Cleaning and Maintenance
15 - 67
ping) for a minimum of 1 minute and until visibly clean. Replace wipe if wipe becomes visibly
soiled. During the 1 minute cleaning time, make certain that all seams, recessed areas and
buttons are clean. Refer to
"Cleaning the Pistons and Syringe Interface"
pistons and syringe interface.
3.
Thoroughly dry the injector head using a dry, clean, soft, non-linting wipe.
4.
Wipe all external surfaces with the exception of the syringe pistons and syringe interface
(refer to Figure 15 - 3: Piston) with Sani-Cloth
®
Plus or CaviWipes™ for 1 minute or until vis-
ibly clean. Make certain that all seams, recessed areas, and buttons are clean. Refer to
"Cleaning the Pistons and Syringe Interface"
to clean the syringe pistons and syringe inter-
face.
NOTE:
If in a country where Sani-Cloth
®
Plus or CaviWipes™ are not available, and using an
equivalent cleaning agent (refer to the materials listed above), follow the same instructions
specified in step 4.
5.
Allow the injector head to thoroughly air dry.
6.
Inspect the injector head to ensure all surfaces are clean.
7.
If any visible contrast media or other debris is evident, repeat cleaning instructions steps 2
through 6 until no visible soil is detected.
15.2 Daily and In the Case of Visible Contamination
NOTE:
To clean the syringe pistons or syringe interface, refer to
"Cleaning the Pistons and Syringe
for instructions.
15.2.1 Cleaning the Injector Head
Figure 15 - 2: Injector Head
Electro-Mechanical Hazard - Equipment damage may result.
•
Do not use cleaning agents containing quaternary ammonium compounds (e.g. dimethyl
ethylbenzyl ammonium chloride), such as Sani-Cloth
®
Plus, CaviWipes™, ZEP
®
brand
cleaners, and/or ethyl alcohol, such as Lysol
®
and Clorox
®
on the syringe pistons or
syringe interface.
NOTICE
Содержание Certegra
Страница 2: ......
Страница 4: ...MEDRAD Stellant CT Injection System with Certegra Workstation ii ...
Страница 18: ...MEDRAD Stellant CT Injection System with Certegra Workstation 2 10 ...
Страница 34: ...MEDRAD Stellant CT Injection System with Certegra Workstation 6 26 ...
Страница 36: ...MEDRAD Stellant CT Injection System with Certegra Workstation 7 28 ...
Страница 54: ...MEDRAD Stellant CT Injection System with Certegra Workstation 10 46 ...
Страница 58: ...MEDRAD Stellant CT Injection System with Certegra Workstation 11 50 ...
Страница 60: ...MEDRAD Stellant CT Injection System with Certegra Workstation 12 52 ...
Страница 70: ...MEDRAD Stellant CT Injection System with Certegra Workstation 13 62 ...
Страница 86: ...MEDRAD Stellant CT Injection System with Certegra Workstation 16 78 ...
Страница 103: ......