TEMP COMMAND KEY: Pressing the
SET
key and then the
TEMP
command key allows the user with the Up and Down arrow keys to
establish the desired set point temperature in the Incubator.
•
High and low limit alarm set points are factory-set at ±1ºC. If the
temperature deviates from these limits, an audible, continuous alarm
will sound and the temperature lamp will flash.
•
While pressing any key will silence the alarm for a one-hour period,
the user should immediately investigate the reason for the alarm and
take steps to correct the problem. Note that indicating light continues
to flash. If problem is not corrected, the alarm will resume after one
hour has passed.
•
By pressing the CAL (#3) and TEMP (#5) keys sequentially users
can calibrate temperature. The Up and Down arrow keys will change
the readout to correspond with the readings obtained from known
standards of temperature measurement.
NOTE:
DECREASES IN TEMPERATURE AND CO2 DUE TO NORMAL DOOR OPENINGS ARE
TOLERATED WITHOUT INITIATING THE ALARMS.