Menu
Description
Password level /
BCP command
Min.thr-
Minimum value for negative flow
If the minimum value for the negative flow falls
below the set value, an alarm message is gen-
erated.
The value of this parameter is set as a per-
centage (0 - 125%) of the scale end value.
If the parameter is 0, the alarm generation is
deactivated.
L 3
FRANN
T1 max
Maximum value for temperature T1
If the maximum value for the temperature is
exceeded, an alarm message is generated.
If the parameter is 0, the alarm generation is
deactivated.
L 3
TMP1X
T1 min
Minimum value for temperature T1
If the minimum value for the temperature is
exceeded, an alarm message is generated.
If the parameter is 0, the alarm generation is
deactivated.
L 3
TMP1N
Hysteresis
Hysteresis threshold for minimum and max-
imum flow
The value of this parameter is expressed as a
percentage of the scale end value and can be
set from 0 - 25%.
L 3
ATHYS
mA v.alarm
Alarm value for output current
The alarm value for the output current can be
triggered when an empty line, a broken coil or
an ADC error occurs.
The value is set as a percentage (0 - 125%) of
the 0 - 20 mA current.
It is recommended to set the value of this func-
tion to 10% so that the maximum deviation is a
maximum of 2 mA in all cases.
L 3
OCACV
Hz v.alarm
Alarm value for frequency value
The value is set as a percentage (0 - 125%) of
the frequency.
L 3
OFACV
Operation
Menu navigation > Alarms menu
12.08.2020
CombiFlow PF75H/S
86