap NOTE If the clot removal line does not remove the obstruction flush the bubble trap by i...

Страница 380: ...eters and press Enter c Move to the CO oximetry frame d Select the tHb parameter and press Enter This turns all CO ox parameters off e Press Next Screen to access the second screen of parameters f Pre...

Страница 381: ...6 Reinstall the screw 7 Reconnect the base model sample tubing 8 Press OK To reconnect the CO ox sample path turn the CO ox parameters on invert the sample connector The instructions in the procedure...

Страница 382: ...he following results reagents sensors and CO ox bar codes electronics roll printer probe pumps thermal control system messages If the problem remains after you perform the suggested solutions contact...

Страница 383: ...as described in Pump Functions Test page 4 58 or or The result is above up arrow or below down arrow the established control range 1 Verify that the control is prepared handled mixed and sampled corre...

Страница 384: ...heck reagent and instrument reporting categories for transcription or setup errors 6 Verify that the controls are not outdated or deteriorated Use controls only within their expiration date 7 Verify t...

Страница 385: ...agents are installed in the correct positions on the manifold 6 Perform a wash 7 Perform a successful two point calibration 8 Retest the sample and verify the results is printed next to the result on...

Страница 386: ...tain whether a value outside the range is likely 2 Check for sample movement during measurement 3 Check the fluid in the sample path for bubbles 4 Perform a wash 5 Ensure that you use the correct samp...

Страница 387: ...sample type the patient sample or the system NOTE When analyzed as a patient sample aqueous materials with dyes cause the If Blood Question Data message to appear 1 Ensure that you analyzed a patient...

Страница 388: ...erform a tHb slope calibration d Analyze the sample again e If the message reappears go to step 6 4 Review the last calibration 5 Obtain and analyze a new sample from the patient 6 If the message reap...

Страница 389: ...rt 1 Ensure that the sample probe is installed correctly 2 Check that the O rings are installed on the sample port and are not deteriorated Fluid is leaking in the measurement module 1 Verify that all...

Страница 390: ...5 Inspect the sample path in the measurement module for obstructions and leaks 6 Verify that all sensor O rings and sensors are correctly aligned with the sample path and that the O rings are not dete...

Страница 391: ...correctly 9 Clean the roller cage and shaft and replace the pump tubing if required No waste is flowing out of the waste outlets into the waste bottle 1 Check for obstructions in the sample entry com...

Страница 392: ...that all sensor O rings and sensors are correctly aligned with the sample path and that the O rings are not deteriorated NOTE The reference sensor has two O rings 7 Test the sample pump flow rate as...

Страница 393: ...Section 1 2 Check for obstructions in the sample entry components as described in Removing Obstructions from the Sample Entry Components page 4 79 3 Check for obstructions in the measurement module a...

Страница 394: ...Check the fill solution level in the sensors and empty and refill if required 5 Initiate the Auto Clean sequence 6 Perform two successful two point calibrations 7 Run controls and verify the results T...

Страница 395: ...ors for leaks and replace if required 5 Ensure the seal is in place between the gas tank and regulator 6 If the leak is between the gas tank and regulator tighten the yoke screw or replace the plastic...

Страница 396: ...sor Refer to the appropriate procedure in Section 3 More than two sensors have a D4 Offset Error 1 If the pH or ISE sensors show offset errors check the reference sensor for the correct solution level...

Страница 397: ...ppear on screen when you scan with the bar code scanner Incorrect symbology selected in system setup incorrect or defaced bar code labels or faulty bar code scanner cable or port incorrect field in fo...

Страница 398: ...d Appendix D Interfacing to External Devices 2 Check the cable connections at both ends 3 Check the cable for crimps or deterioration 4 Connect the external device LIS or HIS to another port as descri...

Страница 399: ...Clean the roll printer as described in Cleaning the Roll Printer in Section 3 3 Repeat the Roll Printer Test 4 Replace the roll of printer paper as described in Replacing the Printer Paper in Section...

Страница 400: ...t module temperature warning message The measurement module is outside the 37 00 0 15 C range The system can accept sample analysis 1 Ensure that the measurement module door is closed 2 Perform the Te...

Страница 401: ...ly 6 Verify that the location of the system meets the specifications for ambient operating temperatures as described in Appendix H Installation 7 Check the air filter to ensure that the air vent is no...

Страница 402: ...gned correctly with the sample path 4 Check the sample probe and the sample tubing for damage or incorrect installation 5 Verify that the reagent and sample pumps are operating correctly and test the...

Страница 403: ...tructions from the CO ox Sample Path in Section 4 5 Check for clots in the hemolyzer See Cleaning the Hemolyzer in Section 3 6 Clean the sample chamber as described in Cleaning the Sample Chamber in S...

Страница 404: ...with the sample type the patient sample or the system NOTE When analyzed as a patient sample aqueous materials with dyes cause the If Blood Question Data message to appear 1 Ensure that you analyzed a...

Страница 405: ...the message reappears contact your Service Representative Insufficient COox Sample The CO ox sample chamber does not detect the sample during the predefined time limit The CO ox measurement cannot be...

Страница 406: ...ate result on reports 1 Check the sample source and determine whether the presence of an interfering substance is likely See Appendix E Performance Characteristics for a list of interfering substances...

Страница 407: ...ensure that the air vent is not obstructed and that the filter is clean See Replacing the Air Filter in Section 3 8 Verify that the fan is operating Meas Module Temperature Warning The measurement mod...

Страница 408: ...the message appears repeatedly with capillary samples replace the capillary seal Sample Temperature Out of Range The measurement module block temperature is not in range at the end of the measurement...

Страница 409: ......

Страница 410: ......

Страница 411: ...3 0 5 2 3 5 5 0 5 0 5 3 5 5 0 5 5 5 5 5 0 5 0 5 2 5 2 5 0 0 1 2 5 5 0 5 0 5 0 5 4 5 0 0 5 3 1 5 0 0 5 5 0 2 5 5 0 5...

Страница 412: ...7 0 2 5 1 7 2 23 7 2 32 2 11 2 7 2 32 2 2 13 21 7 2 32 2 0 1 11 13 21 7 0 2 23 02 7 0 2 21 7 0 2 1 3 1 7 2 0 2 0 3 6 32 2 2 7 0 4 2 7 5 0 4 2 7 612 2 7 12 0 612 2 7 12 25 0 7 6 1 7 0 2 7 2 612 2 6 7...

Страница 413: ...Data select fields to appear on the screen and in the printed reports Operating Setup and Report Formats select units of measure for parameter results Operating Setup and Units Defaults select names f...

Страница 414: ...ce and System Information view system serial number ID install date software version and phone number for technical assistance Service and System Information You use the following keys while performin...

Страница 415: ...n Cancel discontinue the setup procedure Enter accept data you type in fields and selections you make on screens Enter also lets you move forward through fields Enter is located on the keypad Home ret...

Страница 416: ...n contain 16 fields The default for each field is On which means that all fields appear on the Patient Information screen You can choose to have a field not appear on the form by turning the field off...

Страница 417: ...per minute of the patient p50 partial pressure of oxygen at which hemoglobin is 50 saturated Sex gender of the patient Birthdate date patient was born Physician ID number used to identify the physicia...

Страница 418: ...does not appear on the Patient Information screen make a field a required field a Move to the Required check box for the field you want b Press Enter The check box becomes dark which indicates the fie...

Страница 419: ...f is evaluated during quality control testing 1 Access the Select Custom Panels screen from the Menu screen a Select 6 System Setup and press Enter b Select 4 Panels and press Enter c Select 2 Custom...

Страница 420: ...s for Analysis page 5 12 NOTE Default panels are used in patient sample analysis only All parameters are evaluated during quality control analysis 840 All Parameters pH pO2 pCO2 844 All Parameters pH...

Страница 421: ...H Na K Ca Cl glucose lactate pH Ca 864 All Parameters pH Blood gas CO ox pH lytes metabolites pH Ca pH Blood gas lytes metabolites pH pO2 pCO2 Na K Ca Cl glucose lactate tHb FO2Hb pH pO2 pCO2 tHb FO2H...

Страница 422: ...rs you want the system to analyze If you select a parameter to be off the system does not measure or calculate results during analysis and the parameter name does not appear on the Results screen or t...

Страница 423: ...peat steps a and b for each parameter you want to change 3 Press Done 4 You can define another setup function or press Exit Menu to return to the Ready screen When you select a measured parameter to b...

