MEDRAD
®
Stellant CT Injection System with Certegra
®
Workstation
6 - 24
6.2 Injector Head Components
6.2.1 Syringe Heat Maintainer
It is recommended that contrast be at 35 degrees Celsius prior to loading it into the syringe. Once the
syringe is loaded, use the syringe heat maintainer to maintain the contrast temperature.
B
Syringe B
Saline syringe
1
Heat Maintainer
(Side A or B)
Refer to
for more information.
2
Volume
(Side A or B)
•
Syringe Installed
- Indicates the volume loaded in the
syringe.
•
Auto Load Active
- Indicates the volume to be loaded
into the syringe.
•
Syringe not Present
- No indicator.
3
Auto Load
Activates the Fill A, Fill B, and the +/- buttons.
4
Fill A
Fills syringe A to the displayed volume.
5
Fill B
Fills syringe B to the displayed volume.
6
+/-
(Side A or Side B)
Adjusts the Auto Load volume to the nearest multiple of 5 mL, and
then in increments/decrements of 5 mL.
7
Prime
Activates the tube priming function.
8
Move Piston
Activates the forward and reverse piston controls. Times out after
ten seconds of inactivity.
9
Piston Control
(Side A or B)
Advances and retracts the piston (variable speed).
10
Check for Air
Reminds user to verify that air is purged from the syringe and
tubing. (Must be lit to enable arming.)
11
Protocol Lock
Indicates the protocol lock on display is enabled. (Must be lit to
enable arming.)
12
Arm
Arms the system.
13
Abort
Terminates and disarms the injection.
14
Start/Hold
Initiates injection.
Holds injection for a maximum of 20 minutes.
15
Arm Lights
Refer to
for more information.
16
Manual Knob
(Side A or B)
Permits an operator to manually move the piston when the injector
is not armed.
Name
Description
Mechanical Hazard - Minor or moderate patient and/or worker injury may result.
•
Do not use the syringe heat maintainer if the fault indicator light is illuminated.
CAUTION
Содержание Certegra
Страница 2: ......
Страница 4: ...MEDRAD Stellant CT Injection System with Certegra Workstation ii ...
Страница 18: ...MEDRAD Stellant CT Injection System with Certegra Workstation 2 10 ...
Страница 34: ...MEDRAD Stellant CT Injection System with Certegra Workstation 6 26 ...
Страница 36: ...MEDRAD Stellant CT Injection System with Certegra Workstation 7 28 ...
Страница 54: ...MEDRAD Stellant CT Injection System with Certegra Workstation 10 46 ...
Страница 58: ...MEDRAD Stellant CT Injection System with Certegra Workstation 11 50 ...
Страница 60: ...MEDRAD Stellant CT Injection System with Certegra Workstation 12 52 ...
Страница 70: ...MEDRAD Stellant CT Injection System with Certegra Workstation 13 62 ...
Страница 86: ...MEDRAD Stellant CT Injection System with Certegra Workstation 16 78 ...
Страница 103: ......