Operation Manual
15 - 77
15.4.1 Fluid Delivery Setup Configurable Items
15.4.2 Contrast Type Setup
15.4.2.1 Adding a New Contrast Type
NOTE:
The system is not loaded with any contrast agent information at the time of manufacture. The operator
must create the list of contrast agents by adding new contrast types. Refer to contrast manufacturer’s
most recent dosing information from the contrast package insert when adding contrast agents.
1.
Select
Contrast Types
on the Fluid Delivery Setup screen.
2.
Select
Add New
(1) on the Contrast Configuration screen.
Configurable Item
Description
Priming Source
Sets the syringe to be used for priming the tubing. Select A to configure Syringe A (contrast) to
be the priming source. Select B to configure Syringe B (saline) to be the priming source.
Test Inject
Enable or disable test injection. Sets the default flow rate, volume, and duration for the test
injection.
Protocol Programming
Mode
Sets the programming mode to enable the system to calculate the Duration, Flow Rate, or
Volume. Programming Modes:
Q
Calculate Flow Rate: System calculates the flow rate based on volume and duration.
Q
Calculate Volume: System calculates the volume based on flow rate and duration.
Q
Calculate Duration: System calculates the duration based on flow rate and volume.
Q
Disable: Default setting. System calculates the duration based on flow rate and volume.
Total Volume Display
Sets the total volume to display the volumes in Syringe A and Syringe B or the total in Syringe A
and B combined.
Auto Retract
Enables or disables the Auto Retract feature to retract the piston after a syringe is removed.
Auto Advance
Enable or disables the Auto Advance feature to advance the piston after a syringe is installed.
Auto Load Purge
Volume
Auto Load Purge Volume can be set to 10 mL, 5 mL, 2 mL or 0 mL.
Contrast Types
Set or add types of contrast. Refer to
"Section 15.4.2 - Contrast Type Setup"
for more
information.
Reset Protocol after
Injection
Sets whether or not the protocol is reset to the factory default protocol after an injection
completes.
KVO Interval
Sets the interval at which the KVO pulse is delivered. To disable KVO, set the interval to “None.”
Slow Forward Load
Rate
Sets the rate at which fluid is manually loaded or purged using the forward and reverse piston
control buttons.
Slow Reverse Load Rate
Fast Forward Load Rate
Fast Reverse Load Rate
Содержание MEDRAD MRXperion
Страница 1: ...Operation Manual...
Страница 2: ......
Страница 4: ...MEDRAD MRXperion...
Страница 10: ...vi MEDRAD MRXperion...
Страница 18: ...2 8 MEDRAD MRXperion...
Страница 22: ...3 12 MEDRAD MRXperion...
Страница 30: ...4 20 MEDRAD MRXperion...
Страница 38: ...5 28 MEDRAD MRXperion...
Страница 42: ...6 32 MEDRAD MRXperion...
Страница 46: ...7 36 MEDRAD MRXperion...
Страница 62: ...10 52 MEDRAD MRXperion...
Страница 78: ...13 68 MEDRAD MRXperion...
Страница 80: ...14 70 MEDRAD MRXperion...
Страница 92: ...15 82 MEDRAD MRXperion...
Страница 106: ...17 96 MEDRAD MRXperion...
Страница 118: ...18 108 MEDRAD MRXperion...
Страница 125: ......