Страница 424: ...lues and press Enter The Units and Values screen appears as shown in Figure 5 6 2 Select the measurement units a Move to the frame that contains the measurement unit you want to change b Select the un...

Страница 425: ...these values for certain calculations Table 5 5 lists the default values and valid ranges for ctHb O2 binding factor p50 and ctO2 a v ctHb 2 0 25 0 20 250 1 2 15 5 15 0 g dL g L mmol L p50 0 0 100 0...

Страница 426: ...ng all the fields the Incomplete Data Entry message box appears Press OK and complete the required fields If you type an invalid value in a field and press Done the Invalid Entry message box appears P...

Страница 427: ...etup function or press Exit Menu to return to the Ready screen Use this procedure to define the reference and action ranges for each parameter The system uses reference ranges to determine if a result...

Страница 428: ...860 Cl mmol L 98 106 40 160 850 860 Glucose mg dL 66 93 10 999 860 Lactate mmol L 0 5 2 0 0 0 30 0 860 Tietz NW ed Fundamentals of Clinical Chemistry 3rd ed Philadelphia WB Saunders 1987 864 891 Weisb...

Страница 429: ...ess Enter c Select 1 Ref Action Ranges and press Enter The Reference and Action Ranges screen appears as shown in Figure 5 9 2 Move to the field that you want to change 3 Type the new value and press...

Страница 430: ...matically at the end of analysis The default is 1 Entry values are 1 3 NOTE This option is available only when the roll printer is selected Auto Print Lets you control whether the system prints a repo...

Страница 431: ...appears prompting you to set up another report want to select options for another report Yes The Printing Options screen reappears Repeat steps 2 thru 5 to select options for another report do not wan...

Страница 432: ...f each report You need an alphanumeric keyboard to type the information The 800 systems provide a variety of printed report formats Refer to Appendix F Printed Reports for an example of each report fo...

Страница 433: ...report a Select the report type you want b Press Print Report Format A copy of the report format prints on the printer you selected create information lines for a report a Press Enter Header The Ente...

Страница 434: ...inting Options and press Enter The Printing Options screen appears as shown in Figure 5 12 2 Move to the Roll Printer frame 3 Select On or Off and press Enter 4 Press Done The next Printing Options sc...

Страница 435: ...e Printing Options screen appears as shown in Figure 5 13 2 Move to the Paper Spool frame 3 Select On or Off and press Enter 4 Press Done The next Printing Options screen appears 5 Press Done again 6...

Страница 436: ...you can password protect the Operating Setup menu Operators must then enter the menu options password to access the Operating Setup menu They can still access other menu functions without a password R...

Страница 437: ...g maintenance tasks the Maintenance Task Done dialog box The operator ID also prints on the patient sample QC sample and maintenance log reports If you want to protect access to sample analysis select...

Страница 438: ...action define a new operator password a Type the ID in the Operator ID field and press Enter b Type the password in the Operator Password field and press Enter edit a password or ID retype the ID in t...

Страница 439: ...and can be used to access sample analysis Define a menu options password only if you want to password protect access to menu functions Refer to Selecting Password Protection page 5 26 to select menu o...

Страница 440: ...same password for the supervisor and menu passwords or create a different password for each c Type the password in the Menu Password field and press Enter d Type the same password in the Menu Passwor...

Страница 441: ...e date that appears on reports and on the screen uses three letters for the month for example 24 Jun 1994 YY MM DD 94 06 24 NOTE The date that appears on reports and on the screen is completely numeri...

Страница 442: ...If you press Done before completing all required fields the Incomplete Data Entry message box appears Press OK and complete the required fields If you type an invalid value in a field and press Done...

Страница 443: ...etup and press Enter b Press 5 Systems Options and press Enter The System Options screen appears as shown in Figure 5 18 2 Move to the Auto Clean Time field 3 Type the hour and minutes in the format H...

Страница 444: ...the start date 3 Take the appropriate action keep the existing maintenance schedule go to step 10 change the frequency of a maintenance task go to step 4 4 Select Edit Frequency 5 Use the arrow keys...

Страница 445: ...e 5 20 2 Highlight the Patient Sample Limit field 3 Type a number from 100 to 5000 The default is 1500 Set the limit slightly higher than the usual number of samples stored on the system before your s...

Страница 446: ...een from the Menu screen a Press 6 System Setup and press Enter b Press 5 Systems Options and press Enter The System Options screen appears as shown in Figure 5 21 2 Move to the Beeper Volume frame 3...

Страница 447: ...fault value is Yes No The operator moves capillary samples to the measurement module by turning the sample pump 1 Access the System Options screen from the Menu screen a Press 6 System Setup and press...

Страница 448: ...s to an LIS or data management system The default setting is On Off Results are not automatically transmitted The operator must press to transmit results or press to prevent transmission of results 1...

Страница 449: ...lue is high Low pH 0 01 H nmol L 1 pCO2 mmHg kPa 1 0 1 pO2 mmHg kPa 1 0 1 High pH 0 001 H nmol L 0 1 pCO2 mmHg kPa 0 1 0 01 pO2 mmHg kPa 0 1 0 01 1 Access the System Options screen from the Menu scree...

Страница 450: ...tem processes patient sample results when a sensor is out of calibration When the system detects a sensor that has excessive calibration drift it cannot analyze samples until the drift is corrected By...

Страница 451: ...25 2 Move to the Report with Cal Drift frame 3 Select Yes or No and press Enter 4 Press Done Yes A password message appears prompting you to enter the menu password and an operator ID a Type the menu...

Страница 452: ...ibes the slope and offset correlation values Slope 1 0 Slope of the line for the method you want the 800 system to match Offset 0 0 Difference between the 800 system value and the method you want to m...

Страница 453: ...parameter 3 Press Done when you finish 4 Define another setup function or press Exit Menu to return to the Ready screen If you press Done before completing all the fields the Incomplete Data Entry mes...

Страница 454: ...setup parameter or value The report is a record of your setup definitions which can be used later when you want to make another change 1 Access the Print Setup Report screen from the Menu screen a Se...

Страница 455: ...or the last 150 QC samples cumulative statistics such as the mean standard deviation and coefficient of variation total number of samples When you create a new QC setup file you enter the information...

Страница 456: ...and press Enter b Select 1 QC Setup and press Enter c Select 1 QC Files and press Enter 2 Select the file number you want to create and press Enter You cannot select File 13 as a QC file 3 Press Done...

Страница 457: ...ional bar code scanner scan the target range bar code for each parameter from QC Expected Values If you scan the wrong bar code a message appears prompting you to scan the correct parameter bar code P...

Страница 458: ...een Cancel The cursor returns to the last field you completed If you press Done before you complete both the low and high fields for a parameter a message appears telling you the empty field is an inv...

Страница 459: ...kette before performing this procedure Refer to Archiving QC Data page 5 68 for more information When you change QC lots replace the information shown in Table 5 18 with the data from the new QC mater...

Страница 460: ...appears prompting you to delete the current QC file setup data You must delete the current QC file data when you change lots Otherwise the QC file will contain data from the old and the new lots chang...

Страница 461: ...xisting QC file ranges move to the field requiring change press Clear Entry and type the new data Press Enter after you complete each field editing all of the existing QC file ranges press the Erase A...

Страница 462: ...urns to the last field you completed If you press Menu before you press Done in the QC File Setup screen a message appears prompting you to save your changes save the changes Yes The Menu screen appea...

Страница 463: ...s disabled or not in calibration Auto ID does not work Off The operator assigns the QC sample results to the appropriate file The default value is Off 1 Access the Global QC Settings screen from the M...

Страница 464: ...to have results automatically sent to an LIS or data management system On The system automatically accepts QC results and when Auto Send is on sends the results to an LIS or data management system Of...

Страница 465: ...ly transmits QC results to an LIS or data management system The default setting is On Off Results are not automatically transmitted The operator must press Send to transmit results or press Do Not Sen...

Страница 466: ...TUP File 1 QC ID 473842 Level 2 Lot number 41561 Expiration date Dec 30 95 MEAN TARGET RANGE ACTION RANGE pH 7 419 7 399 7 439 7 389 7 449 pCO2 43 4 38 4 48 4 35 9 50 9 pO2 103 1 96 1 110 1 92 6 113 6...

Страница 467: ...If drift exceeds the acceptable limits the results are flagged on the screen and on the report and are reported in the status log The drift limits defined in Table 5 22 are for cal points only The dri...

