It may be appropriate to use a constant time for the damping filter,
for example for a pulsating flow.
The flow is averaged over a number of readings. The damping filter
dampens the measurement noise and the sudden change in the
flow rate. Increasing the damping parameter increases the stability
of the measurement.
Longer times produce a stable average. For short times, the meas-
urement follows the measured values closely and is therefore more
unstable.
7.3.4.1
Measure
menu - additional BCP commands
The following menus are also available as an BCP command in the
command input of the BCP software:
Menu
Description
Password level /
BCP command
Measure cut-off
threshold 2
Threshold for low flow 2
The low flow shutdown threshold setting is the
same as
Cut-off
.
L 3
MFCT2
High immunity input
noise filter
Input noise filter with high immunity to
interference
When this feature is enabled, an immunity to
measurement is activated that is approximately
1%.
L 4
HIINP
Dynamic sample
analysis
Dynamic analysis
L 4
DINSA
Dynamic sample time
Dynamic analysis time
L 4
DYNST
Damping filter time-dependent
Operation
Menu navigation > Measure menu
12.08.2020
CombiFlow PF75H/S