6.
Ensure that the O-ring is in place. Replace the O-ring if it is worn or damaged.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
NOTE:
If the sensor does not insert easily into the measurement module, slide the
remaining sensors to the right to create more space.
7.
Reinstall the sensor:
a. Gently align the top of the sensor with the sensor contact.
b. Snap the body of the sensor down into place.
c. Press the tab on the spring-loaded latch to release the latch.
8.
Close the measurement module door.
9.
Press
Continue
.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
10. Press
No
.
11. Allow the system to warm up for at least 15 minutes.
12. Check the sample path temperature from the Menu screen:
a. Select
3 Troubleshooting
and press
Enter
.
b. Select
2 Temp/pAtm
and press
Enter
.
c. Press
Start Test
.
d. Press
Stop Test
.
e. Check the screen for the sample path temperature.
f. If the temperature control system is off, press
Reset Control
.
g. Press
Exit Test
.
13. Perform a two-point calibration from the Menu screen:
a. Select
1 Calibration
and press
Enter
.
b. Select
2 Two-point
and press
Enter
.
3
Menu Code
2
(from the Main Menu)
1
Menu Code
2
(from the Main Menu)
Содержание Rapidlab 800
Страница 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Страница 13: ......
Страница 14: ......
Страница 20: ......
Страница 21: ......
Страница 22: ......
Страница 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Страница 24: ...4 0 0 4 1 0 4 4 4 1 4 4 4 4 4 0 4 0 4 0 4 4 4 0 4 4 2 4 0 2 4 4 0 3 4 0 3 4 4 3 0 3 4 0 0 3 4 0 0 3 4 4 1 34 4...
Страница 25: ......
Страница 26: ......
Страница 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Страница 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Страница 49: ...Screen elements are the components on the screen that enable you to interact with the system software...
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Страница 100: ......
Страница 164: ......
Страница 165: ......
Страница 166: ......
Страница 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Страница 168: ...6 0 6 3 6 45 6 6 6 3 6 0 0 0 6 6 3 0 6 0 0 6 6 4 6 1 0 6 0 0 6 6 6 6 6 2 0 4 6 2 6 0 0 6 6 4 6 4 6 1 2 6 6...
Страница 169: ......
Страница 170: ......
Страница 171: ......
Страница 172: ......
Страница 213: ......
Страница 214: ......
Страница 235: ......
Страница 236: ......
Страница 280: ......
Страница 281: ......
Страница 282: ......
Страница 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Страница 285: ......
Страница 286: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 339: ......
Страница 340: ......
Страница 364: ......
Страница 365: ......
Страница 366: ......
Страница 409: ......
Страница 410: ......
Страница 411: ...3 0 5 2 3 5 5 0 5 0 5 3 5 5 0 5 5 5 5 5 0 5 0 5 2 5 2 5 0 0 1 2 5 5 0 5 0 5 0 5 4 5 0 0 5 3 1 5 0 0 5 5 0 2 5 5 0 5...
Страница 496: ......
Страница 497: ......
Страница 498: ......
Страница 578: ......
Страница 620: ......
Страница 660: ......
Страница 661: ......
Страница 662: ......
Страница 663: ......
Страница 664: ......
Страница 665: ......
Страница 666: ......
Страница 671: ......
Страница 672: ......
Страница 687: ......
Страница 688: ......