Страница 468: ...the drift limit the range is plus or minus 2 from the expected value 3 Type the new drift limit and press Enter 4 Repeat steps 2 and 3 for each parameter for which you want to enter or change the dri...

Страница 469: ...t 6 Calibration Setup and press Enter c Select 3 Gas Values and press Enter The Calibration Gas Values screen appears as shown in Figure 5 36 2 Move the cursor to the field that you want to change 3 T...

Страница 470: ...libration to avoid excessive drift in subsequent calibrations Flexible Time schedules the one and two point calibrations and lets you define a minimum time interval for one point calibrations Auto Rep...

Страница 471: ...propriate field b Type the amount of time you want between calibrations and press Enter Auto Repeat a Move to the Auto Repeat frame b Select On and press Enter to automatically repeat calibrations Sel...

Страница 472: ...on system Table 5 25 lists the ports you can use with each device Serial port 1 the ticket printer Serial port 2 and 3 the 270 CO oximeter and an LIS or data management system Parallel port other prin...

Страница 473: ...Code Options screen appears Select the appropriate symbology Select if Code 128 is the only symbology you require Then press Symbology 128 is always enabled when you install the bar code scanner You c...

Страница 474: ...nd the LIS HIS or data management system baud rate speed at which data is sent or received when devices are communicating through a serial channel parity method used to detect errors during transmissi...

Страница 475: ...nction or press Exit Menu to return to the Ready screen want to return to the last field edited Cancel When the system establishes the connection to an external device the Device Connected to Port __...

Страница 476: ...pes of system and service information such as the system model number the serial number and the service contact You can view the system information but you cannot change any of the information Refer t...

Страница 477: ...tilities functions Backup copy system files from the hard disk to diskettes that you can use to restore files or to copy troubleshooting trace log information for service representatives if required A...

Страница 478: ...ress Enter If you archive before you analyze the first QC sample of the current month you can print statistical summary reports of the previous month s QC data before you archive 2 Perform one of the...

Страница 479: ...een appears If you insert a diskette that already contains data a message box appears indicating that the system will overwrite the data currently on the diskette you want to overwrite the data Contin...

Страница 480: ...kette drive and press Continue The system reads the diskette When the system is ready a message appears containing archive diskette information 3 Press Continue 4 Remove the diskette when the system p...

Страница 481: ...ppears QC Sample Report a Select Print QC Sample Report and press Enter b Press OK QC Statistical Summary a Select Print Statistical Summary and press Enter b Press OK 10 When you finish viewing archi...

Страница 482: ...rol data calibration diagnostic and maintenance data setup data workload statistics and cycle counts Establish a backup schedule that meets your laboratory s data requirements The frequency with which...

Страница 483: ...tte message box appears while the system performs the backup 5 Complete the backing up process a message appears prompting you to insert a new diskette a Remove the diskette from the diskette drive b...

Страница 484: ...a formatted diskette and press Continue Use this procedure to copy data from your backup diskettes to the system hard disk CAUTION When you restore data from a backup diskette the restore process repl...

Страница 485: ...ystem A message appears on the screen and on the roll printer directing you to wait before you disconnect the power 7 Wait at least 1 minute 8 Disconnect the power cord from the power supply 9 Restart...

Страница 486: ...s before installing new software Refer to the procedure Backing Up System Data page 5 72 1 Access the Install Software screen from the Menu screen a Select 7 System Utilities and press Enter b Select...

Страница 487: ...rs The system identifies the diskette and displays the Install Identification message box Do not remove the diskette 5 Press Continue to proceed using the diskette the system just read The system copi...

Страница 488: ...system reads the last program diskette it is likely that at least one of the program diskettes was not installed Cancel the installation and start the procedure again When the system finishes the sof...

Страница 489: ...ains installed CAUTION Occasionally the system may shut down if you cancel the installation If this happens follow the instructions given in the Caution statement in step 7 and then continue through s...

Страница 490: ...3 Type the search criteria and press after you complete each field Press Done when you finish The Copying Files screen appears with the Begin Copying Files dialog box 4 Insert a clean formatted diske...

Страница 491: ...he contents of the column These headings list patient demographics or parameter names as found on an 800 system The copied file includes all headings even if there are no patient results in the column...

Страница 492: ...les detected in blood gas and CO ox sample 5 Bubbles detected in blood gas sample 6 Bubbles detected in CO ox sample Blank No exceptions 1 Entered value 2 Sensor has a drift D2 error 3 Interference de...

Страница 493: ...educes reagent consumption by turning off automatic calibrations While in standby the system performs the scheduled auto clean and purges the sensors on a regular basis 1 Access the Set Standby screen...

Страница 494: ...s the required calibrations and returns to the Ready screen If you turn off the power while the system is in standby the system returns to standby when you turn on the power If the system is scheduled...

Страница 495: ...ystem message indicates that the system has successfully completed the shutdown procedures Unplugging the system before this message appears can damage the system 3 When you see this message disconnec...

Страница 496: ......

Страница 497: ......

Страница 498: ......

Страница 499: ...04 7 00 26 04 321 0 301 7 2 021 7 1 1 20 22 612 1 7 6 0 2 0 1 2 7 3 0 21 6 7 00 26 5 31 1 7 2 2 1 6 0 12 1 0 00 26 7 3 1 11 0 1 7 2 2 7 5 2 0 7 2 2 2 2 0 2 0 7 2 2 0 0 7 2 2 6 0 12 1 2 2 612 7 2 2 0 2...

Страница 500: ...4 6 5 0 0 0 6 1 0 1 0 6 0 0 0 1 0 6 0 6 2 5 3 0 4 6 6 4 1 0 0 6 6 4 1 0 1 0 6 0 6 2 5 3 0 6 0 0 6 0 0 6 0 0 6 0 0 6 0 0 6 0 0 6 0 0 6 0 0 6 0 0 6 0 0 6 0 0 0 6 00 0 6 4 0 1 0 0 6 1 6 0 6 4 1 6 100 3...

Страница 501: ...0 2 3 0 0 1 7 2 0 1 0 7 0 0 5 0 7 0 1 5 0 7 0 0 0 5 7 0 0 7 0 5 7 7 5 0 2 05 7 7 4 7 7 45 7 7 0 50 7 0 0 1 7 1 7 0 0 0 1 7 0 1 7 0 0 7 7 0 0 6 1 7 1 7 0 0 7 0 0 1 7 0 0 0 0 7 1 0 0 7 0 2 0 2 7 0 7 45...

Страница 502: ...1 5 0 0 0 5 0 5 3 5 34 01 0 0 0 5 34 01 0 5 34 0 0 0 0 5 34 0 0 5 34 04 5 5 0 3 0 5 0 0 5 0 0 01 0 5 3 5 1 1 0 0 5 5 0 5 1 52 01 5 0 34 0 0 5 1 34 0 0 5 0 5 1 34 0 0 5 52 30 0 3 5 34 1 0 0 5 34 2 4 5...

Страница 503: ...r sources of contamination when handling potentially infectious agents are as follows needlesticks hand to mouth contact hand to eye contact direct contact with superficial cuts open wounds and other...

Страница 504: ...ith 15 v v bleach Do not eat drink smoke or apply cosmetics while in the laboratory Do not mouth pipet any liquid including water Do not place tools or any other items in your mouth Do not use the bio...

Страница 505: ...es Argentina 54 11 4 762 7000 Bayer S A Produtos Diagn sticos Rua Domingos Jorge 1100 04779 900 S o Paulo SP Brazil 55 11 5694 5574 Bayer Australia Limited Diagnostics Business Group 2 Keith Campbell...

Страница 506: ...No 615 Col del Valle Delegaci n Benito Ju rez M xico D F C P 03100 525 7283301 Bayer Hellas S A Akakion 54A Marousi Athens 151 25 Greece 01 68 83 648 Bayer B V Division Diagnostics Energieweg 1 3641 R...

Страница 507: ...Bayer House Strawberry Hill Newbury RG14 1JA United Kingdom 44 0 1635 563000 Qu mica Farmac utica Bayer S A Divisi n Diagn sticos Calabria 268 08029 Barcelona Espa a 34 93 495 65 00 Bayer Corporation...

Страница 508: ...iate Agreement Addendum must indicate the term in months for additional service coverage The customer may obtain warranty service for systems during normal business hours by contacting the Bayer Diagn...

Страница 509: ...of system During the warranty period Bayer Diagnostics reserves the right to change the design or construction of specific models of systems without incurring any obligation to make such changes avail...

