The pCO
2
sensor is based upon the electrode described by Severinghaus and
Bradley.
15
It is a complete electrochemical cell that consists of a measuring
electrode and an internal reference electrode. The measuring electrode, which is a
pH electrode, is surrounded by a chloride bicarbonate solution. A membrane
permeable to gaseous CO
2
separates this solution from the sample. The internal
reference electrode, which contains a silver/silver chloride electrode surrounded by
the chloride-bicarbonate solution, provides a fixed potential.
Sample Path
Measuring Electrode
Contact
Internal Reference
Electrode Contact
As the sample comes in contact with the membrane, CO
2
diffuses into
the chloride-bicarbonate solution, which causes a change in the hydrogen ion
activity.
CO
2
+ H
2
O
HCO
3
–
+ H
+
The internal pH electrode detects the change in hydrogen concentration occurring
in the chloride bicarbonate solution and generates a half-cell potential. This
potential, when compared to the fixed potential of the reference electrode, results
in a measurement that reflects pH change in the chloride bicarbonate solution. The
change in pH is related to the log of the partial pressure of CO
2
.
Oxygen (O
2
) is essential for cell and tissue metabolism in the body. The
cardiopulmonary system is responsible for transporting oxygen to the cells.
Oxygen transport involves four major steps: convection and diffusion from the air
into the pulmonary circulation, combination of O
2
from the lungs with hemoglobin
in red blood cells, transportation of the O
2
through the arteries to the cell, and
finally the release into the tissues and utilization of O
2
at the cellular level.
Содержание Rapidlab 800
Страница 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Страница 13: ......
Страница 14: ......
Страница 20: ......
Страница 21: ......
Страница 22: ......
Страница 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Страница 24: ...4 0 0 4 1 0 4 4 4 1 4 4 4 4 4 0 4 0 4 0 4 4 4 0 4 4 2 4 0 2 4 4 0 3 4 0 3 4 4 3 0 3 4 0 0 3 4 0 0 3 4 4 1 34 4...
Страница 25: ......
Страница 26: ......
Страница 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Страница 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Страница 49: ...Screen elements are the components on the screen that enable you to interact with the system software...
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Страница 100: ......
Страница 164: ......
Страница 165: ......
Страница 166: ......
Страница 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Страница 168: ...6 0 6 3 6 45 6 6 6 3 6 0 0 0 6 6 3 0 6 0 0 6 6 4 6 1 0 6 0 0 6 6 6 6 6 2 0 4 6 2 6 0 0 6 6 4 6 4 6 1 2 6 6...
Страница 169: ......
Страница 170: ......
Страница 171: ......
Страница 172: ......
Страница 213: ......
Страница 214: ......
Страница 235: ......
Страница 236: ......
Страница 280: ......
Страница 281: ......
Страница 282: ......
Страница 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Страница 285: ......
Страница 286: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 339: ......
Страница 340: ......
Страница 364: ......
Страница 365: ......
Страница 366: ......
Страница 409: ......
Страница 410: ......
Страница 411: ...3 0 5 2 3 5 5 0 5 0 5 3 5 5 0 5 5 5 5 5 0 5 0 5 2 5 2 5 0 0 1 2 5 5 0 5 0 5 0 5 4 5 0 0 5 3 1 5 0 0 5 5 0 2 5 5 0 5...
Страница 496: ......
Страница 497: ......
Страница 498: ......
Страница 578: ......
Страница 620: ......
Страница 660: ......
Страница 661: ......
Страница 662: ......
Страница 663: ......
Страница 664: ......
Страница 665: ......
Страница 666: ......
Страница 671: ......
Страница 672: ......
Страница 687: ......
Страница 688: ......