IMPLEMENTATION OF A MEASUREMENT
5.2.3 Mounting the impact device
Press the placement ring of the impact device firmly and without wobbling onto
the test specimen. The impact direction must correspond to the set-adjusted
direction.
5.2.4 Measuring
Press release button above on the impact device. Sample and impact device
must be held steady and stable in this case.
For every measuring point, 5 measurements should be carried out whose
deviation should not exceed ±15 HL .
The minimum distance between two measuring points, and the minimum
distance between a measuring point and the edge of the sample, should
correspond to the specifications in 1.
Impact device type
Separation distance
between the center points
of two impressions
Separation distance between
the center point of an
impression and the edge of
the sample
Not less than /mm
Not less than /mm
D / DC
3
5
DL
3
5
D+15
3
5
G
4
8
C
2
4
Table 1