Страница 510: ...days of shipment to the customer s facility unless otherwise specified 4 The customer has not performed appropriate customer maintenance procedures as outlined in the system operator s manual 5 The s...

Страница 511: ...warranties express or implied accompanying either the leasing of the equipment or its sale to the customer at the expiration or termination of this agreement In addition the warranties of merchantabil...

Страница 512: ...anual Rapidlab 800 Interface Specification 106363 Brochure Specimen Collection for Critical Blood Analyte Testing 673709000 Printer Paper 473385000 Buffer 7 382 7 3 CO ox Zero Buffer 473386000 Buffer...

Страница 513: ...r Code Scanner LS4004 110889 Bar Code Scanner LS4004 110890 Cable Bar Code Scanner LS4004 858086001 Cable Bar Code Scanner LS2020 673715000 Air Dust Filter 673712000 Capillary Seal 013896701 Sample Do...

Страница 514: ...ensor 476246000 pO2 Ready Sensor 476270000 Potassium Ready Sensor 476266000 Sodium Ready Sensor 476268000 Calcium Ready Sensor 476279000 Chloride Ready Sensor 476378000 Glucose Ready Sensor 476379000...

Страница 515: ...over 477833000 Analysis Report Ticket 477434000 Cal Calibration Gas 5 CO2 12 O2 477438000 Slope Calibration Gas 10 CO2 823736001 Gas Tank Seal 478740000 Sample Ground Temperature Sensor 013662001 Wast...

Страница 516: ...050 CO ox Lamp 570018 CO ox O ring Gasket Kit for hemolyzer and sample chamber 673700000 Clot Removal Kit tubing diameter 0 022mm 570049 Sample Chamber 106371 Bubble Trap Kit 015552701 CO ox Spares Ki...

Страница 517: ...ory Workers from Instrument Biohazards and Infectious Disease Transmitted by Blood Body Fluids and Tissue Approved Guideline NCCLS Document M29 A Vol 17 No 20 Dec 1997 8 National Committee for Clinica...

Страница 518: ...ne 3rd ed Chicago Year Book Medical Publishers 1978 273 281 22 National Committee for Clinical Laboratory Standards Definitions of Quantities and Conventions Related to Blood pH and Gas Analysis Appro...

Страница 519: ...gy 6th ed St Louis CV Mosby 1982 388 392 39 National Committee for Clinical Laboratory Standards Fractional Oxyhemoglobin Oxygen Content and Saturation and Related Quantities in Blood Terminology Meas...

Страница 520: ...Chicago Year Book Medical Publishers 1989 140 50 Malley W Clinical blood gases application and noninvasive alternatives Philadelphia WB Saunders 1990 97 98 51 Shapiro BA Harrison RA Cane RD Kozlowski...

Страница 521: ...r code scanner Bayer Diagnostics data management system line printer laboratory information system LIS hospital information system HIS Figure D 1 identifies the various ports available on the 800 syst...

Страница 522: ...igure D 1 4 Connect the 25 pin connector to port 1 2 or 3 on the 270 5 Tighten the hold down screws on the connectors 6 Restore power to the 800 system and to the 270 7 Refer to Configuring for Extern...

Страница 523: ...anual to specify whether the 270 automatically sends results as soon as they are available or whether the operator can select the results to transmit Use this procedure to connect an 800 system to the...

Страница 524: ...bar code scanner cable To prevent electrical shock and damage to either system disconnect the 800 system from the AC power source before installing the bar code scanner Refer to Shutting Down and Res...

Страница 525: ...ring for External Devices in Section 5 to configure the 800 system for the data management system Use this procedure to connect an 800 system to a line printer Materials Required printer to 800 system...

Страница 526: ...efore installing the cable Refer to Shutting Down and Restarting the System in Section 5 for the procedure to disconnect the 800 system 1 Contact your LIS or HIS system manager to determine the approp...

Страница 527: ...ation about the 800 series systems system specifications limitations reference methods 840 system performance characteristics 850 system performance characteristics 860 system performance characterist...

Страница 528: ...4V 50 60Hz ambient operating barometric pressure 400 825 mmHg 53 0 110 0 kPa system dimensions base model height 50 8 cm 20 inches width 55 9 cm 22 inches depth 48 3 cm 19 inches weight 29 5 kg 65 lbs...

Страница 529: ...0 800 0 0 106 67 0 1 0 01 1 0 1 pAtm mmHg kPa 400 825 53 3 110 0 1 0 1 1 0 1 Table E 3 lists the units reporting ranges and display resolutions for the electrolyte parameters measured or calculated b...

Страница 530: ...or calculated from a sample analyzed on the 800 system and a connected CO oximeter p50 mmHg kPa 15 0 75 0 2 00 10 00 0 1 0 01 ctO2 mL dL mL L mmol L 0 40 0 0 400 0 17 8 0 1 1 0 1 ctO2 a mL dL mL L mm...

Страница 531: ...0 1 DO2 mL min L min mmol mi n 0 3500 0 00 3 50 0 156 2 1 0 01 0 1 Table E 5 lists the units reporting ranges and display resolutions for parameters measured by the 800 system CO ox module tHb g dL 2...

Страница 532: ...g 15 0 75 0 0 1 kPa 2 00 10 00 0 01 ctO2 Hb mL dL mL L mmol L 0 40 0 0 400 0 17 8 0 1 1 0 1 Table E 6 lists the units reporting ranges and display resolutions for the temperature corrected and respira...

Страница 533: ...Qt T decimal fraction 0 100 0 00 1 00 1 0 01 Qsp Qt est T decimal fraction 0 100 0 00 1 00 1 0 01 Table E 7 lists the units reporting ranges and display resolutions for the metabolic parameters calcu...

Страница 534: ...0 0 43 9 50 0 111 0 0 1 0 1 FIO2 decimal fraction 15 0 100 0 0 150 1 000 0 1 0 001 ctHb g dL g L mmol L 2 0 27 0 20 270 1 3 16 8 0 1 1 0 1 OBF 1 35 1 40 0 01 Flow L min 0 00 99 99 0 01 Resp Rate b min...

Страница 535: ...d samples are tonometered with 95 N2 and 5 CO2 FCOHb Gas Chromatography for COHb 15 and Ishizawa method using a Cary 4 Spectrophotometer for COHb 15 FMetHb Modified Evelyn Malloy37 method utilizing a...

Страница 536: ...any of these situations occur Reagents other than those recommended are used Expiration dates of reagents have been exceeded Reagents are not used according to Bayer Diagnostics recommendations Standa...

Страница 537: ...y control QC materials and calibration verification materials CVM were analyzed on the 840 systems The results are presented here Precision on aqueous quality control materials was estimated using thr...

Страница 538: ...n For testing syringe capillary microsyringe and microcapillary modes blood was collected in heparinized vacuum tubes It was tonometered at 37 0 C to each of three levels to prepare samples for pH ana...

Страница 539: ...48 7 238 100 1 pH Only 15 0 004 7 235 7 235 100 0 2 Syringe 12 0 003 7 427 7 420 100 1 Capillary 14 0 003 7 429 7 423 100 1 Microcapillary 12 0 004 7 380 7 374 100 1 Microsyringe 15 0 002 7 400 7 397...

Страница 540: ...21 4 100 2 Microsyringe 24 0 15 21 6 21 4 101 2 Expired Gas 24 0 14 21 6 21 4 100 7 3 Syringe 24 0 22 35 8 35 7 100 4 Capillary 18 0 53 35 3 35 7 98 8 Microcapillary 23 0 77 36 2 35 7 101 3 Microsyrin...

Страница 541: ...lary 24 1 15 29 0 28 5 101 7 Microsyringe 24 0 17 28 6 28 5 100 3 Expired Gas 24 0 15 28 2 28 5 98 8 2 Syringe 24 0 19 50 0 49 9 100 3 Capillary 18 0 23 50 5 49 9 101 2 Microcapillary 24 0 71 49 0 49...

Страница 542: ...101 4 Microsyringe 24 0 80 150 6 149 7 100 6 Expired Gas 24 0 25 150 4 149 7 100 5 5 Syringe 24 1 20 377 9 377 9 100 0 Capillary 18 2 40 379 0 377 9 100 3 Microcapillary 24 4 08 380 7 377 9 100 7 Micr...

Страница 543: ...with your 850 system Quality control materials and calibration verification materials were analyzed on the 850 systems The results are presented here Precision on aqueous quality control materials was...

