Cleaning and Maintenance
15 - 69
1.
Ensure the injector head has been thoroughly cleaned. Refer to
.
2.
Use Sani-Cloth
®
Plus or CaviWipes™ to thoroughly wipe all external surfaces with the
exception of the syringe piston and syringe interface. Make certain that all seams, recessed
areas, and buttons are clean.
NOTE:
If in a country where Sani-Cloth
®
Plus or CaviWipes™ are not available, and using an
equivalent disinfecting agent (refer to the materials listed above), follow the same instructions
specified in step 2.
3.
Using Sani-Cloth
®
Plus or CaviWipes™, allow surfaces to remain visibly wet for 3 minutes. If
needed, use additional wipes to ensure the surfaces remain wet for the full duration.
NOTE:
If in a country where Sani-Cloth
®
Plus or CaviWipes™ are not available, and using an
equivalent disinfecting agent (refer to the materials listed above), allow the surface to remain
visibly wet per the manufacturer’s instructions.
4.
Allow the injector head to thoroughly air dry.
15.3 Daily
The following procedures are recommended for daily inspection of the system.
Before use each day, the system should be visually inspected, using the procedures outlined in this
section, to ensure that all surfaces are free from defects and visible contamination. If any defects are
detected, either repair the system, or call the local office for Bayer or the local authorized dealer for
service. Do not use the system until the problem is corrected. If visible contamination is evident on any
surface of the device, follow the cleaning process outlined below before using the system.
15.3.1 Inspecting the Injection System
Frequency:
Daily
•
Inspect the housing for any damage or cracks that could allow fluid to leak inside or weaken
the structural integrity of the unit.
•
Inspect all cables connected to the system. Look for cuts, cracks, worn spots, or other obvi-
ous damage to the cables. Ensure all connectors are properly seated.
•
Inspect the pedestal, base, overhead counterpoise system, and support arm for cracks and
other defects that could weaken the structure.
•
Ensure all mounting bolts and screws are secure.
•
Inspect the pivot points. The head and support arm must pivot freely. The injector head
should rotate on the support arm no more than 270°. The support arm should not rotate on
the center post more than 350°.
•
Ensure the casters roll smoothly with no binding or scraping.
•
Ensure all locking mechanisms on the casters are functional.
NOTE:
All relevant guidelines for facility, local, or national safety recommendations related to cable
routing and installation should be followed.
NOTE:
Contact Bayer or a local dealer for service or repairs.
15.3.2 Cleaning and Disinfecting the Injector Head
Frequency:
Daily
1.
Ensure the injector head has been thoroughly cleaned. Refer to
.
2.
Ensure the injector head has been thoroughly disinfected. Refer to
.
Содержание Certegra
Страница 2: ......
Страница 4: ...MEDRAD Stellant CT Injection System with Certegra Workstation ii ...
Страница 18: ...MEDRAD Stellant CT Injection System with Certegra Workstation 2 10 ...
Страница 34: ...MEDRAD Stellant CT Injection System with Certegra Workstation 6 26 ...
Страница 36: ...MEDRAD Stellant CT Injection System with Certegra Workstation 7 28 ...
Страница 54: ...MEDRAD Stellant CT Injection System with Certegra Workstation 10 46 ...
Страница 58: ...MEDRAD Stellant CT Injection System with Certegra Workstation 11 50 ...
Страница 60: ...MEDRAD Stellant CT Injection System with Certegra Workstation 12 52 ...
Страница 70: ...MEDRAD Stellant CT Injection System with Certegra Workstation 13 62 ...
Страница 86: ...MEDRAD Stellant CT Injection System with Certegra Workstation 16 78 ...
Страница 103: ......