Overview
2
BCOT751 is a smart isolated
conductivity transmitter with automatic
temperature compensation.
The device is fully programmable
with all its parameters accessible by PC
via its communication interface.
Mounting and Wiring
Mounting
BCOT751 can be easily mounted on every
35 mm rail conforming to EN50022.
Wiring
♦
Connect a conductivity cell
to terminals 5 and 6(7) and
a temperature compensation sensor
to terminals 7(6) and 8.
Terminals 6 and 7
are internally connected.
♦
Get the analog output signal
from terminals 2(+) and 4(-).
♦
Wire the relay output
via terminals 3 and 4.
♦
Connect the right power supply
voltage (see 'Specifications')
via terminals 1(+) and 4(-).
♦
In order to minimize measuring errors,
make sure the connecting screws
are tightened enough.