Страница 544: ...n run standard deviation TotSD total standard deviation Precision on aqueous calibration verification materials was estimated using four 850 systems As many as seven runs per instrument were made over...

Страница 545: ...vels to prepare samples for pH analysis and five levels to prepare samples for pCO2 and pO2 analysis Blood was spiked diluted to each of three levels to prepare samples for sodium potassium calcium an...

Страница 546: ...002 7 229 7 225 100 1 pH Lytes 18 0 003 7 220 7 218 100 0 2 Syringe 24 0 001 7 399 7 405 99 9 Capillary 24 0 002 7 397 7 393 100 1 Microcapillary 24 0 003 7 402 7 396 100 1 Microsyringe 24 0 002 7 38...

Страница 547: ...7 21 4 96 9 Microsyringe 18 0 42 21 3 21 4 99 4 Expired Gas 18 0 77 21 0 21 4 98 3 3 Syringe 18 0 43 35 1 35 7 98 2 Capillary 24 0 25 36 4 35 7 102 0 Microcapillary 18 0 50 34 5 35 7 96 6 Microsyringe...

Страница 548: ...illary 18 0 35 29 0 28 5 101 8 Microsyringe 18 0 25 28 5 28 5 100 2 Expired Gas 18 0 16 27 9 28 5 98 0 2 Syringe 24 0 22 50 5 49 9 101 2 Capillary 18 0 34 49 2 49 9 98 6 Microcapillary 24 0 41 49 4 49...

Страница 549: ...7 100 2 Microsyringe 18 0 62 149 4 149 7 99 8 Expired Gas 18 0 31 149 5 149 7 99 8 5 Syringe 21 1 40 377 4 377 9 99 9 Capillary 21 2 41 377 5 377 9 99 9 Microcapillary 18 5 50 378 9 377 9 100 3 Micros...

Страница 550: ...e 24 0 49 150 9 151 0 99 9 Capillary 24 0 46 150 9 151 2 99 8 Microcapillary 24 0 80 151 2 151 1 100 0 Microsyringe 24 0 49 151 0 151 3 99 8 pH Lytes 24 0 75 151 3 151 6 99 7 3 Syringe 24 0 39 174 2 1...

Страница 551: ...ge 24 0 020 3 49 3 51 99 5 Capillary 24 0 034 4 14 4 14 100 0 Microcapillary 24 0 037 3 81 3 78 100 7 Microsyringe 24 0 016 3 81 3 82 99 7 pH Lytes 24 0 013 3 74 3 74 100 2 3 Syringe 24 0 109 7 21 7 2...

Страница 552: ...ge 21 0 012 1 87 1 84 102 0 Capillary 24 0 017 1 85 1 82 101 8 Microcapillary 24 0 010 1 65 1 65 99 7 Microsyringe 21 0 008 1 66 1 58 104 8 pH Lytes 18 0 023 1 64 1 57 104 5 3 Syringe 21 0 028 2 48 2...

Страница 553: ...Syringe 24 0 7 117 116 100 5 Capillary 24 0 3 117 116 100 5 Microcapillary 24 1 1 117 116 100 8 Microsyringe 24 0 6 116 116 100 4 pH Lytes 24 1 4 118 119 99 6 3 Syringe 24 0 5 140 141 99 9 Capillary 2...

Страница 554: ...erfering substance up to the test concentration shown in the following two tables The interference was calculated by comparing the spiked sample to an unspiked sample immediately preceding it To evalu...

Страница 555: ...se concentrations therapeutic levels of dopamine do not interfere with the glucose measurement The following anticoagulants were found not to interfere with glucose recovery at the indicated concentra...

Страница 556: ...with the glucose measurement Sodium Fluoride 1000 mg dL each 25 mg dL 1 4 mmol L Acetominophen 2 mg dL 7 mg dL 0 4 mmol L Sodium Fluoride Potassium Oxalate 1000 mg dL each 25 mg dL 1 4 mmol L Increase...

Страница 557: ...te recovery in heparinized blood The following table lists substances that were found not to interfere with the lactate measurement At the concentrations listed these compounds produced less than 0 3...

Страница 558: ...g dL Theophylline 9 mg dL Penicillamine 25 mg dL Isoniazid 2 mg dL Uric Acid 10 mg dL The following anticoagulants were found not to interfere with lactate recovery at the indicated concentrations but...

Страница 559: ...period Two replicates of each control level were analyzed in each run For glucose precision on aqueous calibration verification materials was estimated using fifteen 860 systems Data from seven to si...

Страница 560: ...21 0 50 3 236 151 3 0 40 0 77 K 1 387 2 68 0 014 0 020 2 316 5 04 0 014 0 022 3 236 7 34 0 033 0 050 Cl 1 387 119 5 0 57 1 14 2 316 100 0 0 34 0 74 3 236 77 6 0 41 0 57 Ca 1 386 1 47 0 006 0 011 2 31...

Страница 561: ...marizes the results of the 860 system precision for CVM levels 1 and 4 pH 1 565 6 782 0 002 0 004 4 551 7 830 0 007 0 009 pCO2 1 575 107 6 1 48 3 11 4 548 13 0 0 37 0 74 pO2 1 563 26 0 1 63 2 02 4 539...

Страница 562: ...e expired gas mode 10 mL of tonometry gas were drawn into a 12 mL syringe and aspirated into the 860 system for analysis Multiple runs were made using five levels of expired gas on four 860 systems Th...

Страница 563: ...Microsyringe 24 0 10 14 2 14 3 99 5 Expired Gas 18 0 09 14 7 14 3 102 9 2 Syringe 24 0 12 21 8 21 4 101 7 Capillary 24 0 19 21 6 21 4 100 9 Microcapillary 24 0 33 21 6 21 4 101 2 Microsyringe 24 0 28...

Страница 564: ...yringe 24 1 46 71 0 71 3 99 5 Expired Gas 24 0 24 70 5 71 3 98 8 WRSD within run standard deviation 1 Syringe 24 0 21 28 5 28 5 99 9 Capillary 24 0 25 28 5 28 5 100 1 Microcapillary 24 0 44 28 6 28 5...

Страница 565: ...150 1 149 7 100 2 Microsyringe 24 0 95 150 6 149 7 100 6 Expired Gas 24 0 19 150 0 149 7 100 2 5 Syringe 24 1 52 378 8 377 9 100 2 Capillary 24 1 55 378 7 377 9 100 2 Microcapillary 24 4 69 378 9 377...

Страница 566: ...36 0 98 147 1 147 4 99 9 Microsyringe 33 0 58 148 3 148 4 99 9 pH Lyte 24 0 57 148 4 148 6 99 9 171 5 Syringe 24 0 38 172 1 171 6 100 3 Capillary 24 1 25 171 6 171 2 100 2 Microcapillary 36 2 44 171...

Страница 567: ...ge 48 0 028 3 75 3 73 100 6 Capillary 48 0 027 3 72 3 69 100 7 Microcapillary 48 0 027 3 63 3 66 99 3 Microsyringe 48 0 038 3 78 3 78 100 2 pH Lyte 48 0 022 3 54 3 51 100 9 3 Syringe 24 0 142 8 47 8 5...

Страница 568: ...ringe 48 0 014 1 67 1 65 100 8 Capillary 44 0 011 1 66 1 65 100 5 Microcapillary 48 0 026 1 67 1 65 100 9 Microsyringe 48 0 014 1 67 1 67 100 4 pH Lyte 48 0 022 1 66 1 66 100 1 2 38 Syringe 48 0 036 2...

Страница 569: ...7 Syringe 24 0 6 116 117 98 9 Capillary 24 1 0 118 118 100 3 Microcapillary 36 1 0 117 116 100 2 Microsyringe 36 1 2 116 117 99 4 pH Lyte 24 0 6 117 117 99 5 142 Syringe 24 1 2 141 142 99 8 Capillary...

Страница 570: ...ary 36 2 5 79 79 99 8 Microsyringe 45 1 9 84 82 102 3 pH Lyte 24 1 1 81 80 101 1 207 Syringe 33 3 5 205 205 100 4 Capillary 33 2 9 206 203 101 5 Microcapillary 36 3 9 206 205 100 8 Microsyringe 45 3 8...

Страница 571: ...ringe 36 0 046 2 83 2 89 98 1 Capillary 36 0 074 2 67 2 80 95 4 Microcapillary 36 0 145 2 69 2 73 98 6 Microsyringe 35 0 057 2 70 2 84 95 1 pH Lyte 36 0 066 2 73 2 79 98 0 3 94 Syringe 18 0 019 3 80 3...

