Calibrating
10
Calibration by setting cell constant
♦
To use this method,
set parameter
Calibration Mode
to
c.set
(
cal c.set#13#10
).
♦
Assign the precise cell constant value,
taken from the certificate
provided with your sensor
or from the sensor label,
to the parameter
Cell Constant
(
const x.xxxxxx#13#10
).
Calibration by adjusting measured value
♦
To use this method,
set parameter
Calibration Mode
to
c.cal
(
cal c.cal#13#10
).
♦
To take temperature compensation
into account, select the desired
compensation method
via parameter
Temperature Compensation
.
♦
Place the conductivity cell
into calibration solution
and wait for the measured value
to stabilize (
c.v#13#10
).
♦
Assign the calibration solution’s
conductivity value
to the
Conductivity Value
parameter
(
c.v xx.xx#13#10
).