Choosing a Site
The NANOpure DIamond UF features a removable con-
trol panel display which allows the system to be mounted
almost anywhere within the laboratory. Use the wall
bracket for wall mounted systems as a template to drill
mounting holes. (The NANOpure DIamond UF does not
include screws and fasteners for mounting.) Please refer
to the “General Specifications” section for clearance
requirements.
Installation
NOTE
The outlet of a gravity feed storage
reservoir must be above or at the
same level as the inlet of the
NANOpure DIamond UF.
********************
************************
Caution
Wall composition, condition and
construction as well as fastener type
must be considered when mounting
this unit. The mounting surface and
fasteners selected must be capable of
supporting a minimum of 150 lbs. (68
kg). Inadequate support and/or fasten-
ers may result in damage to mounting
surface and/or equipment. If you are
unsure of mounting surface
composition, condition and construc-
tion or correct fasteners, consult your
building maintenance group or
contractor.
Warning
Do not locate the NANOpure DIamond
UF directly over equipment that
requires electrical service. Routine
maintenance of this unit may involve
water spillage and subsequent electri-
cal shock hazard if improperly located.
Do not use in the presence of
flammable materials; fire or explosion
may result. This device contains
components which may ignite such
materials.