Страница 572: ...m temperature about 23 C You should determine your own performance characteristics in your laboratory with your 800 system To test for interferences blood was spiked with an interfering substance up t...

Страница 573: ...lipid Bilirubin Fetal Hemoglobin 20 mg dL at 21 Table E 42 lists substances that showed interference using the criteria stated in Table E 40 Evans Blue at 5 mg L O2Hb MetHb 2 1 1 4 Methylene Blue at 2...

Страница 574: ...ere Precision on aqueous quality control materials was estimated using four 800 systems One run per instrument was made over a eight day period Three replicates of each control level were analyzed in...

Страница 575: ...llary modes blood was collected in heparinized vacuum tubes It was tonometered at 37 C and or adjusted chemically Three levels of tHb were prepared by separating the red cells from the plasma and reco...

Страница 576: ...0 00 0 22 0 05 WRSD within run standard deviation 1 0 1 tHb Capillary 68 15 18 0 36 0 35 Syringe 66 15 00 0 31 0 11 FO2Hb Capillary 68 92 31 0 09 0 25 Syringe 66 92 27 0 23 0 29 FCOHb Capillary 68 4 5...

Страница 577: ...Capillary 72 90 03 0 73 0 24 Syringe 68 92 31 1 21 0 64 FCOHb Capillary 72 6 56 0 53 0 16 Syringe 68 4 48 0 95 0 72 FMetHb Capillary 72 3 41 0 08 0 21 Syringe 68 3 21 0 01 0 20 FHHb Capillary 72 0 00...

Страница 578: ......

Страница 579: ...s examples of the patient sample reports that you can print on the 800 system roll printer The examples include all possible parameters patient sample data fields temperature corrected values entered...

Страница 580: ...C ctHb xxx xx mmol L xx xxx xxxx x Hct xxx xx ctO2 a xxx xx mL dL xxx xx xxxx x BO2 xxx xx mL dL xxx xx xxxx x pO2 xxx xx mmHg p50 xxxx x mmHg sO2 xxx xx xxx xx xxxx x FO2Hb xxx xx xxx xx xxxx x FCOHb...

Страница 581: ...t xxx xx mmol L HCO3 std xxx xx mmol L ctCO2 xxx xx mmol L BE B xxx xx mmol L BE ecf xxx xx mmol L OXYGEN STATUS 37 C Corrected 38 5 C ctHb xxx xx g dL Hct xxx xx ctO2 a xxx xx mL dL BO2 xxx xx mL dL...

Страница 582: ...C Corrected 38 5 C pCO2 xxx xx xxx xx mmHg pO2 xxx xx xxx xx mmHg ACID BASE 37 C Corrected 38 5 C pH xx xxx xxx xx HCO3 act xxx xx mmol L HCO3 std xxx xx mmol L ctCO2 xxx xx mmol L BE B xxx xx mmol L...

Страница 583: ...n ID xxxxxxxxxxxx Age xxx Location xxx SYRINGE SAMPLE ACID BASE 37 C Corrected 38 5 C pH xxx xx xxx xx pCO2 xxx xx xxx xx pO2 xxx xx xxx xx HCO3 act xxx xx HCO3 std xxx xx ctCO2 xxx xx BE B xxx xx BE...

Страница 584: ...pH xx xxx xxx xx nmol L pCO2 xxx xx xxx xx mmHg pO2 xxx xx xxx xx mmHg ACID BASE 37 C Corrected 38 5 C HCO3 act xxx xx mmol L HCO3 std xxx xx mmol L ctCO2 xxx xx mmol L BE B xxx xx mmol L BE ecf xxx...

Страница 585: ...nt sample data fields temperature corrected values entered values and CO oximeter values for 800 systems The examples include the following reports Line Printer Report A Line Printer Report B Line Pri...

Страница 586: ...xxxx x p50 xxx xx mmHg xxx xx xxx xx sO2 xxx xx xxx xx xxx xx FO2Hb xxx xx xxx xx xxx xx FCOHb xxx xx xxx xx xxx xx FMetHb xxx xx xxx xx xxx xx FHHb xxx xx xxx xx xxx xx OXYGEN STATUS 38 5 C pO2 A a...

Страница 587: ...etHb xxx xx xxx xx xxxx x FHHb xxx xx xxx xx xxxx x OXYGEN STATUS 37 C OXYGEN STATUS 38 5 C ctO2 a xxx xx mL dL xxx xx xxx xx BO2 xxx xx mL dL xxx xx xxx xx pO2 xxx xx mmHg xxx xx xxx xx xxx xx p50 xx...

Страница 588: ...COHb xxx xx FMetHb xxx xx FHHb xxx xx OXYGEN STATUS 37 C OXYGEN STATUS 38 5 C ctO2 a xxx xx mL dL BO2 xxx xx mL dL pO2 xxx xx mmHg xxx xx xxx xx xxx xx p50 xxx xx mmHg pO2 A a T mmHg xxx xx pO2 a A T...

Страница 589: ...you do not complete a field on the Patient Information Screen the field does not appear on the ticket printer Refer to Defining the Printer Report Format in Section 5 for information about how to sel...

Страница 590: ...BE B xxxxx mmol L BE ecf xxxxx mmol L OXYGEN STATUS 37 C CORRECTED 38 5 C ctHb xxxxx mmol L Hct xxxxx ctO2 a xxxxx mL dL BO2 xxxxx p50 xxxxx sO2 xxxxx mL dL FO2Hb xxxxx FCOHb xxxxx FMetHb xxxxx FHHb x...

Страница 591: ...n from the Menu screen a Select 5 Operating Setup and press Enter b Select 8 Correlation and press Enter The Correlation Coefficients screen appears 2 Ensure that the slope values are set to 1 000 and...

Страница 592: ...ation y mx b where m is the slope and b is the offset The calculation must be made in the manner described The calculation of a correction equation is the mathematical inverse of the traditional corre...

Страница 593: ...1 for the system specifications ambient operating temperature 15 32 C ambient operating relative humidity 5 85 non condensing power rating 400VA maximum voltage requirements 100V 120V 85V to 132V 50...

Страница 594: ...the installation tray and check the contents against the printed list included in the box 9 Open the system a Turn the fasteners one half turn to the left with a Phillips screwdriver b Lift the top o...

Страница 595: ...oor using a small flat blade screwdriver and continue with step e your system has a voltage bobbin continue with step a a Take the voltage bobbin from the customizing box and the fuse kit from the ins...

Страница 596: ...Notch Voltage Selection Bobbin Window Voltage Bobbin 100 VAC 120 VAC e Check the system fuses f Pull one of the fuse holders out of the fuse compartment as shown in Figure H 3 Fuse Fuse Holder Fuse co...

Страница 597: ...Insert the power cord into the power input connector on the right side panel 16 Rotate the system so that the front panel faces you 17 Adjust the Display assembly a Remove the packing material from th...

Страница 598: ...gas tanks with the regulator nipple engaging the opening in the tank valve g Attach and tighten the yoke screws firmly h Attach and tighten the tubing adapter fittings into each needle valve Ensure t...

Страница 599: ...eft to the right or behind the system n To prevent restrictions in the tubing and uneven gas flow secure the slope and cal gas tubing so that the tubing does not cross under the system o Reinstall the...

Страница 600: ...e electrode compartment with 4M KCl solution and tap to remove bubbles b Continue to fill the internal reference sensor compartment until the 4M KCl fill solution flows into the reservoir c Use the he...

Страница 601: ...e sensor body with 3 drops of the Na K Cl Ca fill solution and empty e Slowly add the fill solution until the sensor is almost full f Screw the internal electrode into place g Tap the sensor with your...

Страница 602: ...the small tubing to position 2 33 Install the sample pump and waste pump tubing a Disconnect the right side of the sample pump tubing b Place the tubing over the top of the roller cage c Hold the righ...

Страница 603: ...atch the movement of wash solution to verify that the wash reagent flows through the system during the wash sequence that automatically follows the prime sequence 38 Verify the gas flow rates for cal...

Страница 604: ...s in Section 3 CAUTION Do not remove or return the sensors to the measurement module without first discharging static buildup Touch the inner surface of the module frame to discharge static buildup 43...

Страница 605: ...tative Unauthorized use of these procedures can void the warranty or service contract For detailed information about the software refer to Learning About the System in Section 1 and Operating the Syst...

