Temperature Compensation
8
♦
To use
the current measured temperature
for temperature compensation,
set parameter
Temperature Compensation
to
sens
(
t.comp sens#13#10
).
♦
For manual temperature compensation,
using the value of the parameter
Default Temperature
,
set parameter
Temperature Compensation
to
fixed
(
t.comp fixed#13#10
).
Set a
Default Temperature
value
as close as possible
to the operating temperature value.
♦
To turn off temperature compensation,
set parameter
Temperature Compensation
to
off
(
t.comp off#13#10
).
Input Filtration
Low-pass filter
This first-order filter acts ONLY within
a certain band around filter output value.
This has been designed to cut periodic
noises outside the communication signal
spectrum.
♦
Filter operation is defined by two
parameters:
Filter Time
(defines filter time constant) and
Filter Band
(defines filter active band
around filter output value).
♦
If the newly measured value differs
from the filter output by more than
Filter Band
, the filter resets
with a new initial output value
(newly measured value).