Refer to Sample Collection Devices and Anticoagulants in Section 1 for the specific
requirements on sample handling and anticoagulants.
)'($"#" #!! &# # ## "$!#
$"#
#!# "#
% #!!
Sodium Fluoride
1000 mg/dL
1 mmol/L (9 mg/dL)
Sodium Fluoride/
Potassium Oxalate
1000 mg/dL
1 mmol/L (9 mg/dL)
Acetominophen
2 mg/dL
0.35 mmol/L (3.2 mg/dL)
!" #! "
Quality control materials and calibration verification materials were analyzed on
the 860 systems. The results are presented here.
Precision on aqueous quality control materials was estimated using fifteen 860
systems. Seventeen runs per instrument were made over a twenty-two day period.
Two replicates of each control level were analyzed in each run.
For glucose, precision on aqueous calibration verification materials was estimated
using fifteen 860 systems. Data from seven to sixteen runs per instrument were
available. The data were collected over a two-month period. Two replicates of each
control level were analyzed in each run.
For lactate, precision on aqueous calibration verification materials was estimated
using fifteen 860 systems. Six runs per instrument were made over an eight day
period. Two replicates of each control level were analyzed in each run.
Table E-29 summarizes the results of the 860 system precision on QC materials.
)'( !" "$#"
!#!
%
#
pH6.5
1
387
7.157
0.002
0.004
2
316
7.423
0.001
0.004
3
235
7.617
0.003
0.006
pCO
2
1
293
72.5
1.12
1.87
Содержание Rapidlab 800
Страница 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Страница 13: ......
Страница 14: ......
Страница 20: ......
Страница 21: ......
Страница 22: ......
Страница 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Страница 24: ...4 0 0 4 1 0 4 4 4 1 4 4 4 4 4 0 4 0 4 0 4 4 4 0 4 4 2 4 0 2 4 4 0 3 4 0 3 4 4 3 0 3 4 0 0 3 4 0 0 3 4 4 1 34 4...
Страница 25: ......
Страница 26: ......
Страница 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Страница 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Страница 49: ...Screen elements are the components on the screen that enable you to interact with the system software...
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Страница 100: ......
Страница 164: ......
Страница 165: ......
Страница 166: ......
Страница 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Страница 168: ...6 0 6 3 6 45 6 6 6 3 6 0 0 0 6 6 3 0 6 0 0 6 6 4 6 1 0 6 0 0 6 6 6 6 6 2 0 4 6 2 6 0 0 6 6 4 6 4 6 1 2 6 6...
Страница 169: ......
Страница 170: ......
Страница 171: ......
Страница 172: ......
Страница 213: ......
Страница 214: ......
Страница 235: ......
Страница 236: ......
Страница 280: ......
Страница 281: ......
Страница 282: ......
Страница 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Страница 285: ......
Страница 286: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 339: ......
Страница 340: ......
Страница 364: ......
Страница 365: ......
Страница 366: ......
Страница 409: ......
Страница 410: ......
Страница 411: ...3 0 5 2 3 5 5 0 5 0 5 3 5 5 0 5 5 5 5 5 0 5 0 5 2 5 2 5 0 0 1 2 5 5 0 5 0 5 0 5 4 5 0 0 5 3 1 5 0 0 5 5 0 2 5 5 0 5...
Страница 496: ......
Страница 497: ......
Страница 498: ......
Страница 578: ......
Страница 620: ......
Страница 660: ......
Страница 661: ......
Страница 662: ......
Страница 663: ......
Страница 664: ......
Страница 665: ......
Страница 666: ......
Страница 671: ......
Страница 672: ......
Страница 687: ......
Страница 688: ......