28
INSTALLATION:
(Con’t)
INSTALLATION OF GAS FILTER INTERFACE ASSEMBLY KIT #018-514-00:
KIT #018-514-00 INCLUDES:
•
4, Hose Clamps (#250-331-00)
•
1, In-Line Filter (#525-034-00)
•
1, 10 ft. (4.57m) length of 1/4” (6.35mm) ID PVC tubing (#720-418-00)
INSTALLATION PROCEDURE:
See next page for illustration.
•
Turn the unit power switch to the
OFF
position.
•
Connect tubing to CO
2
tank gas outlet.
•
Connect 1/4” (6.35 mm) ID tubing to hose barb on one side of in-line filter
and secure with hose clamp; connect tubing to in-line filter hose barb on its
other side and secure with hose clamp.
•
Connect tubing with in-line filter in place to the appropriate gas inlet fitting
on the back of the unit.
•
Set the supply regulators at 10 psi.
•
Check all of the connections for leaks.
•
Turn the unit power switch to the
ON
position.
WARNING:
HIGH CONCENTRATIONS OF CARBON DIOXIDE PRODUCE
METABOLIC ABNORMALITIES, DISTURBANCES OF THE CENTRAL
NERVOUS SYSTEM AND CARDIAC INSTABILITY. UNCONSCIOUSNESS
MAY OCCUR AT CONCENTRATIONS ABOVE 10%.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com