8. Getting started
ECR calibration must be performed in a perfectly dark environment. Any ambient light or stray light will lead to a bad LUT.
How to start the ECR calibration process?
1. Open the main page of the XDS RACU user interface.
2. Select a projector.
3. Navigate to
Projector > Calibration
and click it.
Image 8-95
Calibration: button
Calibration page is displayed.
Image 8-96
Calibration page
4. If you wish to calibrate more than one projector, select the desired projectors or select all.
Note:
Multiple projector selection is possible in the Calibration page.
5. Click the
Start Calibration
button.
82
R59770509 MES ACURAS HEAD - AUTO ALIGNMENT HEAD 24/11/2010
Содержание MES ACURAS HEAD
Страница 1: ...MES Acuras Head Auto Alignment Head User and Installation guide R9843500 R59770509 03 24 11 2010...
Страница 8: ......
Страница 20: ...2 Safety 12 R59770509 MES ACURAS HEAD AUTO ALIGNMENT HEAD 24 11 2010...
Страница 22: ...3 Compliancy 14 R59770509 MES ACURAS HEAD AUTO ALIGNMENT HEAD 24 11 2010...
Страница 28: ...4 Packaging 20 R59770509 MES ACURAS HEAD AUTO ALIGNMENT HEAD 24 11 2010...
Страница 112: ...10 Operating AutoGeometry 104 R59770509 MES ACURAS HEAD AUTO ALIGNMENT HEAD 24 11 2010...
Страница 126: ...Index 118 R59770509 MES ACURAS HEAD AUTO ALIGNMENT HEAD 24 11 2010...
Страница 127: ...List of tables LIST OF TABLES 1 1 Documentation 3 R59770509 MES ACURAS HEAD AUTO ALIGNMENT HEAD 24 11 2010 119...
Страница 128: ...List of tables 120 R59770509 MES ACURAS HEAD AUTO ALIGNMENT HEAD 24 11 2010...