7.3.2
Units
menu
The functionality and available capabilities of the flow
sensor vary depending on the order configuration and
type code.
Depending on the configuration of the flow sensor or
activated functions, individual menus and selection
functions may differ from the representations or be
hidden.
NOTICE!
Accuracy losses when changing the units of meas-
urement
The total counters and part counters are updated
depending on the setting of the units.
The rounding caused by this can lead to loss of accu-
racy.
Example:
–
Total counter T + = 0.234 l (unit with 3 decimal
places)
–
Change to unit m³
–
Total counter T + = 0.001 m³ (0.234 liters have
been lost due to rounding)
Menu
Description
Password level /
BCP command
Sens.diameter unit
of measure
Nominal diameter unit
n
mm
n
inches
L 2
SDIUM
FR.unit
Flow rate measuring system
n
Metric measurement system
n
Anglo-American measurement system
L 2
FRMUT
Pl1 unit
Measurement system for Impulse 1
n
Metric measurement system
n
Anglo-American measurement system
L 2
PL1UT
Pl2 unit
Measurement system for Impulse 2
n
Metric measurement system
n
Anglo-American measurement system
L 2
PL2UT
Operation
Menu navigation > Units menu
12.08.2020
CombiFlow PF75H/S
76