Страница 606: ...800 CO ox module 3 Open the box Cut the tape along the edge of the top flap Open the flaps 4 Remove the product inserts Unpacking Instructions Ship Damage Instructions Contact Bayer Diagnostics Repre...

Страница 607: ...g you to wait before you disconnect the power Do not unplug the system until the following message appears on the screen synching disks done This is an operating system message indicating that the sys...

Страница 608: ...he edge of the work surface to access the front plug 4 Remove the two plugs mounted to the right side wall of the base model hydraulic wall a Push on the plug stems that protrude from the interior of...

Страница 609: ...ted in the groove located under the chassis shelf of the 800 base model Ensure that the two bosses on the CO ox module are seated in the two receptacles of the base model and that the screws are align...

Страница 610: ...strument label 1 Access the preheater on the measurement module and disconnect the sample tubing 2 Remove the sample connector a Remove the screw located in the front of the preheater block that attac...

Страница 611: ...pin on the spindle arm bracket while pushing the cover all the way to the left b Bend the tab on the cover or pivot it to allow the tab to slide over the pivot pin 1 Locate the groove on the left side...

Страница 612: ...nd set up your system Refer to System Administration in Section 5 for more information about setting up the system 10 Initiate a two point calibration from the Menu screen a Select 1 Calibration and p...

Страница 613: ...rs to the measurement module without first discharging static buildup Touch the inner surface of the module frame to discharge static buildup 1 Take the appropriate action 840 or 850 Go to step 2 860...

Страница 614: ...een a Select 2 Maintenance and press Enter b Select 2 Condition and press Enter Wear safety glasses gloves and a laboratory coat when handling the reagents 5 To clean the sample pathway a Insert an as...

Страница 615: ...t 10 Press Exit Test 11 Empty rinse and fill the four reagent bottles with reagent water 12 Install the reagent bottles containing reagent water in place on the system 13 Perform a Prime sequence as d...

Страница 616: ...he sample port the drip tray and the mount with reagent water Ensure that the three O rings are in place 7 Reinstall the sample port matching the tab on the sample port to the notch in the retainer ri...

Страница 617: ...over as described in Emptying the Waste Bottle in Section 3 5 Remove the printer paper Refer to Replacing the Printer Paper in Section 3 6 Tape the printer cover in place 1 Shut down the system from t...

Страница 618: ...clockwise b Disconnect the gas regulators from the gas tanks by unscrewing the yoke screws c Verify that the gas tank seals are in place on the regulators d Remove the gas tanks to a well ventilated o...

Страница 619: ...ose of the tanks according to your laboratory protocol Bayer Diagnostics recommends that you remove the valve stems before disposal 2 Pack the 800 system in the original shipping carton If the origina...

Страница 620: ......

Страница 621: ...specific substance The transducer mechanism converts the potential generated by the molecular recognition mechanism to an electrical signal In the 800 systems this is accomplished through potentiomet...

Страница 622: ...loped by the electrochemical cell is determined by calculating the difference in potential between the measuring electrode and the reference electrode Ecell Emeas Eref Elj where Ecell electrochemical...

Страница 623: ...on The membrane separates an inner reference element which is immersed in a fixed electrolyte solution from the sample During analysis a membrane potential develops as a result of the interaction of t...

Страница 624: ...potential K a constant from various sources such as the liquid junction R gas constant T absolute temperature Z ionic charge F Faraday s constant ai activity of the ion in the sample This equation st...

Страница 625: ...solution controlling the ionic strength of calibrating solutions to 160 mmol kg sets the activity coefficients of ionic species in the calibrating solutions equal to those of blood plasma water at sam...

Страница 626: ...solution and the sample The membrane completes the conductive path to the sample from the fixed half cell potential that is required for the measurement The Ag wire conducts the half cell potential of...

Страница 627: ...sample comes in contact with the two electrodes a known voltage is applied between the anode and the cathode The analyte to be measured is either an oxidizable or reducible species If the analyte is...

Страница 628: ...1909 Sorenson13 converted the numbers mathematically to simplify their use and described the notation pH pH log 10 cH where H is the molar concentration of hydrogen ions Using this formula a hydrogen...

Страница 629: ...ane that is highly sensitive and specific for hydrogen ions separates the sample from the solution Silver Silver Chloride Wire Buffer Solution Sample Path As the sample comes in contact with the membr...

Страница 630: ...e is proportional to the content of dissolved CO2 HCO3 the value for pCO2 can be used along with pH not only to calculate HCO3 but also to aid in the differentiation of acid base abnormalities The mea...

Страница 631: ...auses a change in the hydrogen ion activity CO2 H2O HCO3 H The internal pH electrode detects the change in hydrogen concentration occurring in the chloride bicarbonate solution and generates a half ce...

Страница 632: ...measurement of pO2 is significant in evaluating the degree of hypoxemia a deficiency of O2 in arterial blood present in a patient The laboratory reference value for pO2 is usually 95 mmHg 12 7 kPa fo...

Страница 633: ...f oxygen in the electrolyte solution is determined 16 The 850 and 860 systems analyze blood samples for sodium Na potassium K chloride Cl and calcium Ca in addition to pH and the blood gases These sys...

Страница 634: ...e potassium sensor The potassium sensor membrane is designed as a charge separator As shown in Figure I 9 the positively charged potassium ions selectively interact with the membrane when the sample i...

Страница 635: ...omplete electrochemical cell The sensor contains a silver silver chloride wire surrounded by an electrolyte solution that has a fixed concentration of sodium and chloride ions The membrane a specially...

Страница 636: ...n electrolyte solution that has a fixed concentration of potassium ions The membrane which consists of the ionophore valinomycin immobilized in a plasticized PVC polyvinyl chloride matrix separates th...

Страница 637: ...erence sensor The final measured potential is directly proportional to the chloride ion concentration in the sample The potential developed by the electrochemical cell varies with the ion activity in...

Страница 638: ...surrounded by an electrolyte solution that has a fixed concentration of calcium ions A membrane consisting of an ionophore imbedded in a polyvinyl chloride membrane separates the electrolyte solution...

Страница 639: ...Glycolysis is the term commonly used to describe the conversion of glucose to lactic acid Under normal circumstances glycolysis occurs during muscle contraction where the rate of metabolism outpaces...

Страница 640: ...removed from the total differential measurement A microporous cover membrane separates the electrodes from the sample Sample Path Counter Electrode Glucose Measuring Electrode Reference Electrode Inte...

Страница 641: ...rmine the concentration of total hemoglobin hemoglobin derivatives and dyshemoglobins in whole blood The presence of dyshemoglobins and toxins changes the oxygen binding capacity of hemoglobin and the...

Страница 642: ...nsport and in the evaluation of anemia The total hemoglobin reference range for a normal adult population is 12 0 to 18 0 g dL Total hemoglobin in the CO ox module is determined using the following eq...

Страница 643: ...b ctHb 100 Carboxyhemoglobin COHb is hemoglobin covalently bound to carbon monoxide Hemoglobin has over 200 times greater affinity for carbon monoxide than for oxygen Hemoglobin bound to carbon monoxi...

Страница 644: ...on or substance at a specific wavelength and C is the concentration of the substance These equations are based on the work of VanAssendelft33 34 and Benesch Benesch and Yung 35 The 800 systems CO oxim...

Страница 645: ...is warmed to 37 C and then measured The sample chamber also acts a fluid detector sensing when sufficient sample reaches the chamber for measurement This section describes the specific reported parame...

Страница 646: ...gin In metabolic acidosis HCO3 levels decrease causing an increase in H which leads to a decrease in pH Conversely in metabolic alkalosis HCO3 levels increase causing a decrease in H which leads to an...

Страница 647: ...ting on all patient reports The calculations for both versions of base excess are derived from the following relationships which are based on the NCCLS recommendations 22 The base excess of extracellu...

Страница 648: ...e value obtained is equivalent to fractional oxyhemoglobin 26 Hemoglobin oxygen saturation sO is a ratio of the amount of hemoglobin bound to oxygen to the total amount of hemoglobin able to bind oxyg...

Страница 649: ...8 The oxygen content of hemoglobin with hemoglobin oxygen saturation and oxygen capacity is a useful parameter for determining the amount of oxygen in the blood that is actually available to the tissu...

Страница 650: ...n the range of 1 30 to 1 40 specified in the CO ox module setup Half saturation of hemoglobin by oxygen p50 indicates the partial pressure of oxygen when oxygen has saturated 50 of the available hemog...

