Communication Protocol
7
invalid command.
command not recognized
parity error.
parity error detected
not a number.
attempt to write symbols
for numerical parameter
point error.
value resolution greater
than parameter’s one
out of range.
value out of range
read only.
parameter is read-only
can't save.
problem with writing
in non-volatile memory
Other device responses
♦
BCOT751 responses in case
of incorrect protocol use
are given on the left.
♦
When
Error Info
value is
-1
,
the device substitutes
any command for
error
reading (see 'Errors').
Reset
To reset the device,
send command
reset
.
Errors
Value
Parameters
Error type
-1
all
incorrect memory
0
all
no error
1
Cell Constant
out of range
2
Filter Time
out of range
3
Filter Band
out of range
4
Temperature Coefficient
out of range
5
Default Temperature
out of range
11
Hysteresis
out of range *
12
Set Point
-
Hysteresis
under range *
13
Set Point
+
Hysteresis
over range *
21
Hysteresis
out of range **
22
Set Point
-
Hysteresis
under range **
23
Set Point
+
Hysteresis
over range **
31
Hold Time On Error
out of range
41
Wire Resistance
out of range
* with
Relay Output Link
=
temp
** with
Relay Output Link
=
cond
♦
Writing in the device
may incur discrepancies
in parameter values
that must be resolved
before further operation.
♦
To check for an error,
send command
error
.
♦
In case of value
-1
,
try resetting by sending
command
reset
.
♦
If the problem persists,
send command
error
0
to restore the default
(factory) settings.