Communication Protocol 

6

 

 Notes: 

 

BCOT751 adds 3 spaces
in the beginning of the response. 

 

BCOT751 returns decimal point
even when the value is integer. 

 

#13 (CR) is byte 0x0D;
#10 (LF) is byte 0x0A. 

 Protocol examples: 

PC or other device: 

  BCOT751 response: 

reading filter time 

f.t#13#10

f.t 0015.#13#10

writing filter time of 30 

f.t 30#13#10

f.t 0030.#13#10

reading measured conductivity 

c.v#13#10

c.v 027.5#13#10

Protocol architecture 

 

The protocol is based 
on UART protocol with: 

 

Baud Rate - 9600 bps; 

 

Data bits - 8; 

 

Parity Control - Even; 

 

Stop bit - 1. 

 

ASCII protocol is used 
for communicating,  
and the information 
is exchanged in frames. 

 

Each frame consists of 1, 
or 2 words separated  
by byte 32 (SPACE),  
and ends with bytes 13 (CR) 
and 10 (LF). The first word  
in the frame denotes  
the parameter 'Symbol'  
as taken from Table 1 and 
the second word (if needed)  
is the parameter 'Value', 
both spelled with only  
small Latin letters (except for 

Conductivity Unit

 

value), digits, 

dots, and/or the '-' sign. 

Reading from a device 

 

If the frame consists of 
only 1 word, it is recognized 
as a command for reading.  

 

The device responds to it by 
returning the same word and 
its value, according to Table 1.  

Writing in a device 

 

If the frame consists of
2 words, it is recognized  
as a command for writing. 

 

With writing, transferred are
the same 2 words that  
would have been received  
at the respective command 
for reading from the device 
(except for the 

reset 

command). 

Содержание BCOT751

Страница 1: ...SJ SWEDEN tel 46 0 40 88009 fax 46 0 40 929877 e mail info basi se BCOT751 OPERATION MANUAL Please read this Operation Manual before mounting and operating Save the Manual for future references SMART...

Страница 2: ...n every 35 mm rail conforming to EN50022 Wiring Connect a conductivity cell to terminals 5 and 6 7 and a temperature compensation sensor to terminals 7 6 and 8 Terminals 6 and 7 are internally connect...

Страница 3: ...nctionality is installed see Specifications Changing Point Position value reflects on the real value of all parameters with the conductivity measurement unit E g changing Point Position value from 0 t...

Страница 4: ...Resistance of the connection wires Output Parameters Output Configuration o conf Analog output signal type Analog Output Link o lnk Defines input linked to the analog output Low Limit o lo Input valu...

Страница 5: ...00000 1 cm 5 6 00 00 99 99 3 used for long wire resistance compensation i 0 20 i 4 20 u 0 10 u 2 10 i 0 20 0 20 mA i 4 20 4 20 mA u 0 10 0 10 V u 2 10 2 10 V cond temp cond conductivity input temp tem...

Страница 6: ...ed in frames Each frame consists of 1 or 2 words separated by byte 32 SPACE and ends with bytes 13 CR and 10 LF The first word in the frame denotes the parameter Symbol as taken from Table 1 and the s...

Страница 7: ...all incorrect memory 0 all no error 1 Cell Constant out of range 2 Filter Time out of range 3 Filter Band out of range 4 Temperature Coefficient out of range 5 Default Temperature out of range 11 Hyst...

Страница 8: ...rn off temperature compensation set parameter Temperature Compensation to off t comp off 13 10 Input Filtration Low pass filter This first order filter acts ONLY within a certain band around filter ou...

Страница 9: ...rs Set Point Direction and Hysteresis When an error is detected the relay output deactivates and the analog output is held for a certain time defined by the parameter Hold Time On Error The analog out...

Страница 10: ...const x xxxxxx 13 10 Calibration by adjusting measured value To use this method set parameter Calibration Mode to c cal cal c cal 13 10 To take temperature compensation into account select the desire...

Страница 11: ...together with household waste material If disposed of within European Union this product should be treated and recycled in accordance with the laws of your jurisdiction implementing Directive 2012 19...

Страница 12: ...ng IP20 Warranty and Support article number serial number manufacturing date QC check mark passed stamp BASI Instrument AB p o box 53 SE 275 06 VOLLSJ SWEDEN tel 46 0 40 88009 fax 46 0 40 929877 e mai...

Отзывы: