MEDRAD
®
Stellant CT Injection System with Certegra
®
Workstation
4 - 16
4.4 Programming Mode
After values for any two parameters are entered, the system calculates the value for the third. By
default, values for flow rate and volume are entered, and the system calculates duration. Programming
Mode is used to toggle the value calculated by the system. The system can calculate the flow rate, vol-
ume, or duration. To enable Programming Mode, refer to
for more information.
4.5 Basic Informatics
If an optional Informatics platform is included with the system, refer to the Certegra
®
Applications and
Workstation Accessories manual for overview and functionality.
4.6 Using the Pedestal with Integrated IV Pole
The pedestal is intended to support the injector head in a CT scan room. Do not attempt to use the ped-
estal for any other purpose.
4.6.1 Pedestal Operating Instructions
•
To extend Integrated IV Pole:
Turn adjustment knob counter-clockwise to loosen, then raise
the top of the Integrated IV Pole to the desired height.
•
To lock the Integrated IV Pole:
Turn the adjustment knob clockwise until tight. Do not
overtighten.
•
To store:
Turn the adjustment knob counter-clockwise and retract the Integrated IV Pole.
Mechanical Hazard - Minor or moderate patient and/or worker injury may result.
•
Tighten all screws, clamps, and knobs during assembly and as needed during use. Loose
components may cause the pedestal to collapse.
•
Do not move or manipulate the injector mount by pulling or pushing the integrated IV pole.
Using the pole to maneuver the injector could bend the pole or cause an imbalance of the
system. Maneuver the injector as recommended. Refer to
•
Do not adjust the integrated IV pole or move the injector when anything is hung from the
pole.
•
Use care and diligence in folding and unfolding the hooks, raising and lowering the pole,
and tightening the adjustment knob. Keep hands and fingers clear of all pinch point areas.
•
The folding hooks are designed to hold a maximum weight of 5 lbs / 2,265 grams each.
Do not exceed the weight limits.
•
Use extreme care in raising and lowering the integrated IV pole. The pole should raise and
lower easily without force. If raising and lowering becomes difficult, clean the pole as
described in
"Cleaning the Pedestal With Integrated IV Pole, Overhead Counterpoise
.
Electric Shock Hazard - Minor or moderate patient and/or worker injury may result.
•
Only plug the system into a direct mains access point. Do not plug the system’s power
cord into an extension cord or multi-outlet power strip.
Procedure Delay Hazard - Serious patient and/or worker injury or death may result.
•
Turn off any equipment that could generate a high level electrostatic discharge.
CAUTION
Содержание Certegra
Страница 2: ......
Страница 4: ...MEDRAD Stellant CT Injection System with Certegra Workstation ii ...
Страница 18: ...MEDRAD Stellant CT Injection System with Certegra Workstation 2 10 ...
Страница 34: ...MEDRAD Stellant CT Injection System with Certegra Workstation 6 26 ...
Страница 36: ...MEDRAD Stellant CT Injection System with Certegra Workstation 7 28 ...
Страница 54: ...MEDRAD Stellant CT Injection System with Certegra Workstation 10 46 ...
Страница 58: ...MEDRAD Stellant CT Injection System with Certegra Workstation 11 50 ...
Страница 60: ...MEDRAD Stellant CT Injection System with Certegra Workstation 12 52 ...
Страница 70: ...MEDRAD Stellant CT Injection System with Certegra Workstation 13 62 ...
Страница 86: ...MEDRAD Stellant CT Injection System with Certegra Workstation 16 78 ...
Страница 103: ......