Страница 651: ...g equation is used ctCO2 cHCO3 0 0307 pCO2 Hematocrit Hct is the ratio of the volume of packed red blood cells to the volume of whole blood 53 Hematocrit with other parameters such as total hemoglobin...

Страница 652: ...at the bedside However they do not have a high level of correlation with the actual measurement of arterial and mixed venous blood and should be used with discretion A more reliable method is the Qs...

Страница 653: ...on that remains relatively stable when FIO2 changes It is useful in predicting oxygen tension in alveolar gas The following equation29 is used pO2 a A T pO2 a T pO2 A T where pO2 a T is the temperatur...

Страница 654: ...y approximate the balance of these unmeasured ions An anion gap result is of twofold value in a clinical laboratory Primarily abnormal anion gap results indicate electrolyte imbalance or other conditi...

Страница 655: ...e oxygen content of venous blood ctO2 v is a determination of the total oxygen carried by the venous blood including the oxygen bound to hemoglobin and the oxygen dissolved in plasma and in the fluid...

Страница 656: ...en uptake is a determination of the volume of oxygen consumed by the body per minute 48 The system determines the oxygen consumption rate as follows VO2 ctO2 a v Qt 10 Oxygen delivery DO2 which is als...

Страница 657: ...the physiologic shunt The best alternative method for reflecting changes in the physiologic shunt is the estimated shunt Qsp Qt est T value which is applicable to most hypoxemic patients with cardiov...

Страница 658: ...135 0 148 0 mmol L K 3 50 5 30 mmol L Ca 1 13 1 32 mmol L Cl 98 106 mmol L Glucose 66 8 93 2 mg dL 3 7 5 2 mmol L Lactate 0 5 2 0 mmol L Tietz NW ed Fundamentals of clinical chemistry 3rd ed Philadelp...

Страница 659: ...ce Checklist charts for you to record maintenance activities performed on the 800 system Make photocopies of these charts as necessary and record your maintenance activities according to the maintenan...

Страница 660: ......

Страница 661: ......

Страница 662: ......

Страница 663: ......

Страница 664: ......

Страница 665: ......

Страница 666: ......

Страница 667: ...sistor 25 C Maximum silicate mg L 0 05 0 1 1 0 Particulate matter m smaller than 0 22 m not applicable not applicable water is passed through a 0 22 m filter Organic compounds pretreat with activated...

Страница 668: ...methods 1 Figure 1 Water Purification System Maintaining your Purification System Efficient operation and regularly sched uled maintenance of your water purifi cation system is the key to optimizing t...

Страница 669: ...ocument C3 A2 2 Tietz Norbert W Fundamentals of clinical chemistry 3rd ed Philadelphia W B Saunders Company 1987 1010 p 3 Kaplan Lawrence A Pesce Amadeo J Clinical Chemistry theory analysis and correl...

Страница 670: ...yer Diagnostics Limited Hong Kong Bayer Diagnostics India Limited Gujarat India Bayer S P A Divisione Diagnostici 20151 Milan Italy Bayer Diagnostics Korea Ltd Seoul Korea Bayer de M xico S A de C V D...

Страница 671: ......

Страница 672: ......

Страница 673: ...stems or methods before you report patient results Analyze The key that begins sample analysis Analyze mode The normal operating state in which you can analyze patient and QC samples and perform calib...

Страница 674: ...system that includes the measurement module fluidic components and reagents For example the base model for an 865 is an 860 biosensor A device that utilizes an active biomolecule such as an enzyme to...

Страница 675: ...ioner A solution used during the conditioning procedure that cleans and conditions the sensors CO oximeter CO ox A device that spectrophotometrically measures the absorption of whole blood at several...

Страница 676: ...rded QC data into an active file or update the statistics You can retrieve the discarded data using the Data Recall function if the data was accidentally discarded Also called File 14 diskette A 3 5 i...

Страница 677: ...rough F5 located below the screen that are associated with the labels displayed on the bottom line of the screen The key functions can change as you move from screen to screen flag A symbol that appea...

Страница 678: ...highlight bar The area of the screen that appears in reverse video to indicate the selected choice in a scroll list or menu HIS Hospital information system A computer system that is used for data man...

Страница 679: ...esses the Calibration Troubleshooting Maintenance Data Recall Operating Setup System Setup System Utilities and Service Setup menus Menu screen The screen that appears when you are in Menu mode From t...

Страница 680: ...ey that tightens the printer paper on the spool Parameter Names The Operating Setup menu option that lets you select the chemical symbols to identify parameters Parameters The System Setup menu option...

Страница 681: ...Operating Setup menu option that lets you select a report type turn on automatic printing enter the number of copies to be printed and select a printer Purge A function that moves solutions through t...

Страница 682: ...t sample or calibration result is outside the expected range reference range A range of values with upper and lower limits that the system uses to evaluate patient sample results The reference range r...

Страница 683: ...s used for serial ports to connect printers and other devices to the system sample connector The component that splits the sample path to allow a sample to go to both the measurement module and the CO...

Страница 684: ...theoretical value for a calibration solenoid valve The component associated with opening and closing the path within the fluidic system to control the passage of fluids and gases Standby The System Ut...

Страница 685: ...d a barometer test Troubleshooting The Menu screen option that lets you access diagnostic tests Units Values The Operating Setup menu option that lets you select measurement units that the system uses...

Страница 686: ...unction For example during sample analysis the system displays a data entry form you use to enter patient demographic data and when analysis is finished it displays results Workload Stats The Data Rec...

Страница 687: ......

Страница 688: ......

Страница 689: ...Additional Messages message 4 6 air filter removing illustration 3 32 replacing 3 32 replacing CO ox 3 33 alveolar oxygen tension I 32 alveolar arterial oxygen tension difference I 33 Analyze key 1 24...

Страница 690: ...e 1 46 barometer 1 47 2 59 drift limits 1 48 5 57 valid entry ranges 5 57 frequency 5 60 glucose biosensor 1 46 interrupting 2 60 manually initiated 1 47 one point 1 45 1 64 to 1 74 options 5 57 to 5...

Страница 691: ...9 to 4 54 D3 Slope Error Lac 4 19 D3 Slope Error pH Na K Ca Cl 4 18 to 4 54 D3 Slope Error pO2 pCO2 4 17 to 4 54 D3 Slope Error tHb 4 20 D4 Offset Error Glu 4 23 to 4 54 D4 Offset Error Lac 4 23 D4 Of...

Страница 692: ...o 4 115 Excessive Bubbles in CO ox Sample message 2 33 4 7 Excessive Bubbles in CO ox Zero message 4 7 Excessive Bubbles in tHb Slope 4 115 Excessive Bubbles in tHb Slope message 4 7 Excessive Scatter...

Страница 693: ...5 32 limitations on blood gas analysis 1 33 line printer configuring an 800 system for 5 62 connecting to an 800 system D 5 examples of printed reports F 7 LIS canceling an assay order 2 4 communicat...

Страница 694: ...9 measuring pH I 8 measuring pO2 I 11 other reported parameters I 25 potentiometry I 1 to I 38 operator password default 5 28 operator passwords defining 5 27 to 5 29 Operator Passwords Screen 5 28 ot...

Страница 695: ...31 symbols 2 30 screen 2 30 transmitting to an LIS HIS or data management system 5 38 troubleshooting 4 97 patient temperature correction I 32 pCO2 measuring I 9 sensor I 11 performance 840 data E 11...

Страница 696: ...bration verification 1 56 how often to run 1 51 materials 1 50 options 5 45 to 5 56 statistics 1 51 to 1 52 reagent delivery components illustration 1 19 reagent fittings cleaning 3 53 to 3 54 removin...

Страница 697: ...1 27 illustration 1 9 1 23 message box 1 27 option button 1 27 scroll list 1 27 sensor icons 1 27 1 28 text field 1 27 timing bar 1 27 Select Custom Panels screen 5 9 to 5 19 Select Default Panel scr...

Страница 698: ...alyzed 5 12 password protection 5 26 to 5 27 operator passwords 5 27 to 5 29 Patient Information form 5 6 to 5 8 printer report format 5 22 printing options 5 20 to 5 21 reference and action ranges 5...

Страница 699: ...e thermal control 4 112 tubercle bacillus A 1 two point calibration report example 2 64 Unexpected Device Detected message 4 9 Units and Values form screen 5 16 Units and Values screen 5 14 user inter...

Отзывы: